The incidence of implant fractures after knee arthroplasty

被引:15
作者
Gilg, Magdalena M. [1 ]
Zeller, Christian W. [2 ]
Leitner, Lukas [1 ]
Leithner, Andreas [1 ]
Labek, Gerold [3 ]
Sadoghi, Patrick [1 ]
机构
[1] Med Univ Graz, Dept Orthoped Surg, Auenbruggerpl 5, A-8036 Graz, Austria
[2] Univ Hosp Regensburg, Dept Trauma Surg, Franz Josef Str Allee 11, D-93042 Regensburg, Germany
[3] Med Univ Innsbruck, Dept Orthoped Surg, Anichstr 35, A-6020 Innsbruck, Austria
关键词
Implant; Fracture; Breakage; Knee; Arthroplasty; TKA; TKR; Replacement; UNICOMPARTMENTAL KNEE; REVISION SURGERY; ROTATING HINGE; LIMB-SALVAGE; REPLACEMENT; FAILURE; SURVIVORSHIP; REGISTERS; PROSTHESIS; COMMON;
D O I
10.1007/s00167-016-4160-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This study aimed to evaluate the incidence of "implant" fractures after knee arthroplasty using clinical literature and worldwide arthroplasty registers. The hypothesis was that register datasets report higher rates of these rare complications than clinical studies. Calculations were based on the pooled incidence of revision operations after fractures of knee arthroplasties, comparing clinical studies published in MEDLINE-listed journals and annual reports from worldwide arthroplasty registers in a structured literature analysis based on a standardized methodology. Twelve clinical studies and datasets from six worldwide registers were included. Rates of fractures of knee arthroplasties were reported from 0.2 to 2.5 % in clinical studies versus 0.02-0.17 % in worldwide arthroplasty registers. Fractures of knee arthroplasty systems are rare complications, with clinical studies showing higher incidence rates than worldwide arthroplasty registries. Unicompartmental knee arthroplasty (UKA) implanted before 2000, constrained primary or revision constrained total knee arthroplasties, and patellar replacements showed the highest incidence of implant fracture. The results of this analysis can help clinicians to counsel patients on potential complications following knee arthroplasty. III.
引用
收藏
页码:3272 / 3279
页数:8
相关论文
共 36 条
  • [1] Fracture of the metal tibial tray after kinematic total knee replacement - A common cause of early aseptic failure
    Abernethy, PJ
    Robinson, CM
    Fowler, RM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (02): : 220 - 225
  • [2] [Anonymous], STAT REP 2011
  • [3] [Anonymous], ANN REP 2013
  • [4] [Anonymous], 14 YEAR REP
  • [5] [Anonymous], 10 ANN REP
  • [6] [Anonymous], 2013, ANN REPORT
  • [7] St. Georg sledge for medial compartment knee replacement - 461 arthroplasties followed for 4 (1-17) years
    Ansari, S
    Newman, JH
    Ackroyd, CE
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 (05): : 430 - 434
  • [8] Lateral unicompartmental knee replacement - Survivorship and clinical experience over 21
    Ashraf, T
    Newman, JH
    Evans, RL
    Ackroyd, CE
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (08): : 1126 - 1130
  • [9] The Epidemiology of Revision Total Knee Arthroplasty in the United States
    Bozic, Kevin J.
    Kurtz, Steven M.
    Lau, Edmund
    Ong, Kevin
    Chiu, Vanessa
    Vail, Thomas P.
    Rubash, Harry E.
    Berry, Daniel J.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (01) : 45 - 51
  • [10] Risk Factors for Infection Following Total Knee Arthroplasty: A Series of 3836 Cases from One Institution
    Crowe, Brooks
    Payne, Ashley
    Evangelista, Perry J.
    Stachel, Anna
    Phillips, Michael S.
    Slover, James D.
    Inneh, Ifeoma A.
    Iorio, Richard
    Bosco, Joseph A.
    [J]. JOURNAL OF ARTHROPLASTY, 2015, 30 (12) : 2275 - 2278