New examination of the performance of the MetaMax I metabolic analyser with the Douglas-bag technique

被引:13
作者
Medbo, Jon I. [1 ]
Mamen, Asgeir [2 ]
Resaland, Geir K. [2 ]
机构
[1] Natl Inst Occupat Hlth, NO-00338149 Oslo, Norway
[2] Sogn Og Fjordane Univ Coll, Fac Teacher Educ & Sport, Sogndal, Norway
关键词
Douglas-bag technique; measurement; metabolic analyser/metabolic cart; MetaMax I instrument; MetaSoft; oxygen consumption/uptake; pulmonary gas exchange; respiratory exchange ratio; GAS-ANALYSIS SYSTEMS; CARDIORESPIRATORY FITNESS; PHYSICAL-ACTIVITY; OXYGEN-UPTAKE; VALIDITY; RELIABILITY; VALIDATION; ACCURACY; VARIABILITY; B(2);
D O I
10.3109/00365513.2011.649013
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Purpose. To examine the performance of the MetaMax (R) I metabolic analyser of Cortex Biophysik, running a recent version of the software (MetaSoft v. 1.11.05), using the Douglas-bag method as reference. Experiments. First, 182 measurements of the maximal O-2 uptake on school children were analysed both by a former version of the software (MetaMax Analysis) and by a recent version (MetaSoft). In further experiments, seven grown-up subjects of different fitness levels cycled for 5 min at constant powers between 50 and 350 W while the O-2 uptake was measured simultaneously by the MetaMax I and the Douglas-bag method during the last minute of altogether 39 exercises. Results. The calculated maximal O-2 uptake of the 182 school children was on average 3% lower when the data were reanalysed by MetaSoft than when analysed by the former version of the software. There was in addition a 2% variation. In further experiments on grown-up subjects the O-2 uptake reported by the MetaMax I did not differ from that reported by the Douglas-bag system when data for all subjects were pooled, and the random error was 4%. However, the relationships differed slightly between the subjects (p = 0.005), and consequently the random error within each subject was 2 -3%. The respiratory exchange ratio (R-value) reported by the MetaMax I differed systematically from that of the control method. However, a reliable R-value could be calculated from the instrument's raw data. Conclusions. Our data suggest that the MetaMax I running MetaSoft reports unbiased values of the O-2 uptake and with moderate random error.
引用
收藏
页码:158 / 168
页数:11
相关论文
共 41 条
[1]   VARIABILITY OF RESPIRATION AND METABOLISM - RESPONSES TO SUBMAXIMAL CYCLING AND RUNNING [J].
ARMSTRONG, LE ;
COSTILL, DL .
RESEARCH QUARTERLY FOR EXERCISE AND SPORT, 1985, 56 (02) :93-96
[2]  
Astrand P-O, 2003, TXB WORK PHYSL, P276
[3]  
Ballal M A, 1982, Clin Phys Physiol Meas, V3, P57, DOI 10.1088/0143-0815/3/1/005
[4]   Validation of the portable VmaxST system for oxygen-uptake measurement [J].
Brehm, MA ;
Harlaar, J ;
Groepenhof, H .
GAIT & POSTURE, 2004, 20 (01) :67-73
[5]   Validity and reliability of three commercially available breath-by-breath respiratory systems [J].
Carter, J ;
Jeukendrup, AE .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2002, 86 (05) :435-441
[6]   Accuracy and reliability of the ParvoMedics TrueOne 2400 and MedGraphics VO2000 metabolic systems [J].
Crouter, Scott E. ;
Antczak, Amanda ;
Hudak, Jonathan R. ;
Della Valle, Diane M. ;
Haas, Jere D. .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2006, 98 (02) :139-151
[7]  
Douglas CG., 1911, J PHYSL, V42, pP17
[8]   Accuracy and reliabillity of a Cosmed K4b2 portable gas analysis system [J].
Duffield, R ;
Dawson, B ;
Pinnington, H ;
Wong, P .
JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2004, 7 (01) :11-22
[9]  
Eisenmann JC, 2003, J SPORT MED PHYS FIT, V43, P150
[10]   Physical activity, fitness and fatness: relations to mortality, morbidity and disease risk factors. A systematic review [J].
Fogelholm, M. .