Agreement of predicted intraocular lens power using swept-source optical coherence tomography and partial coherence interferometry

被引:4
|
作者
Tana-Sanz, Pedro [1 ]
Rodriguez-Carrillo, Maria Dolores [1 ]
Ruiz-Santos, Maria [1 ]
Montes-Mico, Robert [1 ,2 ]
Ruiz-Mesa, Ramon [1 ]
Tana-Rivero, Pedro [1 ]
机构
[1] Oftalvis Clin, C Angel Lozano 11, Alicante 03001, Spain
[2] Univ Valencia, Opt & Optometry & Vis Sci, Valencia, Spain
关键词
Swept-source optical coherence tomography; Scheimpflug; partial coherence interferometry; optical biometer; intraocular lens power; PURSUING PERFECTION; BIOMETER; REPEATABILITY; DEVICE;
D O I
10.1080/17434440.2021.2008908
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose To analyze the agreement of the predicted intraocular lens (IOL) power obtained with ANTERION, IOLMaster 700 and Pentacam AXL biometers. Methods We calculated the monofocal and trifocal IOL power using the SRK/T, Haigis, Barrett Universal II and Hoffer Q formulas for 106 eyes. IOL power agreement between devices was evaluated using the Bland-Altman method. Results We found significant differences between biometers comparisons (p < 0.001). ANTERION and IOLMaster 700 did not produce significant IOL power differences (p > 0.05), with the same outcomes for medium- and long-eyes. No significant differences were found using the SRK/T, Haigis, or Hoffer Q formulas for short-eyes (p > 0.1). However, Barrett Universal II formula produced significant differences (p < 0.05) and these differences lay between the ANTERION and Pentacam AXL. ANTERION versus IOLMaster 700 comparison showed limits of agreement (LoA) varying from 1.1071D in SRK/T monofocal medium-eyes to 1.6828D in Hoffer Q trifocal all-eyes. The largest LoA (about 3.0D) was found for short-eyes when comparing the Pentacam AXL with the other two devices. Conclusions These devices provided statistically significant but clinically insignificant mean differences in predicted IOL power. However, wide LoA values suggest that for specific eyes these outcomes could be clinically significant.
引用
收藏
页码:1219 / 1234
页数:16
相关论文
共 50 条
  • [1] Predictive accuracy of partial coherence interferometry and swept-source optical coherence tomography for intraocular lens power calculation
    Whang, Woong-Joo
    Yoo, Young-Sik
    Kang, Min-Ji
    Joo, Choun-Ki
    SCIENTIFIC REPORTS, 2018, 8
  • [2] Predictive accuracy of partial coherence interferometry and swept-source optical coherence tomography for intraocular lens power calculation
    Woong-Joo Whang
    Young-Sik Yoo
    Min-Ji Kang
    Choun-Ki Joo
    Scientific Reports, 8
  • [3] Repeatability and agreement of swept-source optical coherence tomography and partial coherence interferometry biometers in myopes
    Ardoy, Elisa Garcia
    Mateos, Nuria
    Roda, Laura
    Sierra, Oscar Torrado
    Baptista, Antonio M.
    Serra, Pedro M.
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2023, 106 (07) : 783 - 792
  • [4] Biometry and intraocular power calculation using a swept-source optical coherence tomography: A repeatability and agreement study
    Hashemi, Hassan
    Miraftab, Mohammad
    Panahi, Parsa
    Asgari, Soheila
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2022, 70 (08) : 2845 - 2850
  • [5] Prediction of postoperative intraocular lens tilt using swept-source optical coherence tomography
    Hirnschall, Nino
    Buehren, Tobias
    Bajramovic, Ferid
    Trost, Michael
    Teuber, Tanja
    Findl, Oliver
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (06): : 732 - 736
  • [6] Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry
    Higashiyama, Tomoaki
    Mori, Hazuki
    Nakajima, Fumi
    Ohji, Masahito
    PLOS ONE, 2018, 13 (04):
  • [7] Evaluation of Refractive Predictive Accuracy in Intraocular Lens Power Calculations: A Comparative Study of Swept-Source Optical Coherence Tomography and Optical Low-Coherence Interferometry
    Al Barri, Leila
    Mercea, Nadina
    Ionela-Iasmina, Yasar
    Munteanu, Mihnea
    Stanca, Horia T.
    JOURNAL OF CLINICAL MEDICINE, 2025, 14 (04)
  • [8] Evaluation of crystalline lens and Chock for intraocular lens tilt using a swept-source optical coherence tomography biometer
    Wang, Li
    de Souza, Rodrigo Guimaraes
    Weikert, Mitchell P.
    Koch, Douglas D.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (01): : 35 - 40
  • [9] Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry
    Kunert, Kathleen S.
    Peter, Monika
    Blum, Marcus
    Haigis, Wolfgang
    Sekundo, Walter
    Schuetze, Juliane
    Bueehren, Tobias
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (01): : 76 - 83
  • [10] Laser Lens Size Measurement Using Swept-Source Optical Coherence Tomography
    Jia, Pingping
    Zhao, Hong
    Qin, Yuwei
    APPLIED SCIENCES-BASEL, 2020, 10 (14):