Highly sensitive single polyaniline nanowire biosensor for the detection of immunoglobulin G and myoglobin

被引:85
作者
Lee, Innam [1 ]
Luo, Xiliang [2 ]
Cui, Xinyan Tracy [2 ]
Yun, Minhee [1 ,2 ]
机构
[1] Univ Pittsburgh, Dept Elect & Comp Engn, Pittsburgh, PA 15261 USA
[2] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA 15260 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
Polyaniline nanowire; Biosensing; Immunoglobulin G; Myoglobin; Surface immobilization; LABEL-FREE DETECTION; ACUTE MYOCARDIAL-INFARCTION; ELECTRICAL DETECTION; PROTEIN INTERACTIONS; CONDUCTING POLYMERS; SENSOR ARRAYS; ANTIBODY; SERUM; NANOPARTICLES; DNA;
D O I
10.1016/j.bios.2011.01.001
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A single polyaniline (PANI) nanowire-based biosensor was established to detect immunoglobulin G (IgG) and myoglobin (Myo), which is one of the cardiac biomarkers. The single PANI nanowires were fabricated via an electrochemical growth method, in which single nanowires were formed between a pair of patterned electrodes. The single PANI nanowires were functionalized with monoclonal antibodies (mAbs) of IgG or Myo via a surface immobilization method, using 1-ethyl-3-(3-dimethyaminopropyl) carbodiimide (EDC), and N-hydroxysuccinimde (NHS). The functionalization was then verified by Raman spectroscopy and fluorescence microscopy. The target proteins of IgG and Myo were detected by measuring the conductance change of functionalized single PANI nanowires owing to the capturing of target proteins by mAbs. The detection limit was found to be 3 ng/mL for IgG and 1.4 ng/mL for Myo. No response was observed when single nanowires were exposed to a non-specific protein, demonstrating excellent specificity to expected target detection. Together with the fast response time (a few seconds), high sensitivity, and good specificity, this single PANI nanowire-based biosensor shows great promise in the detection of cardiac markers and other proteins. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:3297 / 3302
页数:6
相关论文
共 33 条
[1]   Polymers in sensor applications [J].
Adhikari, B ;
Majumdar, S .
PROGRESS IN POLYMER SCIENCE, 2004, 29 (07) :699-766
[2]   Biomolecular immobilization on conducting polymers for biosensing applications [J].
Ahuja, Tarushee ;
Mir, Irfan Ahmad ;
Kumar, Devendra ;
Rajesh .
BIOMATERIALS, 2007, 28 (05) :791-805
[3]   Label-free protein assay based on a nanomechanical cantilever array [J].
Arntz, Y ;
Seelig, JD ;
Lang, HP ;
Zhang, J ;
Hunziker, P ;
Ramseyer, JP ;
Meyer, E ;
Hegner, M ;
Gerber, C .
NANOTECHNOLOGY, 2003, 14 (01) :86-90
[4]   Label-Free Electrical Detection of Cardiac Biomarker with Complementary Metal-Oxide Semiconductor-Compatible Silicon Nanowire Sensor Arrays [J].
Chua, Jay Huiyi ;
Chee, Ru-Ern ;
Agarwal, Ajay ;
Wong, She Mein ;
Zhang, Guo-Jun .
ANALYTICAL CHEMISTRY, 2009, 81 (15) :6266-6271
[5]   SURFACE-ENHANCED RESONANCE RAMAN-SCATTERING FROM CYTOCHROME-C AND MYOGLOBIN ADSORBED ON A SILVER ELECTRODE [J].
COTTON, TM ;
SCHULTZ, SG ;
VANDUYNE, RP .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1980, 102 (27) :7960-7962
[6]   On-chip detection of myoglobin based on fluorescence [J].
Darain, Farzana ;
Yager, Paul ;
Gan, Kai Ling ;
Tjin, Swee Chuan .
BIOSENSORS & BIOELECTRONICS, 2009, 24 (06) :1744-1750
[7]   VALUE OF MYOGLOBIN, TROPONIN-T, AND CK-MB(MASS) IN RULING OUT AN ACUTE MYOCARDIAL-INFARCTION IN THE EMERGENCY ROOM [J].
DEWINTER, RJ ;
KOSTER, RW ;
STURK, A ;
SANDERS, GT .
CIRCULATION, 1995, 92 (12) :3401-3407
[8]   An in situ electrochemical surface plasmon resonance immunosensor with polypyrrole propylic acid film:: Comparison between SPR and electrochemical responses from polymer formation to protein immunosensing [J].
Dong, Hua ;
Cao, Xiaodong ;
Li, Chang Ming ;
Hu, Weihua .
BIOSENSORS & BIOELECTRONICS, 2008, 23 (07) :1055-1062
[9]   Silicon nanowire arrays for label-free detection of DNA [J].
Gao, Zhiqiang ;
Agarwal, Ajay ;
Trigg, Alastair D. ;
Singh, Navab ;
Fang, Cheng ;
Tung, Chih-Hang ;
Fan, Yi ;
Buddharaju, Kavitha D. ;
Kong, Jinming .
ANALYTICAL CHEMISTRY, 2007, 79 (09) :3291-3297
[10]   Application of conducting polymers to biosensors [J].
Gerard, M ;
Chaubey, A ;
Malhotra, BD .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (05) :345-359