Microfabricated Gas Chromatograph for Sub-ppb Determinations of TCE in Vapor Intrusion Investigations

被引:12
作者
Chang, Hungwei [2 ]
Kim, Sun Kyu [2 ]
Sukaew, Thitiporn [2 ]
Bohrer, Forest [2 ]
Zellers, Edward T. [1 ]
机构
[1] Univ Michigan, Ctr Wireless Integrated Microsyst WIMS, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Environm Hlth Sci & Chem, Ann Arbor, MI 48109 USA
来源
EUROSENSORS XXIV CONFERENCE | 2010年 / 5卷
关键词
Micro Gas Chromatograph; MEMS; GC; TCE; Vapor intrusion; preconcentrator; sensor array; nanoparticle;
D O I
10.1016/j.proeng.2010.09.271
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A microfabricated gas chromatograph (mu GC) is described and its application to the analysis of sub-parts-per-billion (ppb) concentrations of trichloroethylene (TCE) in mixtures, relevant to the problem of TCE vapor intrusion (VI) into homes and offices, is demonstrated. The system employs a MEMS focuser, dual MEMS separation columns, and MEMS interconnects along with a microsensor array. These are interfaced to a (non-MEMS) front-end pre-trap and high-volume sampler module to reduce analysis time. The response patterns generated from the sensor array for each vapor are combined with the chromatographic retention time to identify and differentiate the components of VOC mixtures. All functions are controlled by a LabView routine written in house. A chemometric method based on multivariate curve resolution has also been developed for analyzing partially resolved mixture components. First results are presented of the capture, separation, recognition, and quantification of TCE in a mixture. TCE is measured at 0.185 ppb, with a projected detection limit of 0.030 ppb (20-L sample). (C) 2010 Published by Elsevier Ltd.
引用
收藏
页码:973 / 976
页数:4
相关论文
共 10 条
  • [1] Evaluation of multitransducer arrays for the determination of organic vapor mixtures
    Jin, Chunguang
    Kurzawski, Petra
    Hierlemann, Andreas
    Zellers, Edward T.
    [J]. ANALYTICAL CHEMISTRY, 2008, 80 (01) : 227 - 236
  • [2] Chemometric analysis of gas chromatographic peaks measured with a microsensor array: Methodology and performance assessment
    Jin, Chunguang
    Zellers, Edward T.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2009, 139 (02) : 548 - 556
  • [3] HEURISTIC MODEL FOR PREDICTING THE INTRUSION RATE OF CONTAMINANT VAPORS INTO BUILDINGS
    JOHNSON, PC
    ETTINGER, RA
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1991, 25 (08) : 1445 - 1452
  • [4] Kim S. K., 2009, 15th International Conference on Solid-State Sensors, Actuators and Microsystems. Transducers 2009, P128, DOI 10.1109/SENSOR.2009.5285546
  • [5] Recent advancements in the gas-phase MicroChemLab
    Lewis, Patrick R.
    Manginell, Ronald P.
    Adkins, Douglas R.
    Kottenstette, Richard J.
    Wheeler, David R.
    Sokolowski, Sara S.
    Trudell, Daniel E.
    Byrnes, Joy E.
    Okandan, Murat
    Bauer, Joseph M.
    Manley, Robert G.
    Frye-Mason, Gregory C.
    [J]. IEEE SENSORS JOURNAL, 2006, 6 (03) : 784 - 795
  • [6] First-generation hybrid MEMS gas chromatograph
    Lu, CJ
    Steinecker, WH
    Tian, WC
    Oborny, MC
    Nichols, JM
    Agah, M
    Potkay, JA
    Chan, HKL
    Driscoll, J
    Sacks, RD
    Wise, KD
    Pang, SW
    Zellers, ET
    [J]. LAB ON A CHIP, 2005, 5 (10) : 1123 - 1131
  • [7] Serrano G., 2009, 15th International Conference on Solid-State Sensors, Actuators and Microsystems. Transducers 2009, P1654, DOI 10.1109/SENSOR.2009.5285748
  • [8] Assessing the reliability of wall-coated microfabricated gas chromatographic separation columns
    Serrano, Gustavo
    Reidy, Shaelah M.
    Zellers, Edward T.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2009, 141 (01): : 217 - 226
  • [9] Real-time monitoring of sub-ppb concentrations of aromatic volatiles with a MEMS-enabled miniaturized gas-chromatograph
    Zampolli, S.
    Elmi, I.
    Mancarella, F.
    Betti, P.
    Dalcanale, E.
    Cardinali, G. C.
    Severi, M.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2009, 141 (01) : 322 - 328
  • [10] Zellers Edward T., 2007, TRANSDUCERS '07 & Eurosensors XXI. 2007 14th International Conference on Solid-State Sensors, Actuators and Microsystems, P1491, DOI 10.1109/SENSOR.2007.4300427