Influence of phosphate buffer and proteins on the potentiometric response of a polymeric membrane-based solid-contact Pb(II) ion-selective electrode

被引:29
作者
Joon, Narender Kumar [1 ]
He, Ning [1 ]
Wagner, Michal [2 ]
Cardenas, Marite [3 ,4 ]
Bobacka, Johan [1 ]
Lisak, Grzegorz [1 ,5 ,6 ]
机构
[1] Abo Akad Univ, Johan Gadolin Proc Chem Ctr, Lab Analyt Chem, Biskopsgatan 8, Turku 20500, Finland
[2] Tech Univ Denmark, Dept Chem, Kemitorvet 207, DK-2800 Lyngby, Denmark
[3] Malmo Univ, Dept Biomed Sci, Albin Hanssons Vag 35, S-21432 Malmo, Sweden
[4] Malmo Univ, Biofilms Res Ctr Biointerfaces, Fac Hlth & Soc, Albin Hanssons Vag 35, S-21432 Malmo, Sweden
[5] Nanyang Technol Univ, Coll Engn, Sch Civil & Environm Engn, 50 Nanyang Ave, Singapore 639798, Singapore
[6] CleanTech, Residues & Resource Reclamat Ctr, Nanyang Environm & Water Res Inst, 1 Cleantech Loop, Singapore 637141, Singapore
基金
芬兰科学院;
关键词
Phosphate-buffered saline; Potentiometric sensors; Non-equilibrium potentiometry; Protein adsorption; DETECTION LIMIT; CONDUCTING POLYMERS; SERUM-ALBUMIN; SENSORS; ADSORPTION; BINDING; SENSITIVITY; IONOPHORES; BIOSENSOR; TRANSPORT;
D O I
10.1016/j.electacta.2017.08.126
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
In this work, the influence of phosphate buffer and proteins on the potentiometric response of a polymeric membrane-based solid-contact Pb2+-selective electrode (Pb2+-ISE) was studied. The effects of bovine serum albumin (BSA) adsorption at the surface of the ion-selective membrane combined with electrode conditioning in phosphate-buffered saline (PBS) solution was elucidated by potentiometry and electrochemical impedance spectroscopy. The adsorbed BSA at the surface of the Pb2+-ISE slightly lowered the detection limit but did not influence the selectivity of the Pb2+-ISE towards the interfering ions studied (Cu2+, Cd2+). Conditioning of the Pb2+-ISE in 0.01 mol dm(-3) PBS resulted in a super-Nernstian response which was related to fixation/extraction of Pb2+ in the ion-selective membrane via precipitation of Pb-3(PO4)(2) by PO43- anions present in PBS. By conditioning of the Pb2+-ISE in 0.01 mol dm(-3) PBS + 1 mg/ml BSA it was possible to extend the linear response range of the Pb2+-ISE towards lower analyte concentrations. The utilization of this conditioning procedure was validated by determination of Pb2+ concentrations down to ca 20 ppb in aqueous samples by Pb2+-ISEs and by comparing the results with those obtained by ICP-MS. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:490 / 497
页数:8
相关论文
共 44 条
[1]   Biomolecular immobilization on conducting polymers for biosensing applications [J].
Ahuja, Tarushee ;
Mir, Irfan Ahmad ;
Kumar, Devendra ;
Rajesh .
BIOMATERIALS, 2007, 28 (05) :791-805
[2]   Tuned ionophore-based bi-membranes for selective transport of target ions [J].
Akieh-Pirkanniemi, Marceline ;
Lisak, Grzegorz ;
Arroyo, Jesus ;
Bobacka, Johan ;
Ivaska, Ari .
JOURNAL OF MEMBRANE SCIENCE, 2016, 511 :76-83
[3]   Multi-metal binding site of serum albumin [J].
Bal, W ;
Christodoulou, J ;
Sadler, PJ ;
Tucker, A .
JOURNAL OF INORGANIC BIOCHEMISTRY, 1998, 70 (01) :33-39
[4]   Potential stability of all-solid-state ion-selective electrodes using conducting polymers as ion-to-electron transducers [J].
Bobacka, J .
ANALYTICAL CHEMISTRY, 1999, 71 (21) :4932-4937
[5]   Potentiometric ion sensors [J].
Bobacka, Johan ;
Ivaska, Ari ;
Lewenstam, Andrzej .
CHEMICAL REVIEWS, 2008, 108 (02) :329-351
[6]   Rational design of potentiometric trace level ion sensors.: A Ag+-selective electrode with a 100 ppt detection limit [J].
Ceresa, A ;
Radu, A ;
Peper, S ;
Bakker, E ;
Pretsch, E .
ANALYTICAL CHEMISTRY, 2002, 74 (16) :4027-4036
[7]   Potentiometric sensors doped with biomolecules as a new approach to small molecule/biomolecule binding kinetics analysis [J].
Daems, D. ;
De Wael, K. ;
Vissenberg, K. ;
Van Camp, G. ;
Nagels, L. .
BIOSENSORS & BIOELECTRONICS, 2014, 54 :515-520
[8]   Potentiometric sensing of nuclease activities and oxidative damage of single-stranded DNA using a polycation-sensitive membrane electrode [J].
Ding, Jiawang ;
Qin, Wei .
BIOSENSORS & BIOELECTRONICS, 2013, 47 :559-565
[9]   THROMBOGENIC PROPERTIES OF UNTREATED AND POLY(ETHYLENE OXIDE)-MODIFIED POLYMERIC MATRICES USEFUL FOR PREPARING INTRAARTERIAL ION-SELECTIVE ELECTRODES [J].
ESPADASTORRE, C ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1995, 67 (18) :3108-3114
[10]   In vivo chemical sensors: Tackling biocompatibility [J].
Frost, Megan ;
Meyerhoff, Mark E. .
ANALYTICAL CHEMISTRY, 2006, 78 (21) :7370-7377