Reconstruction of high resolution MLC leaf positions using a low resolution detector for accurate 3D dose reconstruction in IMRT

被引:1
作者
Visser, R. [1 ,2 ]
Godart, J. [1 ]
Wauben, D. J. L. [1 ]
Langendijk, J. A. [1 ]
van't Veld, A. A. [1 ]
Korevaar, E. W. [1 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Dept Radiat Oncol, Groningen, Netherlands
[2] Hanze Univ Appl Sci, Res Grp Hlth Ageing Allied Hlth Care & Nursing, Groningen, Netherlands
关键词
MLC; iterative reconstruction; 3D dose reconstruction; high resolution; 2D detector array; QUALITY-ASSURANCE; DOSIMETRY; VERIFICATION; FILM; BEAM; STEP;
D O I
10.1088/0031-9155/61/23/N642
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In pre-treatment dose verification, low resolution detector systems are unable to identify shifts of individual leafs of high resolution multi leaf collimator (MLC) systems from detected changes in the dose deposition. The goal of this study was to introduce an alternative approach (the shutter technique) combined with a previous described iterative reconstruction method to accurately reconstruct high resolution MLC leaf positions based on low resolution measurements. For the shutter technique, two additional radiotherapy treatment plans (RT-plans) were generated in addition to the original RT-plan; one with even MLC leafs closed for reconstructing uneven leaf positions and one with uneven MLC leafs closed for reconstructing even leaf positions. Reconstructed leaf positions were then implemented in the original RT-plan for 3D dose reconstruction. The shutter technique was evaluated for a 6 MV Elekta SLi linac with 5 mm MLC leafs (Agility (TM)) in combination with the MatriXX Evolution detector with detector spacing of 7.62mm. Dose reconstruction was performed with the COMPASS system (v2.0). The measurement setup allowed one row of ionization chambers to be affected by two adjacent leaf pairs. Measurements were obtained for various field sizes with MLC leaf position errors ranging from 1.0mm to 10.0mm. Furthermore, one clinical head and neck IMRT treatment beam with MLC introduced leaf position errors of 5.0mm was evaluated to illustrate the impact of the shutter technique on 3D dose reconstruction. Without the shutter technique, MLC leaf position reconstruction showed reconstruction errors up to 6.0mm. Introduction of the shutter technique allowed MLC leaf position reconstruction for the majority of leafs with sub-millimeter accuracy resulting in a reduction of dose reconstruction errors. The shutter technique in combination with the iterative reconstruction method allows high resolution MLC leaf position reconstruction using low resolution measurements with sub-millimeter accuracy.
引用
收藏
页码:N642 / N649
页数:8
相关论文
共 14 条
[1]   PATIENT-SPECIFIC 3D PRETREATMENT AND POTENTIAL 3D ONLINE DOSE VERIFICATION OF MONTE CARLO-CALCULATED IMRT PROSTATE TREATMENT PLANS [J].
Boggula, Ramesh ;
Jahnke, Lennart ;
Wertz, Hansjoerg ;
Lohr, Frank ;
Wenz, Frederik .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (04) :1168-1175
[2]   3D DVH-based metric analysis versus per-beam planar analysis in IMRT pretreatment verification [J].
Carrasco, Pablo ;
Jornet, Nuria ;
Latorre, Artur ;
Eudaldo, Teresa ;
Ruiz, Agusti ;
Ribas, Montserrat .
MEDICAL PHYSICS, 2012, 39 (08) :5040-5049
[3]   Reconstruction of high-resolution 3D dose from matrix measurements: error detection capability of the COMPASS correction kernel method [J].
Godart, J. ;
Korevaar, E. W. ;
Visser, R. ;
Wauben, D. J. L. ;
van't Veld, A. A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (15) :5029-5043
[4]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[5]   Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT [J].
Korevaar, Erik W. ;
Wauben, David J. L. ;
van der Hulst, Peter C. ;
Langendijk, Johannes A. ;
Van't Veld, Aart A. .
RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) :446-452
[6]   Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposure [J].
Menegotti, L. ;
Delana, A. ;
Martignano, A. .
MEDICAL PHYSICS, 2008, 35 (07) :3078-3085
[7]  
Olson Adam C, 2012, Int J Radiat Oncol Biol Phys, V82, pe77, DOI 10.1016/j.ijrobp.2011.01.033
[8]   Spatial resolution of 2D ionization chamber arrays for IMRT dose verification:: single-detector size and sampling step width [J].
Poppe, Bjoern ;
Djouguela, Armand ;
Blechschmidt, Arne ;
Willborn, Kay ;
Ruehmann, Antje ;
Harder, Dietrich .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (10) :2921-2935
[9]   Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy [J].
Stathakis, Sotirios ;
Myers, Pamela ;
Esquivel, Carlos ;
Mavroidis, Panayiotis ;
Papanikolaou, Nikos .
MEDICAL PHYSICS, 2013, 40 (07)
[10]   Accurate dosimetry with GafChromic™ EBT film of a 6 MV photon beam in water:: What level is achievable? [J].
van Batturn, L. J. ;
Hoffmans, D. ;
Piersma, H. ;
Heukelom, S. .
MEDICAL PHYSICS, 2008, 35 (02) :704-716