On-line preconcentration system for lanthanum determination in urine using FI-ICP-AES

被引:22
|
作者
Vicente, O [1 ]
Masi, A [1 ]
Martinez, L [1 ]
Olsina, R [1 ]
Marchevsky, E [1 ]
机构
[1] Univ Nacl San Luis, Fac Quim Bioquim & Farm, Area Quim Analit, RA-5700 San Luis, Argentina
关键词
flow injection; ICP-AES; preconcentration and determination; lanthanum in urine samples;
D O I
10.1016/S0003-2670(98)00105-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An on-line lanthanum preconcentration and determination system implemented with inductively coupled plasma-atomic emission spectrometry (ICP-AES) associated to flow injection (FI) method was studied. For the retention of lanthanum, quinolin-8-ol and Amberlite XAD-7 were used at pH 9.5. The lanthanum complex was removed from the micro-column with nitric acid. The value of detection limit for the preconcentration of 100 mi of aqueous solution was 0.09 ng ml(-1) with a relative standard deviation (RSD) of 1.5%, calculated with the peak heights obtained. The calibration graph using the preconcentration system for lanthanum was linear with a correlation coefficient of 0.9993 at levels near the detection limits up to at least 100 ng ml(-1). The method was successfully applied to the determination of lanthanum in urine samples. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:201 / 207
页数:7
相关论文
empty
未找到相关数据