In-vitro performance of CAD/CAM crowns with insufficient preparation design

被引:14
作者
Rosentritt, Martin [1 ]
Preis, Verena [1 ]
Behr, Michael [1 ]
Krifka, Stephanie [1 ]
机构
[1] UKR Univ Med Ctr Regensburg, Dept Prosthet Dent, D-93042 Regensburg, Germany
关键词
CAD/CAM; Composite; Ceramic; Fitting; TCML; BOND STRENGTH; MECHANICAL-PROPERTIES; FRACTURE-RESISTANCE; SURFACE-TREATMENT; LIGHT; EXPANSION; POLYMERIZATION; TRANSMISSION; SHRINKAGE; TOUGHNESS;
D O I
10.1016/j.jmbbm.2018.10.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: To compare the debonding and fracture force of different CAD/CAM composite crowns with respect to the influence of water storage (Od vs. 90d/37 degrees C) and types of cementation (adhesive vs. self-adhesive). Methods: Extracted human molars were prepared with a worst-case preparation scenario providing a non retentive design (height similar to 4 mm; angle similar to 15 degrees) and reduced fitting (250 gm). After digitalization, 72 crowns (n = 8 per group; circular wall thickness 1.5 mm/occlusal thickness similar to 2.5 mm) were milled from the composites (CS, LU), one experimental composite (EX), a resin-infiltrated ceramic (VE), and a feldspar ceramic reference (VM). The crowns were adhesively bonded (Scotchbond Universal + Rely X Ultimate, 3M), and two groups (EX, VE) were additionally cemented with a self-adhesive cement (RelyX Unicem, 3M). After 90-d water storage, thermal cycling and mechanical loading (TCML) were performed. Restorations, which failed during storage or TCML, were analyzed using scanning electron microscopy, and surviving restorations were loaded to fracture. To evaluate storage effects, two materials (EX, LU) were investigated without water storage. Results: CS (7 x) and LU (2 x) exhibited debonding during 90-d storage. LU (5 x) debonded during TCML. Cement remained on the inner sides of the crowns in all cases. EX and VE survived storage and TCML without failure or debonding. Two specimens of VM exhibited cracks after TCML. Fracture forces varied between 720 N and 2155 N. Solely the results between VE and VM were not significantly different (p = 0.204). Debonding effects due to water storage were material dependent. Fracture forces in tendency (p > 0.117) were higher for self-adhesive cementation. Conclusions: Debonding and stability of CAD/CAM crowns were material dependent. Water storage affected debonding, and cementation marginally influenced performance and fracture force.
引用
收藏
页码:269 / 274
页数:6
相关论文
共 36 条
  • [1] Effect of Aging Regimens on Resin Nanoceramic Chairside CAD/CAM Material
    Al-Harbi, Fahad A.
    Ayad, Neveen M.
    ArRejaie, Aws S.
    Bahgat, Hala A.
    Baba, Nadim Z.
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2017, 26 (05): : 432 - 439
  • [2] Simultaneous determination of polymerization shrinkage, exotherm and thermal expansion coefficient for dental resin-composites
    Alnazzawi, A.
    Watts, D. C.
    [J]. DENTAL MATERIALS, 2012, 28 (12) : 1240 - 1249
  • [3] Optimum pressure for the high-pressure polymerization of urethane dimethacrylate
    Anh Chi Phan
    Behin, Pascal
    Stoclet, Gregory
    Ruse, N. Dorin
    Jean-Francois Nguyen
    Sadoun, Michael
    [J]. DENTAL MATERIALS, 2015, 31 (04) : 406 - 412
  • [4] Mechanical properties of resin-ceramic CAD/CAM restorative materials
    Awada, Abdallah
    Nathanson, Dan
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2015, 114 (04) : 587 - 593
  • [5] Fracture toughness of chairside CAD/CAM materials - Alternative loading approach for compact tension test
    Badawy, Rania
    El-Mowafy, Omar
    Tam, Laura E.
    [J]. DENTAL MATERIALS, 2016, 32 (07) : 847 - 852
  • [6] Chairside CAD/CAM materials. Part 1: Measurement of elastic constants and microstructural characterization
    Belli, Renan
    Wendler, Michael
    de Ligny, Dominique
    Cicconi, Maria Rita
    Petschelt, Anselm
    Peterlik, Herwig
    Lohbauer, Ulrich
    [J]. DENTAL MATERIALS, 2017, 33 (01) : 84 - 98
  • [7] Micro-shear bond strength of different resin cements to ceramic/glass-polymer CAD-CAM block materials
    Cekic-Nagas, Isil
    Ergun, Gulfem
    Egilmez, Ferhan
    Vallittu, Pekka Kalevi
    Lassila, Lippo Veli Juhana
    [J]. JOURNAL OF PROSTHODONTIC RESEARCH, 2016, 60 (04) : 265 - 273
  • [8] Machinability of CAD-CAM materials
    Chavali, Ramakiran
    Nejat, Amir H.
    Lawson, Nathaniel C.
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2017, 118 (02) : 194 - 199
  • [9] Grinding damage assessment for CAD-CAM restorative materials
    Curran, Philippe
    Cattani-Lorente, Maria
    Wiskott, H. W. Anselm
    Durual, Stephane
    Scherrer, Susanne S.
    [J]. DENTAL MATERIALS, 2017, 33 (03) : 294 - 308
  • [10] Effect of elasticity on stress distribution in CAD/CAM dental crowns: Glass ceramic vs. polymer-matrix composite
    Duan, Yuanyuan
    Griggs, Jason A.
    [J]. JOURNAL OF DENTISTRY, 2015, 43 (06) : 742 - 749