Does the Novel Thulium Fiber Laser Have a Higher Risk of Urothelial Thermal Injury than the Conventional Holmium Laser in an In Vitro Study?

被引:35
作者
Belle, Joshua D. [1 ]
Chen, Ricky [1 ]
Srikureja, Nathaniel [1 ]
Amasyali, Akin S. [1 ]
Keheila, Mohammed [1 ]
Baldwin, D. Duane [1 ,2 ]
机构
[1] Loma Linda Univ Med Ctr, Dept Urol, Loma Linda, CA USA
[2] Loma Linda Univ Hlth, Dept Urol, Room A560,11234 Anderson St, Loma Linda, CA 92354 USA
关键词
thulium; Soltive; laser; ureteroscopy; stone; holmium; PERCUTANEOUS NEPHROLITHOTOMY; LITHOTRIPSY; TEMPERATURE; SETTINGS;
D O I
10.1089/end.2021.0842
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction and Objective: The novel thulium fiber laser (TFL) has been shown to break stones more rapidly than the holmium:YAG laser (HL). However, some evidence suggests that the TFL generates more heat. The purpose of this study is to compare ureteral temperatures generated by these lasers during ureteroscopic laser lithotripsy in a benchtop model.Methods: A 1-cm BegoStone was manually impacted in the proximal ureter of a three-dimensional printed kidney-ureter model and submerged in 35.5 degrees C saline. Lithotripsy was performed using a 7.6F flexible ureteroscope and a 200 mu m laser fiber without a ureteral access sheath. The Dornier 30 W HL, Olympus 100 W HL, and Olympus 60 W TFL were compared. A needle thermocouple to measure temperature was inserted 2 mm from the laser tip. Irrigation was maintained at 35 cc/minute at room temperature using the Thermedx FluidSmart System. Intraluminal temperature was continuously recorded for 60 seconds of laser activation. Five trials were performed for each of four different power settings: 3.6, 10, 20, and 30 W. Analysis of variance and Mann-Whitney U tests were performed with p < 0.05 considered significant.Results: Intraureteral fluid temperature increased as laser power settings increased for all lasers (p < 0.05). The TFL generated higher average ureteral fluid temperatures than the Dornier and Empower HL at all power settings tested (p < 0.001). The maximum temperature for the TFL was higher than the Dornier and Empower HL at all power settings tested (p < 0.001), except at 20 W with the Empower HL. At 30 W, the TFL exceeded 43 degrees C, the threshold for tissue damage.Conclusions: The TFL generated more heat at all settings tested. Supraphysiologic ureteral temperatures may be generated with extended use at high energy settings and low irrigation rates. Understanding the heat generation properties of both lasers could help improve the safety of ureteroscopic laser lithotripsy.
引用
收藏
页码:1249 / 1254
页数:6
相关论文
共 24 条
  • [1] Renal Pelvic Pressure in Percutaneous Nephrolithotomy: The Effect of Multiple Tracts
    Abourbih, Samuel
    Alsyouf, Muhannad
    Yeo, Alexander
    Martin, Jacob
    Vassantachart, Janna M.
    Lee, Michael
    Maldonado, Jonathan
    Lightfoot, Michelle
    Li, Roger
    Olgin, Gaudencio
    Baldwin, D. Duane
    [J]. JOURNAL OF ENDOUROLOGY, 2017, 31 (10) : 1079 - 1083
  • [2] Caliceal Fluid Temperature During High-Power Holmium Laser Lithotripsy in an In Vivo Porcine Model
    Aldoukhi, Ali H.
    Hall, Timothy L.
    Ghani, Khurshid R.
    Maxwell, Adam D.
    MacConaghy, Brian
    Roberts, William W.
    [J]. JOURNAL OF ENDOUROLOGY, 2018, 32 (08) : 724 - 729
  • [3] Aldoukhi AH, 2019, J UROLOGY, V201, pE73
  • [4] Elevated Renal Pelvic Pressures during Percutaneous Nephrolithotomy Risk Higher Postoperative Pain and Longer Hospital Stay
    Alsyouf, Muhannad
    Abourbih, Samuel
    West, Benjamin
    Hodgson, Herbert
    Baldwin, D. Duane
    [J]. JOURNAL OF UROLOGY, 2018, 199 (01) : 193 - 198
  • [5] [Anonymous], 2021, TGA RECALL SFDA NATL
  • [6] Recent advances in infrared laser lithotripsy
    Fried, Nathaniel M.
    [J]. BIOMEDICAL OPTICS EXPRESS, 2018, 9 (09): : 4552 - 4568
  • [7] Gyrus ACMI IN. C, 2021, SOLTIVE PREM SUPERPU
  • [8] Gyrus ACMI. Inc, 2021, SOLTIVE PREM SUPERPU
  • [9] High power holmium:YAG versus thulium fiber laser treatment of kidney stones in dusting mode: ablation rate and fragment size studies
    Hardy, Luke A.
    Vinnichenko, Viktoriya
    Fried, Nathaniel M.
    [J]. LASERS IN SURGERY AND MEDICINE, 2019, 51 (06) : 522 - 530
  • [10] Thulium fiber laser lithotripsy in an in vitro ureter model
    Hardy, Luke A.
    Wilson, Christopher R.
    Irby, Pierce B.
    Fried, Nathaniel M.