The between-day and inter-rater reliability of a novel wireless system to analyse lumbar spine posture

被引:23
|
作者
O'Sullivan, Kieran [1 ]
Galeotti, Luciana [1 ,2 ]
Dankaerts, Wim [2 ,3 ]
O'Sullivan, Leonard [1 ]
O'Sullivan, Peter [4 ]
机构
[1] Univ Limerick, Limerick, Ireland
[2] Catholic Univ, Louvain, Belgium
[3] Univ Coll Limburg, Hasselt, Belgium
[4] Curtin Univ Technol, Perth, WA, Australia
关键词
back pain; posture; reliability; LOW-BACK-PAIN; TRUNK MUSCLE ACTIVATION; SITTING POSTURES; 3-DIMENSIONAL MEASUREMENT; HEALTHY CONTROLS; MOVEMENT; CLASSIFICATION; CURVATURE; MOTION; IMPAIRMENT;
D O I
10.1080/00140139.2010.535020
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Lumbar posture is commonly assessed in non-specific chronic low back pain (NSCLBP), although quantitative measures have mostly been limited to laboratory environments. The BodyGuard is a spinal position monitoring device that can monitor posture in real time, both inside and outside the laboratory. The reliability of this wireless device was examined in 18 healthy participants during usual sitting and forward bending, two tasks that are commonly provocative in NSCLBP. Reliability was determined using intraclass correlation coefficients (ICC), the standard error of measurement (SEM), the mean difference and the minimal detectable change (MDC90). Between-day ICC values ranged from 0.84 to 0.87, with small SEM (5%), mean difference (9%) and MDC90 (14%) values. Inter-rater ICC values ranged from 0.91 to 0.94, with small SEM (4%), mean difference (6%) and MDC90 (9%) values. Between-day and inter-rater reliability are essential requirements for clinical utility and were excellent in this study. Further studies into the validity of this device and its application in clinical trials in occupational settings are required. Statement of Relevance:A novel device that can analyse spinal posture exposure in occupational settings in a minimally invasive manner has been developed. This study established that the device has excellent between-day and inter-rater reliability in healthy pain-free subjects. Further studies in people with low back pain are planned.
引用
收藏
页码:82 / 90
页数:9
相关论文
共 50 条
  • [21] INTRA-RATER AND INTER-RATER RELIABILITY OF THE FIVE IMAGE-BASED CRITERIA OF THE FOOT POSTURE INDEX-6
    Terada, Masafumi
    Wittwer, Ara M.
    Gribble, Phillip A.
    INTERNATIONAL JOURNAL OF SPORTS PHYSICAL THERAPY, 2014, 9 (02): : 187 - 194
  • [22] Quantitative measurements of forward head in college-aged students: A conformational study of intra-rater and inter-rater reliability of a novel Posture Measuring Device
    Gray, Kurt
    Dalal, Rajiv
    Weaver, Jullet Davis
    Randolph, Saungaylia
    JOURNAL OF BODYWORK AND MOVEMENT THERAPIES, 2021, 26 : 233 - 237
  • [23] Assessment of the Human Factors Analysis and Classification System (HFACS): Intra-rater and inter-rater reliability
    Ergai, Awatef
    Cohen, Tara
    Sharp, Julia
    Wiegmann, Doug
    Gramopadhye, Anand
    Shappell, Scott
    SAFETY SCIENCE, 2016, 82 : 393 - 398
  • [24] The inter-rater reliability of clinical tests that best predict the subclassification of lumbar segmental instability: structural, functional and combined instability
    Alyazedi, Faisal M.
    Lohman, Everett B.
    Swen, R. Wesley
    Bahjri, Khaled
    JOURNAL OF MANUAL & MANIPULATIVE THERAPY, 2015, 23 (04) : 197 - 204
  • [25] Evaluation of the within- and between-day intra-tester and inter-tester reliability of positioning subjects into neutral and lordotic sitting postures
    Korakakis, Vasileios
    O'Sullivan, Kieran
    Sotiralis, Yiannis
    Karanasios, Stefanos
    Sideris, Vasilis
    Sideris, Alexandros
    Sakellariou, Konstantinos
    Giakas, Giannis
    EUROPEAN JOURNAL OF PHYSIOTHERAPY, 2019, 21 (03) : 142 - 152
  • [26] Inter-rater reliability of output measures for a posture matching assessment approach: a pilot study with food service workers
    Cann, A. P.
    Connolly, M.
    Ruuska, R.
    MacNeil, M.
    Birmingham, T. B.
    Vandervoort, A. A.
    Callaghan, J. P.
    ERGONOMICS, 2008, 51 (04) : 556 - 572
  • [27] Intra-rater and inter-rater reliability of Surgimap Spine software for measuring spinal postural angles from digital photographs
    Nesma A. Helmya
    Mohsen M. El-Sayyadb
    Omaima M. Kattabeib
    Bulletin of Faculty of Physical Therapy, 2015, 20 (2) : 193 - 199
  • [28] On the reliability of peer-reviewing A study of inter-rater reliability between reviewers for DGPuK conference submissions
    Koch, Thomas
    Geiss, Stefan
    SCM STUDIES IN COMMUNICATION AND MEDIA, 2019, 8 (02): : 203 - 235
  • [29] Reliability of ultrasound assessment of the rectus femoris muscle thickness: intra-rater, inter-rater, and inter-day analysis accounting for age and sex
    Stausholm, Martin Bjorn
    da Silva, Katielle Rodrigues
    Inacio, Pedro Augusto
    de Sa Filho, Alberto Souza
    Lopes-Martins, Pedro Sardinha Leonardo
    Bjordal, Jan Magnus
    Leonardo, Patricia Sardinha
    Lopes-Martins, Rodrigo Alvaro B.
    BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)
  • [30] Intra-rater and inter-rater reliability of a handheld myotonometer measuring myofascial stiffness of lower lumbar myofascial tissue in healthy adults
    Valenti, Fabio
    Meden, Sara
    Frange, Maja
    Vauhnik, Renata
    PEERJ, 2024, 12