Uncertainty evaluation and propagation for spectral measurements

被引:4
作者
Schmaehling, Franko [1 ]
Wuebbeler, Gerd [1 ]
Krueger, Udo [2 ]
Ruggaber, Benjamin [2 ]
Schmidt, Franz [2 ]
Taubert, Richard D. [1 ]
Sperling, Armin [3 ]
Elster, Clemens [1 ]
机构
[1] Phys Tech Bundesanstalt, Abbestr 2-12, D-10587 Berlin, Germany
[2] TechnoTeam Bildverarbeitung GmbH, Werner von Siemens Str 5, D-98693 Ilmenau, Germany
[3] Phys Tech Bundesanstalt, Bundesallee 100, D-38116 Braunschweig, Germany
关键词
color coordinates; photometry; radiometry; spectral measurements; uncertainty; ARRAY SPECTRORADIOMETERS; LED COLOR; COLORIMETRY; CALIBRATION; VARIABLES; ERRORS;
D O I
10.1002/col.22185
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
The measurement of the spectral power distribution (SPD) of a radiation source by array spectroradiometers is a technique that is widely used. In many applications, quantities that are derived from the SPD by a weighted integral over a wavelength interval are of interest. These integral quantities ought to be accompanied by a reliable uncertainty statement, for example, to assess conformity with prescribed limits or in order to judge the consistency of results obtained at different laboratories. We have developed a generally applicable Monte Carlo procedure for evaluating the uncertainty of spectral measurements. The procedure naturally accounts for correlations in the SPD which turn out to be crucial. Means are provided to handle and transfer these large-scale correlation matrices easily. The proposed approach is illustrated by the determination of the SPD of colored LEDs from array spectroradiometer measurements, together with the derived CIE 1931 color coordinates. MATLAB(TM) software implementing the proposed analysis procedure is made available.
引用
收藏
页码:6 / 16
页数:11
相关论文
共 46 条
  • [1] [Anonymous], 2012, MATRIX COMPUTATIONS
  • [2] [Anonymous], 2008, EVALUATION MEASUREME
  • [3] BIPM IEC IFCC ISO IUPAC OIML and IUPAP, 2008, V MESA DAT S1
  • [4] Bureau International des Poids et Mesures (BIPM) International Electrotechnical Commission (IEC) International Federation of Clinical Chemistry (IFCC) International Laboratory Accreditation Cooperation (ILAC) International Organization for Standardization (ISO) International Union of Pure and Applied Chemistry (IUPAC) International Union of Pure and Applied Physics (IUPAP) and International Organization of Legal Metrology (OIML), 2011, EV MEAS DAT S2, V102
  • [5] Monte Carlo-based uncertainty analysis of UV array spectroradiometers
    Cordero, Raul R.
    Seckmeyer, Gunther
    Riechelmann, Stefan
    Damiani, Alessandro
    Labbe, Fernando
    [J]. METROLOGIA, 2012, 49 (06) : 745 - 755
  • [6] Exploitation of spectral direct UV irradiance measurements
    Cordero, Raul R.
    Seckmeyer, Gunther
    Pissulla, Darius
    Labbe, Fernando
    [J]. METROLOGIA, 2009, 46 (01) : 19 - 25
  • [7] Dubard J, 2014, P CIE EXP S MEAS UNC
  • [8] Gardner JL, 2000, COLOR RES APPL, V25, P349, DOI 10.1002/1520-6378(200010)25:5<349::AID-COL5>3.0.CO
  • [9] 2-N
  • [10] Correlated colour temperature - uncertainty and estimation
    Gardner, JL