Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness

被引:124
作者
Miglior, S
Albe, E
Guareschi, M
Mandelli, G
Gomarasca, S
Orzalesi, N
机构
[1] Univ Milano Bicocca, Dipartimento Neurosci & Tecnol Biomed, Dept Ophthalmol, I-20052 Monza, MI, Italy
[2] Univ Milan, Dept Ophthalmol, Dept Med Surg & Dent, Monza, Italy
关键词
D O I
10.1136/bjo.2003.023416
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess reproducibility of central corneal thickness (CCT) measurement by means of ultrasonic pachymetry. Methods: Fifty one volunteers underwent three sessions of CCT measurements, each consisting of three CCT measurements, performed by each of three different observers. Intraand interobserver reproducibility was calculated by means of intraclass correlation coefficient (ICC). The expected range of variability between two independent evaluations was calculated using scatter plots of each test-retest difference against their mean. The standard deviation of the mean differences in the test-retest scores was used to describe the differences in the score spread. Results: The ICC ranges of the intra-and interobserver evaluations were 0.95 - 0.97 and 0.89 - 0.95 respectively; the expected variability was less than or equal to +/-1% and less than or equal to +/- 2% respectively (95% confidence interval). Conclusions: The measurement of CCT by means of ultrasonic pachymetry is highly reproducible.
引用
收藏
页码:174 / 177
页数:4
相关论文
共 44 条
[31]   MEASUREMENT OF INTRAOCULAR-PRESSURE - STUDY OF ITS REPRODUCIBILITY [J].
PHELPS, CD ;
PHELPS, GK .
ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE, 1976, 198 (01) :39-43
[32]  
SALZ JJ, 1983, OPHTHALMIC SURG LAS, V14, P750
[33]   Assessment of the diurnal variation in central corneal thickness and intraocular pressure for patients with suspected glaucoma [J].
Shah, S ;
Spedding, C ;
Bhojwani, R ;
Kwartz, J ;
Henson, D ;
McLeod, D .
OPHTHALMOLOGY, 2000, 107 (06) :1191-1193
[34]   Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic [J].
Shah, S ;
Chatterjee, A ;
Mathai, M ;
Kelly, SP ;
Kwartz, J ;
Henson, D ;
McLeod, D .
OPHTHALMOLOGY, 1999, 106 (11) :2154-2160
[35]  
SNEDECOR GW, 1989, STAT METHODS, P242
[36]   Applanation tonometry and correction according to corneal thickness [J].
Stodtmeister, R .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 1998, 76 (03) :319-324
[37]   SYSTEMATIC-ERROR IN COMPUTERIZED PACHYMETRY [J].
STUCCHI, CA ;
GENNARI, G ;
AIMINO, G ;
DICARLO, I ;
BAUCHIERO, L .
OPHTHALMOLOGICA, 1993, 207 (04) :208-214
[38]  
SUDESH S, 1993, ACTA OPHTHALMOL, V71, P185
[39]  
Van Veldhuisen PC, 2000, AM J OPHTHALMOL, V130, P429
[40]  
VILLASENOR RA, 1986, OPHTHALMOLOGY, V93, P327