Determination of Patient Radiation Dose During Ureteroscopic Treatment of Urolithiasis Using a Validated Model

被引:45
作者
Lipkin, Michael E. [1 ]
Wang, Agnes J. [1 ]
Toncheva, Greta [2 ,4 ]
Ferrandino, Michael N. [1 ]
Yoshizumi, Terry T. [2 ,3 ,4 ]
Preminger, Glenn M. [1 ]
机构
[1] Duke Univ, Med Ctr, Div Urol, Dept Surg, Durham, NC 27710 USA
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[3] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27710 USA
[4] Duke Univ, Med Ctr, Div Radiat Safety, Durham, NC 27710 USA
关键词
ureter; nephrolithiasis; radiologic health; ureteroscopy; fluoroscopy; ATOMIC-BOMB SURVIVORS; COMPUTED-TOMOGRAPHY; CANCER-RISKS; EXPOSURE; MANAGEMENT;
D O I
10.1016/j.juro.2011.10.159
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: We measured organ specific radiation dose rates and determined effective dose rates during simulated ureteroscopy using a validated model. To calculate the effective dose, patients were exposed to ureteroscopic management of stones at our institution. Materials and Methods: A validated anthropomorphic male phantom was placed on a fluoroscopy table and underwent simulated ureteroscopy. High sensitivity metal oxide semiconductor field effect transistor dosimeters were placed at 20 organ sites in the phantom and used to measure organ specific radiation doses. These dose rates were multiplied by the appropriate tissue weighting factor and summed to calculate effective dose rates. Also, we retrospectively reviewed the charts of patients who underwent ureteroscopy at our institution. A total of 30 nonobese males with data on fluoroscopy time were included in analysis. The median effective dose was determined by multiplying median fluoroscopy time by the effective dose rate. Results: The skin entrance was exposed to the highest absorbed dose rate, followed by the small intestine (mean +/- SD 0.3286 +/- 0.0054 and 0.1882 +/- 0.0194 mGy per second, respectively). The mean effective dose rate was 0.024 +/- 0.0019 mSv per second. Median fluoroscopy time was 46.95 seconds (range 12.9 to 298.8). The median effective dose was 1.13 mSv (range 0.31 to 7.17). Conclusions: The fluoroscopy used during ureteroscopy contributes to overall radiation exposure in patients with nephrolithiasis. Nonobese males are exposed to a median of 1.13 mSv during ureteroscopy, similar to that of abdominopelvic x-ray. More data are needed to determine clinical implications but urologists must be aware and decrease patient radiation during ureteroscopy.
引用
收藏
页码:920 / 924
页数:5
相关论文
共 19 条
[1]  
[Anonymous], 2007, ANN ICRP, V37, P1, DOI 10.1016/j.icrp.2007.11.001
[2]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[3]   Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007 [J].
de Gonzalez, Amy Berrington ;
Mahesh, Mahadevappa ;
Kim, Kwang-Pyo ;
Bhargavan, Mythreyi ;
Lewis, Rebecca ;
Mettler, Fred ;
Land, Charles .
ARCHIVES OF INTERNAL MEDICINE, 2009, 169 (22) :2071-2077
[4]   Radiation Exposure in the Acute and Short-Term Management of Urolithiasis at 2 Academic Centers [J].
Ferrandino, Michael N. ;
Bagrodia, Aditya ;
Pierre, Sean A. ;
Scales, Charles D., Jr. ;
Rampersaud, Edward ;
Pearle, Margaret S. ;
Preminger, Glenn M. .
JOURNAL OF UROLOGY, 2009, 181 (02) :668-672
[5]   Comparison of a Reduced Radiation Fluoroscopy Protocol to Conventional Fluoroscopy During Uncomplicated Ureteroscopy [J].
Greene, Daniel J. ;
Tenggadjaja, Christopher F. ;
Bowman, Ryan J. ;
Agarwal, Gautum ;
Ebrahimi, Kamyar Y. ;
Baldwin, D. Duane .
UROLOGY, 2011, 78 (02) :286-290
[6]   Effective dose determination using an anthropomorphic phantom and metal oxide semiconductor field effect transistor technology for clinical adult body multidetector array computed tomography protocols [J].
Hurwitz, Lynne M. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Frush, Donald P. ;
Nguyen, Giao ;
Toncheva, Greta ;
Lowry, Carolyn .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2007, 31 (04) :544-549
[7]   Flexible Ureterorenoscopy and Holmium Laser Lithotripsy for the Management of Renal Stone Burdens That Measure 2 to 3 cm: A Multi-Institutional Experience [J].
Hyams, Elias S. ;
Munver, Ravi ;
Bird, Vincent G. ;
Uberoi, Jayant ;
Shah, Ojas .
JOURNAL OF ENDOUROLOGY, 2010, 24 (10) :1583-1588
[8]  
Krupp N, 2010, J ENDOUROL, V24, P1067, DOI 10.1089/end.2010.0040
[9]  
Lipkin ME, 2010, J ENDOUROL, V24, pA83
[10]   Effective dose: how should it be applied to medical exposures? [J].
Martin, C. J. .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (956) :639-647