The Influence of the Implant Macrogeometry on Insertion Torque, Removal Torque, and Periotest Implant Primary Stability: A Mechanical Simulation on High-Density Artificial Bone

被引:7
作者
Tumedei, Margherita [1 ]
Petrini, Morena [1 ]
Pietropaoli, Davide [2 ]
Cipollina, Alessandro [3 ]
La Torre, Castrenze
Di Carmine, Maria Stella [4 ]
Piattelli, Adriano [1 ,5 ,6 ,7 ]
Iezzi, Giovanna [1 ]
机构
[1] Univ G dAnnunzio, Dept Med Oral & Biotechnol Sci, Via Vestini 31, I-66100 Chieti, Italy
[2] Univ Aquila, Oral Dis & Syst Interact Study Grp ODISSY Grp, Dept Life Hlth & Environm Sci,San Salvatore Hosp, Dent Clin,Unit Oral Dis Prevent & Translat Res, I-67100 Laquila, Italy
[3] Via Piacenza 7, I-92019 Sciacca, Italy
[4] Via Fornace 10, I-65014 Loreto Aprutino, Italy
[5] Catholic Univ San Antonio de Murcia UCAM, Biomat Engn, Ave Jeronimos 135, Murcia 30107, Spain
[6] Villa Serena Fdn Res, Via Leonardo Petruzzi 42, I-65013 Citta S Angelo, Italy
[7] Casa Cura Villa Serena Dott L Petruzzi, Via Leonardo Petruzzi 42, I-65013 Citta S Angelo, Italy
来源
SYMMETRY-BASEL | 2021年 / 13卷 / 05期
关键词
dental implant; primary stability; insertion torque; mechanical behaviour; drilling protocol; BEAM COMPUTED-TOMOGRAPHY; SURGICAL TECHNIQUE; TAPERED IMPLANTS; DENTAL IMPLANTS; THREAD; DIAMETER; NECK;
D O I
10.3390/sym13050776
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: The primary stability is a determinant clinical condition for the success of different dental implants macro-design in different bone density using a validated and repeatable in vitro technique employing solid rigid polyurethane blocks. Materials and Methods: Five implants 3.8 x 13 mm(2) for each macro-design (i.e., IK-tapered; IC-cylindric; and IA-active blade shape) were positioned into 20- and 30- pounds per cubic foot (PCF) polyurethane blocks. Bucco-lingual (BL) and mesial-distal (MD) implant stability quotient score (ISQ) was assessed by resonance frequency analysis while, insertion/removal torques were evaluated by dynamometric ratchet. Results: IC implants shown better primary stability in terms of ISQ compared to IA and IK in lower density block (20 PCF), while IK was superior to IA in higher density (30 PCF). IC shown higher removal torque in 30-PCF compared to IA and IC. Conclusions: The study effectiveness on polyurethane artificial bone with isotropic symmetry structure showed that the implants macro-design might represent a key factor on primary stability, in particular on low-density alveolar bone. Clinicians should consider patients features and implant geometry during low-density jaws rehabilitation. Further investigations are needed to generalize these findings.
引用
收藏
页数:11
相关论文
共 37 条
  • [1] The effect of thread pattern upon implant osseointegration
    Abuhussein, Heba
    Pagni, Giorgio
    Rebaudi, Alberto
    Wang, Hom-Lay
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (02) : 129 - 136
  • [2] Alves CC, 2009, INT J PERIODONT REST, V29, P161
  • [3] Effect of microthread design of dental implants on stress and strain patterns: a three-dimensional finite element analysis
    Amid, Reza
    Raoofi, Saeed
    Kadkhodazadeh, Mahdi
    Movahhedi, Mohammad Reza
    Khademi, Maryam
    [J]. BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2013, 58 (05): : 457 - 467
  • [4] Stability of tapered and parallel-walled dental implants: A systematic review and meta-analysis
    Atieh, Momen A.
    Alsabeeha, Nabeel
    Duncan, Warwick J.
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (04) : 634 - 645
  • [5] Influence of surgical technique, implant shape and diameter on the primary stability in cancellous bone
    Bilhan, H.
    Geckili, O.
    Mumcu, E.
    Bozdag, E.
    Sunbuloglu, E.
    Kutay, O.
    [J]. JOURNAL OF ORAL REHABILITATION, 2010, 37 (12) : 900 - 907
  • [6] Chang Pei-Kun, 2012, Int J Oral Maxillofac Implants, V27, pe96
  • [7] Primary Stability of Dental Implants in Low-Density (10 and 20 pcf) Polyurethane Foam Blocks: Conical vs Cylindrical Implants
    Comuzzi, Luca
    Tumedei, Margherita
    Pontes, Ana Emilia
    Piattelli, Adriano
    Iezzi, Giovanna
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2020, 17 (08)
  • [8] Degidi Marco, 2006, J Oral Implantol, V32, P277, DOI 10.1563/0.795.1
  • [9] Implant insertion torque value in immediate loading: A retrospective study
    Del Giudice, Roberto
    Piattelli, Adriano
    Grande, Nicola-Maria
    Cataneo, Enrico
    Crispino, Antonio
    Petrini, Morena
    [J]. MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2019, 24 (03): : E398 - E403
  • [10] Effect of Diameter and Length on Stress Distribution of the Alveolar Crest around Immediate Loading Implants
    Ding, Xi
    Liao, Sheng-Hui
    Zhu, Xing-Hao
    Zhang, Xiu-Hua
    Zhang, Lin
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2009, 11 (04) : 279 - 287