Image quality determination of a novel digital detector for X-ray imaging and cone -beam computed tomography applications

被引:7
作者
Alzahrani, Hanan [1 ,2 ]
Richards, Sion [3 ]
Sedgwick, Iain [3 ]
Seller, Paul [3 ]
Konstantinidis, Anastasios [4 ]
Royle, Gary [5 ]
Ricketts, Kate [1 ]
机构
[1] UCL, Res Dept Tissue & Energy, Div Surg & Intervent Sci, London W1W 7TS, England
[2] Taibah Univ, Appl Med Sci Fac, Medina 42353, Saudi Arabia
[3] Rutherford Appleton Lab, Sci & Technol Facil Council, Harwell Campus, Didcot OX11 0QX, Oxon, England
[4] Christie NHS Fdn Trust, Diagnost Radiol & Radiat Protect, Christie Med Phys & Engn, Manchester M20 4BX, Lancs, England
[5] UCL, Dept Med Phys & Biomed Engn, London WC1E 6BT, England
关键词
QUANTUM EFFICIENCY; RADIOGRAPHY SYSTEMS; NOISE; RADIATION; ENERGY; SCINTILLATORS; ABSORPTION; DQE;
D O I
10.1016/j.nima.2020.163914
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The demand for adequate image quality with low radiation doses for patients has greatly increased. This is especially true in the case of position verification in radiotherapy which requires a high number of images per patient. This study presents a physical characterisation of a new clinical detector named “Lassena (CsI)” based on a thick layer of structured thallium activated caesium iodide and complementary metal-oxide semiconductor technology with active pixel sensor architecture for general X-ray imaging and cone-beam computed tomography (CBCT) applications. We made a critical appraisal of its performance for the first time and determined its signal transfer property (STP) and its detective quantum efficiency (DQE) by acquiring the pre-sampling modulation transfer function (pMTF) and normalised noise power spectrum (NNPS) in addition to the dark current calculation. The investigation was conducted with the application of three X-ray beam qualities: (50 kV (RQA3), 70 kV (RQA5) and 90 kV (RQA7)) in compliance with the International Electrotechnical Commission (IEC 62220-1(2003)) standard. The STP was found to be linear with the coefficient of determination (R2) more than 0.9995 in all cases. The spatial resolution and NNPS results led to acceptable DQE values at all energies; in particular the DQE values at 0.5 line pairs per mm (DQE(0.5)) which were 0.46 for RQA3, 0.52-0.56 for RQA5 and 0.55-0.59 for RQA7. Lastly, the dark current was 2.51 pA/cm2 for a 50μm pixel pitch. For CBCT applications, Lassena (CsI) showed very promising results. © 2020
引用
收藏
页数:7
相关论文
共 41 条
[1]   Megavoltage cone beam computed tomography: Commissioning and evaluation of patient dose [J].
Abou-Elenein, Hassan S. ;
Attalla, Ehab M. ;
Ammar, H. ;
Eldesoky, Ismail ;
Farouk, Mohamed ;
Zaghloul, Mohamed S. .
JOURNAL OF MEDICAL PHYSICS, 2011, 36 (04) :205-212
[2]  
[Anonymous], ESSENTIAL PHYS MED I
[3]  
[Anonymous], [No title captured]
[4]  
[Anonymous], [No title captured]
[5]  
[Anonymous], [No title captured]
[6]   A comparison of digital radiography systems in terms of effective detective quantum efficiency [J].
Bertolini, Marco ;
Nitrosi, Andrea ;
Rivetti, Stefano ;
Lanconelli, Nico ;
Pattacini, Pierpaolo ;
Ginocchi, Vladimiro ;
Iori, Mauro .
MEDICAL PHYSICS, 2012, 39 (05) :2617-2627
[7]  
Beutel J., 2000, HDB MED IMAGING
[8]   Comparison of Methods for Estimating the Conversion Gain of CMOS Active Pixel Sensors [J].
Bohndiek, Sarah E. ;
Blue, Andrew ;
Clark, Andy T. ;
Prydderch, Mark L. ;
Turchetta, Renato ;
Royle, Gary J. ;
Speller, Robert D. .
IEEE SENSORS JOURNAL, 2008, 8 (9-10) :1734-1744
[9]   STUDIES OF X-RAY-ENERGY ABSORPTION AND QUANTUM NOISE PROPERTIES OF X-RAY SCREENS BY USE OF MONTE-CARLO SIMULATION [J].
CHAN, HP ;
DOI, K .
MEDICAL PHYSICS, 1984, 11 (01) :37-46
[10]  
Chan Mf, 2011, Biomed Imaging Interv J, V7, pe11, DOI 10.2349/biij.7.2.e11