Synchrotron radiation computed tomography assessment of calcified plaques and coronary stenosis with different slice thicknesses and beam energies on 3D printed coronary models

被引:29
作者
Sun, Zhonghua [1 ]
Ng, Curtise K. C. [1 ]
Squelch, Andrew [2 ,3 ]
机构
[1] Curtin Univ, Sch Mol & Life Sci, Discipline Med Radiat Sci, GPO Box U1987, Perth, WA 6845, Australia
[2] Curtin Univ, Western Australian Sch Mines Minerals Energy & Ch, Discipline Explorat Geophys, Perth, WA, Australia
[3] Curtin Univ, Curtin Inst Computat, Computat Image Anal Grp, Perth, WA, Australia
关键词
Assessment; computed tomography (CT); coronary artery; plaque; synchrotron radiation; virtual intravascular endoscopy (VIE); 3D printed models; CT ANGIOGRAPHY; DIAGNOSTIC-ACCURACY; CALCIUM SCORE; CHEST-PAIN; ITERATIVE RECONSTRUCTION; COST-EFFECTIVENESS; ARTERY-DISEASE;
D O I
10.21037/qims.2018.09.11
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To investigate the effect of different slice thicknesses and beam energies on the visualization and assessment of coronary artery stenosis caused by calcified plaques using synchrotron radiation computed tomography (CT) based on 3D printed coronary artery models. Methods: Patient-specific 3D coronary models were created based on 3 sample coronary CT angiographic cases with calcified plaques in the left coronary arteries. In addition to the original significant coronary stenosis (>70%) shown on these CT images, stenoses of <50% and >90% were created in the segmented coronary models for simulation of different degrees of stenosis. The coronary lumen and calcification were printed with soft and rigid materials to simulate properties of coronary wall and calcified plaque, respectively. The models were scanned with synchrotron radiation CT with beam energies of 30, 40 and 50 keV and spatial resolution of 0.019x0.019x0.019 mm3 voxel size. Original high-resolution images were reconstructed with slice thicknesses of 0.095, 0.208, 0.302 and 0.491 mm to determine the effect of spatial resolution on plaque and coronary stenosis assessment based on 2D axial and 3D virtual intravascular endoscopy (VIE) images. Results: Three coronary artery models were successfully printed with plaques placed in the coronary arteries to simulate different degrees of stenosis. 2D and 3D VIE images reconstructed with slice thicknesses of 0.095, 0.208 and 0.302 mm allowed for accurate assessment of coronary plaques and lumen stenosis with no significant differences (P>0.05). Synchrotron radiation CT images reconstructed with a slice thickness of 0.491 mm resulted in overestimation of coronary stenosis when compared to other images on 2D and 3D VIE views (<50% vs. 55-72%; 70-79% vs. 80-90%) with significant differences (P<0.05). Similarly, irregular plaque appearances were observed on 2D and 3D VIE images with a slice thickness of 0.491 mm when compared to others using thin slice thicknesses. The scanning protocol with beam energy of 30 keV provided optimal visualization of coronary lumen and plaque appearances. Conclusions: This study shows the feasibility of using 3D printed coronary artery models to simulate calcifications and different degrees of coronary stenosis. High resolution synchrotron radiation CT imaging with the 30 keV beam energy enables accurate assessment of coronary stenosis in the presence of calcification, thus highlighting the importance of high spatial resolution in the diagnosis of calcified coronary plaques.
引用
收藏
页码:6 / 22
页数:17
相关论文
共 48 条
[1]   Atrial Fibrillation: Diagnostic Accuracy of Coronary CT Angiography Performed with a Whole-Heart 230-μm Spatial Resolution CT Scanner [J].
