Comparative Analysis of Optoelectronic Accuracy in the Laboratory Setting Versus Clinical Operative Environment: A Systematic Review

被引:7
作者
Cunningham, Bryan W. [1 ,2 ]
Brooks, Daina M. [1 ]
机构
[1] MedStar Union Mem Hosp, Musculoskeletal Res & Innovat Inst, Dept Orthopaed Surg, Baltimore, MD 21218 USA
[2] Georgetown Univ, Sch Med, Dept Orthopaed Surg, Washington, DC USA
关键词
optoelectronic accuracy; spinal surgery; spinal biomechanics; imaging; navigation; robotics; PEDICLE SCREW PLACEMENT; MINIATURE ROBOTIC GUIDANCE; ASSISTED SPINAL SURGERY; BIOMECHANICAL EVALUATION; TARGET REGISTRATION; FIXATION DEVICES; LEARNING-CURVE; LUMBAR SPINE; HUMAN MOTION; NAVIGATION;
D O I
10.1177/21925682211035083
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objectives: The optoelectronic camera source and data interpolation process serve as the foundation for navigational integrity in robotic-assisted surgical platforms. The current systematic review serves to provide a basis for the numerical disparity observed when comparing the intrinsic accuracy of optoelectronic cameras versus accuracy in the laboratory setting and clinical operative environments. Methods: Review of the PubMed and Cochrane Library research databases was performed. The exhaustive literature compilation obtained was then vetted to reduce redundancies and categorized into topics of intrinsic accuracy, registration accuracy, musculoskeletal kinematic platforms, and clinical operative platforms. Results: A total of 465 references were vetted and 137 comprise the basis for the current analysis. Regardless of application, the common denominators affecting overall optoelectronic accuracy are intrinsic accuracy, registration accuracy, and application accuracy. Intrinsic accuracy equaled or was less than 0.1 mm translation and 0.1 degrees rotation per fiducial. Controlled laboratory platforms reported 0.1 to 0.5 mm translation and 0.1 to 1.0 degrees rotation per array. Accuracy in robotic-assisted spinal surgery reported 1.5 to 6.0 mm translation and 1.5 to 5.0 degrees rotation when comparing planned to final implant position. Conclusions: Navigational integrity and maintenance of fidelity of optoelectronic data is the cornerstone of robotic-assisted spinal surgery. Transitioning from controlled laboratory to clinical operative environments requires an increased number of steps in the optoelectronic kinematic chain and error potential. Diligence in planning, fiducial positioning, system registration and intraoperative workflow have the potential to improve accuracy and decrease disparity between planned and final implant position.
引用
收藏
页码:59S / 74S
页数:16
相关论文
共 137 条
[1]  
Abbasi HR, 2001, STUD HEALTH TECHNOL, V81, P1
[2]   Image-guidance, Robotics, and the Future of Spine Surgery [J].
Ahern, Daniel P. ;
Gibbons, Denys ;
Schroeder, Gregory D. ;
Vaccaro, Alexander R. ;
Butler, Joseph S. .
CLINICAL SPINE SURGERY, 2020, 33 (05) :179-184
[3]   Accuracy and Feasibility of Optoelectronic Sensors for Weed Mapping in Wide Row Crops [J].
Andujar, Dionisio ;
Ribeiro, Angela ;
Fernandez-Quintanilla, Cesar ;
Dorado, Jose .
SENSORS, 2011, 11 (03) :2304-2318
[4]  
[Anonymous], 2002, E17720 ASTM
[5]  
[Anonymous], 1994, Accuracy trueness and precision of measurement methods and results
[6]   Lumbar pedicle: surgical anatomic evaluation and relationships [J].
Attar, A ;
Ugur, HC ;
Uz, A ;
Tekdemir, I ;
Egemen, N ;
Genc, Y .
EUROPEAN SPINE JOURNAL, 2001, 10 (01) :10-15
[7]   The Surgical Navigation System with Optical Position Determination Technology and Sources of Errors [J].
Avrunin, O. G. ;
Alkhorayef, M. ;
Saied, Husham Farouk Ismail ;
Tymkovych, M. Y. .
JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2015, 5 (04) :689-696
[8]   Gait analysis methods in rehabilitation [J].
Baker, Richard .
JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2006, 3 (1)
[9]  
Baria Dinah, 2020, JB JS Open Access, V5, pe0034, DOI 10.2106/JBJS.OA.19.00034
[10]   Quality of motion considerations in numerical analysis of motion restoring implants of the spine [J].
Bowden, Anton E. ;
Guerin, Heather L. ;
Villarraga, Marta L. ;
Patwardhan, Avinash G. ;
Ochoa, Jorge A. .
CLINICAL BIOMECHANICS, 2008, 23 (05) :536-544