Optic axis determination accuracy for fiber-based polarization-sensitive optical coherence tomography

被引:46
作者
Park, BH [1 ]
Pierce, MC [1 ]
Cense, B [1 ]
de Boer, JF [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Wellman Ctr Photomed, Boston, MA 02114 USA
关键词
D O I
10.1364/OL.30.002587
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We present a generalized analysis of fiber-based polarization-sensitive optical coherence tomography with an emphasis on determination of sample optic axis orientation. The polarization properties of a fiber-based system can cause an overall rotation in a Poincare sphere representation such that the plane of possible measured sample optic axes for linear birefringence and diattenuation no longer lies in the QU-plane. The optic axis orientation can be recovered as an angle on this rotated plane, subject to an offset and overall indeterminacy in sign such that only the magnitude, but not the direction, of a change in orientation can be determined. We discuss the accuracy of optic axis determination due to a fundamental limit on the accuracy with which a polarization state can be determined as a function of signal-to-noise ratio. (c) 2005 Optical Society of America.
引用
收藏
页码:2587 / 2589
页数:3
相关论文
共 25 条
[1]   In vivo birefringence and thickness measurements of the human retinal nerve fiber layer using polarization-sensitive optical coherence tomography [J].
Cense, B ;
Chen, HC ;
Park, BH ;
Pierce, MC ;
de Boer, JF .
JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (01) :121-125
[2]   In vivo depth-resolved birefringence measurements of the human retinal nerve fiber layer by polarization-sensitive optical coherence tomography [J].
Cense, B ;
Chen, TC ;
Park, BH ;
Pierce, MC ;
de Boer, JF .
OPTICS LETTERS, 2002, 27 (18) :1610-1612
[3]   Imaging thermally damaged tissue by polarization sensitive optical coherence tomography [J].
de Boer, JF ;
Srinivas, SM ;
Malekafzali, A ;
Chen, ZP ;
Nelson, JS .
OPTICS EXPRESS, 1998, 3 (06) :212-218
[4]   Determination of the depth-resolved Stokes parameters of light backscattered from turbid media by use of polarization-sensitive optical coherence tomography [J].
de Boer, JF ;
Milner, TE ;
Nelson, JS .
OPTICS LETTERS, 1999, 24 (05) :300-302
[5]   Review of polarization sensitive optical coherence tomography and Stokes vector determination [J].
de Boer, JF ;
Milner, TE .
JOURNAL OF BIOMEDICAL OPTICS, 2002, 7 (03) :359-371
[6]   Two-dimensional birefringence imaging in biological tissue by polarization-sensitive optical coherence tomography [J].
deBoer, JF ;
Milner, TE ;
vanGemert, MJC ;
Nelson, JS .
OPTICS LETTERS, 1997, 22 (12) :934-936
[7]   Birefringence characterization of biological tissue by use of optical coherence tomography [J].
Everett, MJ ;
Schoenenberger, K ;
Colston, BW ;
Da Silva, LB .
OPTICS LETTERS, 1998, 23 (03) :228-230
[8]   Measurement and imaging of birefringence and optic axis orientation by phase resolved polarization sensitive optical coherence tomography [J].
Hitzenberger, CK ;
Götzinger, E ;
Sticker, M ;
Pircher, M ;
Fercher, AF .
OPTICS EXPRESS, 2001, 9 (13) :780-790
[9]   Comment on "optical-fiber-based Mueller optical coherence tomography" - Reply [J].
Jiao, SL ;
Wang, LHV .
OPTICS LETTERS, 2004, 29 (24) :2875-2877
[10]   Optical-fiber-based Mueller optical coherence tomography [J].
Jiao, SL ;
Yu, WR ;
Stoica, G ;
Wang, LHV .
OPTICS LETTERS, 2003, 28 (14) :1206-1208