Andreini, Daniele ;
Pontone, Gianluca ;
Mushtaq, Saima ;
Conte, Edoardo ;
Perchinunno, Marco ;
Guglielmo, Marco ;
Volpato, Valentina ;
Annoni, Andrea ;
Baggiano, Andrea ;
Formenti, Alberto ;
Mancini, Maria Elisabetta ;
Beltrama, Virginia ;
Ditali, Valentina ;
Campari, Alessandro ;
Fiorentini, Cesare ;
Bartorelli, Antonio L. ;
Pepi, Mauro .
RADIOLOGY, 2017, 284 (03) :676-684
[2]   Long-term Cost-Effectiveness of Diagnostic Tests for Assessing Stable Chest Pain: Modeled Analysis of Anatomical and Functional Strategies [J].
Bertoldi, Eduardo G. ;
Stella, Steffan F. ;
Rohde, Luis E. ;
Polanczyk, Carisi A. .
CLINICAL CARDIOLOGY, 2016, 39 (05) :249-256
[3]   Comparison of synchrotron radiation angiography with conventional angiography for the diagnosis of in-stent restenosis after percutaneous transluminal coronary angioplasty [J].
Bertrand, B ;
Estève, F ;
Elleaume, H ;
Nemoz, C ;
Fiedler, S ;
Bravin, A ;
Berruyer, G ;
Brochard, T ;
Renier, M ;
Machecourt, J ;
Thomlinson, W ;
Le Bas, JF .
EUROPEAN HEART JOURNAL, 2005, 26 (13) :1284-1291
[4]   Ultra-high-resolution 3D imaging of atherosclerosis in mice with synchrotron differential phase contrast: a proof of concept study [J].
Bonanno, Gabriele ;
Coppo, Simone ;
Modregger, Peter ;
Pellegrin, Maxime ;
Stuber, Annina ;
Stampanoni, Marco ;
Mazzolai, Lucia ;
Stuber, Matthias ;
van Heeswijk, Ruud B. .
SCIENTIFIC REPORTS, 2015, 5
[5]   The effect of calcium score on the diagnostic accuracy of coronary computed tomography angiography [J].
Chen, Chien-Cheng ;
Chen, Chun-Chi ;
Hsieh, I-Chang ;
Liu, Yuan-Chang ;
Liu, Chia-Yi ;
Chan, Tiffany ;
Wen, Ming-Shien ;
Wan, Yung-Liang .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 :37-42
[6]   Coronary artery calcium scoring: Influence of adaptive statistical iterative reconstruction using 64-MDCT [J].
Gebhard, Catherine ;
Fiechter, Michael ;
Fuchs, Tobias A. ;
Ghadri, Jelena R. ;
Herzog, Bernhard A. ;
Kuhn, Felix ;
Stehli, Julia ;
Mueller, Ennio ;
Kazakauskaite, Egle ;
Gaemperli, Oliver ;
Kaufmann, Philipp A. .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 167 (06) :2932-2937
[7]   Cardiothoracic Applications of 3-dimensional Printing [J].
Giannopoulos, Andreas A. ;
Steigner, Michael L. ;
George, Elizabeth ;
Barile, Maria ;
Hunsaker, Andetta R. ;
Rybicki, Frank J. ;
Mitsouras, Dimitris .
JOURNAL OF THORACIC IMAGING, 2016, 31 (05) :253-272
[8]   Technique factors and image quality as functions of patient weight at abdominal CT [J].
Huda, W ;
Scalzetti, EM ;
Levin, G .
RADIOLOGY, 2000, 217 (02) :430-435
[9]   Artifacts at Cardiac CT: Physics and Solutions [J].
Kalisz, Kevin ;
Buethe, Ji ;
Saboo, Sachin S. ;
Abbara, Suhny ;
Halliburton, Sandra ;
Rajiah, Prabhakar .
RADIOGRAPHICS, 2016, 36 (07) :2064-2083
[10]   Three-dimensional printing in congenital heart disease: A systematic review [J].
Lau, Ivan ;
Sun, Zhonghua .
JOURNAL OF MEDICAL RADIATION SCIENCES, 2018, 65 (03) :226-236