Comparison of Rapid Centrifugation Technique with Conventional Centrifugation for Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) Testing

被引:9
作者
Chandler, Ekta [1 ]
Kakkar, Naveen [1 ]
Kaur, Rupinder [1 ]
机构
[1] Christian Med Coll & Hosp, Dept Pathol, Brown Rd, Ludhiana 141008, Punjab, India
关键词
Centrifugation; Conventional; Rapid; PT; APTT; TAT; COAGULATION; PLASMA; SPEED; STAT;
D O I
10.1007/s12288-018-0983-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Prothrombin Time (PT) and activated partial thromboplastin time (APTT) are frequently performed coagulation tests in patients with coagulation disorders especially in critical care areas and in monitoring patients on anticoagulation therapy. In coagulation testing, sample processing especially centrifugation is one of the most critical steps that affect turnaround time(TAT). This study was carried out over a period of 1year. Three hundred paired samples from patients sent for PT and APTT estimation were included. One sample was centrifuged in a regular bench top centrifuge at 1500g for 20min. The other sample was divided into two polypropylene aliquots and centrifuged in a microcentrifuge at 13000g for 3min. The plasma obtained from both methods was tested for PT and APTT using the automated method on STA Compact coagulometer (Stago) using commercial thromboplastin STA(R)-Neoplastine(R) C1 Plus and phospholipid (cephalin), STA(R)-C K PRESTR 5 respectively. Data were analyzed using descriptive statistics, Student t test, correlation coefficient and Bland-Altman plots. Mean PT, INR and APTT forboth centrifugation methods was comparable with no statistically significant difference (p>0.05). PT, INR and APTT also showed good correlation (r>0.98) when compared between the two methods of centrifugation. Bland-Altman comparison between rapid and conventional methods of centrifugation for PT, INR and APTT also showed acceptable agreement. Rapid centrifugation technique for routine coagulation testing can be used safely with a significant reduction in the TAT. This can benefit patients in critical care settings and those on outpatient oral anticoagulant therapy.
引用
收藏
页码:161 / 166
页数:6
相关论文
共 18 条
  • [1] Adcock DM., 2008, CLIN LAB STANDARDS I, V5th
  • [2] [Anonymous], 2010, J BIOMED LAB SCI
  • [3] Azlin I, 2011, MED HEALTH-KUALA LUM, V6, P68
  • [4] A 2-min at 4500 g rather than a 15-min at 2200 g centrifugation does not impact the reliability of 10 critical coagulation assays
    Boissier, Elodie
    Sevin-Allouet, Mathieu
    Le Thuaut, Aurelie
    De Gaalon, Solene
    Trossaert, Marc
    Rozec, Bertrand
    Lakhal, Karim
    Rigal, Jean-Christophe
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (06) : E118 - E121
  • [5] Boudaoud L, 2006, ANN BIOL CLIN-PARIS, V64, P315
  • [6] Rapid laboratory testing for trauma patients: where a perfect result may not be in the best interests of the patient
    Callum, Jeannie L.
    Rizoli, Sandro
    Pendergrast, Jacob
    [J]. TRANSFUSION, 2010, 50 (12) : 2529 - 2532
  • [7] EVALUATION OF STAT AND ROUTINE TURNAROUND TIMES AS A COMPONENT OF LABORATORY QUALITY
    HILBORNE, LH
    OYE, RK
    MCARDLE, JE
    REPINSKI, JA
    RODGERSON, DO
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 91 (03) : 331 - 335
  • [8] PREPARATION OF PLASMA FOR COAGULATION-TESTING - EVALUATION OF THE STATSPIN HIGH-SPEED CENTRIFUGE
    HOPE, E
    MAYORGA, SR
    ROBINSON, S
    GOLDBERG, L
    LEVESON, JE
    MARENGOROWE, AJ
    PEERSCHKE, EIB
    [J]. LABORATORY MEDICINE, 1991, 22 (03) : 190 - 193
  • [9] Laffan MA, 2012, DACIE LEWIS PRACTICA, P393
  • [10] The effect of specimen hemolysis on coagulation test results
    Laga, Alvaro C.
    Cheves, Tracey A.
    Sweeney, Joseph D.
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2006, 126 (05) : 748 - 755