Comparison between laser scanning tomography and computerised image analysis of the optic disc

被引:12
作者
Azuara-Blanco, A
Spaeth, GL
Nicholl, J
Lanzl, IM
Augsburger, JJ
机构
[1] Univ Nottingham, Queens Med Ctr, Dept Ophthalmol, Nottingham NG7 2RD, England
[2] Thomas Jefferson Univ, Jefferson Med Coll, Wills Eye Hosp, Glaucoma Serv, Philadelphia, PA 19107 USA
[3] Thomas Jefferson Univ, Jefferson Med Coll, Wills Eye Hosp, Retina Serv, Philadelphia, PA 19107 USA
关键词
D O I
10.1136/bjo.83.3.295
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims-To study the interchangeability of the measurements of the optic disc topography obtained by one computerised image analyser and one confocal laser tomographic scanner. Methods-One eye of 28 patients with glaucoma or glaucoma suspects was studied. All cases had simultaneous stereoscopic disc photographs taken with the fundus camera Topcon TRC-SS and optic disc examination with the Heidelberg retina tomograph (HRT) during the same visit. The optic disc photographs were digitised and analysed with the Topcon ImageNet (TI) system. Three variables of the optic disc topography provided by the TI and the HRT were compared-cup volume (CV), rim area (RA), and cup area to disc area ratio (CA/DA). Results-The mean values of CV and RA provided by the TI (0.52 (SD 0.32) mm(3) and 1.58 (0.39) mm(2), respectively) were greater (p<0.01) than the mean values of CV and RA determined by the HRT (0.32 (0.25) mm(3), and 1.33 (0.47) mm(2), respectively). The mean value of CA/DA provided by the TI (0.42 (0.14)) and the HRT (0.42 (0.18)) was similar (p=0.93). Correlation coefficients between measurements obtained by the two methods ranged from 0.53 to 0.73. Conclusion-There was a significant discrepancy in the measurements of rim area and cup volume of the optic disc obtained by a computerised image analyser and a laser scanning tomograph.
引用
收藏
页码:295 / 298
页数:4
相关论文
共 20 条
[1]   Reproducibility of optic disk topographic measurements with the Topcon ImageNet and the Heidelberg retina tomograph [J].
Azuara-Blanco, A ;
Harris, A ;
Cantor, LB .
OPHTHALMOLOGICA, 1998, 212 (02) :95-98
[2]   CORRECTION OF OPTIC DISK MEASUREMENTS ON FUNDUS PHOTOGRAPHS [J].
BENGTSSON, B ;
KRAKAU, CET .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1992, 230 (01) :24-28
[3]   COMPARISON OF CLINICIAN JUDGMENT WITH DIGITIZED IMAGE-ANALYSIS IN THE DETECTION OF INDUCED OPTIC DISK CHANGE IN MONKEY EYES [J].
BURGOYNE, CF ;
QUIGLEY, HA ;
VARMA, R .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1995, 120 (02) :176-183
[4]   LASER SCANNING TOMOGRAPHY AND STEREOPHOTOGRAMMETRY IN 3-DIMENSIONAL OPTIC DISC ANALYSIS [J].
BURK, ROW ;
ROHRSCHNEIDER, K ;
TAKAMOTO, T ;
VOLCKER, HE ;
SCHWARTZ, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1993, 231 (04) :193-198
[5]  
CIOFFI GA, 1993, OPHTHALMOLOGY, V100, P57
[6]  
Coleman AL, 1996, J GLAUCOMA, V5, P384
[7]   VARIABILITY OF DEPTH MEASUREMENTS OF THE OPTIC-NERVE HEAD AND PERIPAPILLARY RETINA WITH COMPUTERIZED IMAGE-ANALYSIS [J].
DANDONA, L ;
QUIGLEY, HA ;
JAMPEL, HD .
ARCHIVES OF OPHTHALMOLOGY, 1989, 107 (12) :1786-1792
[8]   Comparison between tomographic scanning evaluation and photographic measurement of the neuroretinal rim [J].
Dichtl, A ;
Jonas, JB ;
Mardin, CY .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1996, 121 (05) :494-501
[9]   REPRODUCIBILITY OF TOPOGRAPHIC MEASUREMENTS OF THE NORMAL AND GLAUCOMATOUS OPTIC-NERVE HEAD WITH THE LASER TOMOGRAPHIC SCANNER [J].
DREHER, AW ;
TSO, PC ;
WEINREB, RN .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1991, 111 (02) :221-229
[10]  
Lichter P R, 1976, Trans Am Ophthalmol Soc, V74, P532