Quality of dynamic radionuclide renal imaging: multicentre evaluation using a functional renal phantom

被引:28
作者
Heikkinen, JO [1 ]
Kuikka, JT [1 ]
Ahonen, AKA [1 ]
Rautio, PJ [1 ]
机构
[1] Mikkeli Cent Hosp, Dept Nucl Med, Etela Savo Hosp Dist, FIN-50100 Mikkeli, Finland
关键词
renal; imaging; phantom; nuclear; medicine;
D O I
10.1097/00006231-200109000-00008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to evaluate the quality of dynamic radionuclide renal imaging in Finland. Nineteen nuclear medicine departments participated. A functional renal phantom was imaged and interpreted as a patient-like study. Reconstruction and printouts were performed according to the clinical routine of each laboratory. Three nuclear medicine specialists anonymously evaluated the quality of the image sets. The visual scores of the experts were ranked from one to five. In addition, several numerical time-activity parameters were calculated and compared between the laboratories. The average visual scores of the experts for the image sets were 3.2 +/-0.5 (range 2.4-4.2). Only two laboratories received a score value of 4 or higher. The average error for the time to reach maximum activity (T-max) ranged from -29 to +18% and for the washout time to reach half activity from maximum activity (T-1/2) ranged from -43 to +66%. These results suggest that the difference in calculated parameters between laboratories is most probably due to variations in study protocols and analysis programmes. The need for external quality assurance and quality improvement in nuclear renal imaging is evident and is recommended for regular use. ((C) 2001 Lippincott Williams & Wilkins).
引用
收藏
页码:987 / 995
页数:9
相关论文
共 22 条
[1]   CLINICAL-EXPERIENCE WITH TC-99M-MAG3, MERCAPTOACETYLTRIGLYCINE, AND A COMPARISON WITH TC-99M-DTPA [J].
ALNAHHAS, AA ;
JAFRI, RA ;
BRITTON, KE ;
SOLANKI, K ;
BOMANJI, J ;
MATHER, S ;
CARROLL, MA ;
ALJANABI, M ;
FRUSCIANTE, V ;
AJDINOWIC, B ;
FIORE, F ;
DEMENA, S ;
NIMMON, CC .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1988, 14 (9-10) :453-462
[2]   WELL-TEMPERED DIURESIS RENOGRAPHY - ITS HISTORICAL DEVELOPMENT, PHYSIOLOGICAL AND TECHNICAL PITFALLS, AND STANDARDIZED TECHNIQUE PROTOCOL [J].
CONWAY, JJ .
SEMINARS IN NUCLEAR MEDICINE, 1992, 22 (02) :74-84
[3]   TOWARDS STANDARDIZATION IN GAMMA-CAMERA RENOGRAPHY [J].
COSGRIFF, PS ;
LAWSON, RS ;
NIMMON, CC .
NUCLEAR MEDICINE COMMUNICATIONS, 1992, 13 (08) :580-585
[4]  
Fleming JS, 1999, J NUCL MED, V40, P1503
[5]   RENAL NUCLEAR-MEDICINE [J].
FOMMEI, E ;
VOLTERRANI, D .
SEMINARS IN NUCLEAR MEDICINE, 1995, 25 (02) :183-194
[6]   Quality of myocardial perfusion single-photon emission tomography imaging: multicentre evaluation with a cardiac phantom [J].
Heikkinen, J ;
Ahonen, A ;
Kuikka, JT ;
Rautio, P .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (10) :1289-1297
[7]   A FINNISH MULTICENTER QUALITY ASSURANCE PROJECT IN BONE-SCINTIGRAPHY AND BRAIN SPET - A PHANTOM STUDY [J].
HEIKKINEN, J ;
KUIKKA, JT ;
AHONEN, A ;
KURVELIN, J ;
HARTIKAINEN, K ;
KVIST, G .
NUCLEAR MEDICINE COMMUNICATIONS, 1994, 15 (10) :795-805
[8]   Quality of brain perfusion single-photon emission tomography images: multicentre evaluation using an anatomically accurate three-dimensional phantom [J].
Heikkinen, J ;
Kuikka, JT ;
Ahonen, A ;
Rautio, P .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (10) :1415-1422
[9]   A dynamic phantom for radionuclide renography [J].
Heikkinen, JO .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (03) :N39-N53
[10]   A quantitative comparison of some FADS methods in renal dynamic studies using simulated and phantom data [J].
Houston, AS ;
Sampson, WFD .