Single-Pulse Measurement Electronics for Accurate Dosimetry in X-ray Radiation Therapy

被引:2
作者
Pettinato, Sara [1 ,2 ]
Olivieri, Riccardo [3 ]
Salvatori, Stefano [1 ,2 ]
机构
[1] Univ Niccolo Cusano, Fac Engn, Rome, Italy
[2] CNR, Ist Struttura Mat, Rome, Italy
[3] San Giovanni Addolorata Hosp, Uosd Clin Risk Management, Rome, Italy
来源
2021 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (IEEE MEMEA 2021) | 2021年
关键词
dosimetry; CVD-diamond; radiation therapy; X-ray pulses; synchronized integration; state configurable timer; GATED INTEGRATOR; RADIOTHERAPY; CANCER;
D O I
10.1109/MeMeA52024.2021.9478706
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work aims at describing a dose-measurement system based on a high-quality single-crystal synthetic diamond coupled to specifically-designed front-end electronics and timing circuitry. By exploiting the pulsed nature of radiation sourced by LINACs, the electronics performs a synchronous detection of charge-carriers generated by the diamond detector on which pulsed X-rays impinge. Synchronous detection would assure superior level of accuracy, sensitivity, and dynamics in comparison to conventional continuous-integration solutions, since the measurements are affected by environmental noise to a lesser extent. In addition, preliminary characterization described in this paper highlights the feasibility of a measurement system capable of providing information on the dose delivered at the level of a single X-ray pulse. In this context, the proposed system would be able to respond to the specific needs of accurate dose diagnostics for modern radiotherapy treatments where radiation-dose is characterized by a complex shaped distribution and steep gradients.
引用
收藏
页数:6
相关论文
共 29 条
[1]   Synthetic single crystal diamond dosimeters for Intensity Modulated Radiation Therapy applications [J].
Almaviva, S. ;
Ciancaglioni, I. ;
Consorti, R. ;
De Notaristefani, F. ;
Manfredotti, C. ;
Marinelli, Marco ;
Milani, E. ;
Petrucci, A. ;
Prestopino, G. ;
Verona, C. ;
Verona-Rinati, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (01) :191-194
[2]  
Ashour M.G., 2017, J CANC THERAPY, V8, P1160
[3]   Comparison of toxicity after IMRT and 3D-conformal radiotherapy for patients with pancreatic cancer - A systematic review [J].
Bittner, Martin-Immanuel ;
Grosu, Anca-Ligia ;
Brunner, Thomas B. .
RADIOTHERAPY AND ONCOLOGY, 2015, 114 (01) :117-121
[4]   Beam intensity modulation to reduce the field sizes for conformal irradiation of lung tumors: A dosimetric study [J].
Brugmans, MJP ;
van der Horst, A ;
Lebesque, JV ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (04) :893-904
[5]   A low-cost gated integrator boxcar averager [J].
Collier, JL ;
Goddard, BJ ;
Goode, DC ;
Marka, S ;
Telle, HH .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (09) :1204-1211
[6]   Three-dimensional graphite electrodes in CVD single crystal diamond detectors: Charge collection dependence on impinging β-particles geometry [J].
Conte, G. ;
Allegrini, P. ;
Pacilli, M. ;
Salvatori, S. ;
kononenko, T. ;
Bolshakov, A. ;
Ralchenko, V. ;
Konov, V. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 799 :10-16
[7]   X-ray diamond detectors with energy resolution [J].
Conte, G. ;
Girolami, M. ;
Salvatori, S. ;
Ralchenko, V. .
APPLIED PHYSICS LETTERS, 2007, 91 (18)
[8]  
D'Antonio E, 2018, IEEE INT SYM MED MEA, P112
[9]   Incorporating an improved dose-calculation algorithm in conformal radiotherapy of lung cancer: re-evaluation of dose in normal lung tissue [J].
De Jaeger, K ;
Hoogeman, MS ;
Engelsman, M ;
Seppenwoolde, Y ;
Damen, EMF ;
Mijnheer, BJ ;
Boersma, LJ ;
Lebesque, JV .
RADIOTHERAPY AND ONCOLOGY, 2003, 69 (01) :1-10
[10]  
Fowler J.F., 1966, Radiation Dosimetry, Vsecond