THE INFLUENCE OF BONE-DENSITY ON IN VIVO K X-RAY FLUORESCENCE BONE-LEAD MEASUREMENTS

被引:2
作者
Lodwick, Camille J. [1 ]
Lodwick, Jeffrey C. [2 ]
Spitz, Henry B. [3 ]
机构
[1] Oregon State Univ, Dept Nucl Engn & Radiat Hlth Phys, Corvallis, OR 97331 USA
[2] Occupat Safety & Hlth Adm, US Dept Labor, Salt Lake City, UT 84070 USA
[3] Univ Cincinnati, Dept Mech Ind & Nucl Engn, Cincinnati, OH 45221 USA
来源
HEALTH PHYSICS | 2011年 / 100卷 / 05期
关键词
bones; human; Monte Carlo; Pb-210; x rays; CALIBRATION;
D O I
10.1097/HP.0b013e3181f725af
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Mathematical simulations and benchmark measurements were performed to assess the impact that normal variations in human calcium content have on in vivo K x-ray fluorescence measurements of lead in bone. Four sets of cortical bone tissue simulants were fabricated containing from 20.8% to 23.8% calcium ( by weight) for measurement in a surrogate ( phantom) of the human leg. The net counts detected in the coherent backscatter peak at 88.034 keV using a Cd-109 source indicate a positive trend, with a variability of up to 17% over the range of assessed calcium content. Mathematical simulations confirm this trend and also demonstrate that the contribution of 87.3 keV Pb K-beta 2 counts, which are unresolved in measurements, do not contribute significantly to the coherent peak at low levels of bone-lead content. Both measurements and simulations confirm that calcium is a statistically significant parameter in predicting the K-XRF response and suggest that lead levels may be over-predicted for individuals having low bone density compared to the calibration matrix. Simulations identify a 4.5% negative bias in measured lead values for each 1% increase in calcium weight percent in the bone matrix as compared to the calibration matrix. It is therefore important to accommodate this uncertainty when performing epidemiological studies of populations having a wide range of bone densities. Health Phys. 100(5):502-507; 2011
引用
收藏
页码:502 / 507
页数:6
相关论文
共 50 条
  • [21] Calibration of the 125I-induced x-ray fluorescence spectrometry-based system of in vivo bone strontium determinations using hydroxyapatite as a phantom material: a simulation study
    Da Silva, Eric
    Pejovic-Milic, Ana
    PHYSIOLOGICAL MEASUREMENT, 2017, 38 (06) : 1077 - 1093
  • [22] Investigation of X-ray Fluorescence Computed Tomography (XFCT) and K-Edge Imaging
    Bazalova, Magdalena
    Kuang, Yu
    Pratx, Guillem
    Xing, Lei
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (08) : 1620 - 1627
  • [23] The influence of X-ray radiation on the mineral/organic matrix interaction of bone tissue:: An FT-IR microscopic investigation
    Hübner, W
    Blume, A
    Pushjakova, R
    Dekhtyar, Y
    Hein, HJ
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2005, 28 (01) : 66 - 73
  • [24] ASSESSMENT OF THE MINERAL DENSITY AND MINERAL CONTENT OF THE EQUINE THIRD METACARPAL AND FIRST PHALANX BONE BY DUAL ENERGY X-RAY ABSORPTIOMETRY
    Toth, Peter
    Horvath, Csaba
    Ferencz, Viktoria
    Nagy, Krisztina
    Gligor, Noemi
    Szenci, Otto
    Bodo, Gabor
    ACTA VETERINARIA HUNGARICA, 2010, 58 (03) : 317 - 329
  • [25] Effectiveness of an X-ray shielding sheet with lower shielding ability to enable both bone mineral density determination and morphological diagnosis
    Maeda, Tatsuya
    Hayashi, Hiroaki
    Ando, Miku
    Nishigami, Rina
    Kobayashi, Daiki
    Asahara, Takashi
    Goto, Sota
    Kimoto, Natsumi
    Takegami, Kazuki
    Yamashita, Kazuta
    Higashino, Kosaku
    Murakami, Shuichi
    Konishi, Takeshi
    Maki, Motochika
    RADIATION MEASUREMENTS, 2024, 176
  • [26] Precision lattice spacing measurements using X-ray Cu K alpha doublet
    Fukumori, T
    Imai, K
    Hasegawa, T
    Akashi, Y
    JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 1997, 66 (07) : 1976 - 1978
  • [27] The Influence of Spectral Interferences on Critical Element Determination with Portable X-Ray Fluorescence (pXRF)
    Gallhofer, Daniela
    Lottermoser, Bernd G.
    MINERALS, 2018, 8 (08):
  • [28] Detection of Hand Bone Fractures in X-Ray Images Using Hybrid YOLO NAS
    Medaramatla, Sai Charan
    Samhitha, Chennupati Veda
    Pande, Sagar Dhanraj
    Vinta, Surendra Reddy
    IEEE ACCESS, 2024, 12 : 57661 - 57673
  • [29] Computer-assisted quantification of periaxial bone rotation from X-ray CT
    Tsao, J
    Chiodo, CP
    Williamson, DS
    Wilson, MG
    Kikinis, R
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1998, 22 (04) : 615 - 620
  • [30] Dual-Energy X-Ray Absorptiometry for Measurement of Phalangeal Bone Mineral Density on a Slot-Scanning Digital Radiography System
    Dendere, Ronald
    Potgieter, J. Herman
    Steiner, Stefan
    Whiley, Sarah P.
    Douglas, Tania S.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (12) : 2850 - 2859