Effects of Fluid Viscoelasticity on the Performance of an Axial Blood Pump Model

被引:2
|
作者
Hu, Qi-Hui [1 ]
Li, Jing-Yin [1 ]
Zhang, Ming-Yuan [1 ]
机构
[1] Xi An Jiao Tong Univ, Sch Energy & Power Engn, Fluid Machinery Inst, Xian 710049, Peoples R China
基金
中国国家自然科学基金;
关键词
VENTRICULAR ASSIST DEVICE; TURBULENT PIPE-FLOW; DRAG-REDUCING FLUIDS; REYNOLDS-NUMBER; PREDICTIONS; REDUCTION; STRESS; VELOCIMETRY; VISCOSITY; POLYMERS;
D O I
10.1097/MAT.0b013e3182383077
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An aqueous Xanthan gum solution (XGS) was used as blood analog fluid to explore the influence of fluid viscoelasticity on the performance of an axial blood pump model. For comparison, a 39 wt% Newtonian aqueous glycerin solution (GS), the common fluid in blood pump tests, was also used as a working fluid. The experimental results showed that a higher head curve was obtained using XGS in the pump than using GS. The heads of the XGS that were computed using the viscoelastic turbulence model agreed well with the measured data. In contrast, the standard k-epsilon turbulence model failed to provide satisfactory predictions for the XGS. The computational results revealed that in most parts of the pump model flow fields, the Reynolds shear stress values and turbulent dissipation rates of the XGS were all lower than those of the GS. The hemolysis index of the pump model using the XGS was calculated to be only one-third of that using the GS. ASAIO Journal 2012; 58:32-39.
引用
收藏
页码:32 / 39
页数:8
相关论文
共 50 条
  • [21] STRUCTURAL DESIGN AND NUMERICAL SIMULATION OF THE DIFFUSER FOR MAGLEV AXIAL BLOOD PUMP
    Wu, Huachun
    Gong, Gao
    Wang, Zhiqiang
    Hu, Yefa
    Song, Chunsheng
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2014, 14 (03)
  • [22] Short-Term In Vivo Preclinical Biocompatibility Evaluation of FW-II Axial Blood Pump in a Sheep Model
    Chen, Haibo
    Zhou, Jianye
    Sun, Hansong
    Tang, Yue
    Zhang, Yan
    Liu, Guangmao
    Hu, Shengshou
    ASAIO JOURNAL, 2011, 57 (03) : 177 - 182
  • [23] Computational Parametric Study of the Axial and Radial Clearances in a Centrifugal Rotary Blood Pump
    Rezaienia, Mohammad Amin
    Paul, Gordon
    Avital, Eldad
    Rothman, Martin
    Korakianitis, Theodosios
    ASAIO JOURNAL, 2018, 64 (05) : 643 - 650
  • [24] Measurement of hemodynamic changes with the axial flow blood pump installed in descending aorta
    Eiji Okamoto
    Tetsuya Yano
    Hidekazu Miura
    Yasuyuki Shiraishi
    Tomoyuki Yambe
    Yoshinori Mitamura
    Journal of Artificial Organs, 2017, 20 : 390 - 393
  • [25] Measurement of hemodynamic changes with the axial flow blood pump installed in descending aorta
    Okamoto, Eiji
    Yano, Tetsuya
    Miura, Hidekazu
    Shiraishi, Yasuyuki
    Yambe, Tomoyuki
    Mitamura, Yoshinori
    JOURNAL OF ARTIFICIAL ORGANS, 2017, 20 (04) : 390 - 393
  • [26] Optimising prediction model of centrifugal pump as turbine with viscosity effects
    Li, Wen-Guang
    APPLIED MATHEMATICAL MODELLING, 2017, 41 : 375 - 398
  • [27] Performance Evaluation of a Regenerative Pump under Fluid Viscosity Enhancement
    Nejadrajabali, J.
    Riasi, A.
    Nourbakhsh, S. A.
    MECHANIKA, 2017, 23 (06): : 852 - 858
  • [28] Computational and experimental evaluation of the fluid dynamics and hemocompatibility of the CentriMag blood pump
    Zhang, JT
    Gellman, B
    Koert, A
    Dasse, KA
    Gilbert, RJ
    Griffith, BP
    Wu, ZJJ
    ARTIFICIAL ORGANS, 2006, 30 (03) : 168 - 177
  • [29] Computational Fluid Dynamics Simulation of the Sputnik Pediatric Rotary Blood Pump
    Telyshev, Dmitry
    Denisov, Maxim
    Satyukova, Anna
    Le, Tatyana
    PROCEEDINGS OF 2018 IEEE EAST-WEST DESIGN & TEST SYMPOSIUM (EWDTS 2018), 2018,
  • [30] Particle Tracking Computational Prediction of Hemolysis By Blade of Micro-Axial Blood Pump
    Chen Xin
    Tan Jianping
    Yun Zhong
    MATERIALS SCIENCE AND ENGINEERING APPLICATIONS, PTS 1-3, 2011, 160-162 : 1779 - 1786