A COMPARISON BETWEEN SPECTROSCOPIC PERFORMANCE OF HgI2 AND CdZnTe FRISCH COLLAR DETECTORS

被引:14
作者
Kargar, A. [1 ]
Ariesanti, E. [2 ]
McGregor, D. S. [2 ]
机构
[1] Radiat Monitoring Devices Inc, Watertown, MA 02472 USA
[2] Kansas State Univ, Mech & Nucl Engn Dept, SMART Lab, Manhattan, KS 66506 USA
关键词
CdZnTe spectrometer; HgI2; detector; Frisch collar device; COLLECTION;
D O I
10.13182/NT11-A12281
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In this study, the charge collection efficiencies (CCEs) of a 7.8- X 7.8- X 15.6-mm(3) CdZnTe Frisch collar detector and a 2.1- X 2.1- X 4.1-mm(3) HgI2 Frisch collar detector were measured and compared. Two Frisch collar devices were designed and fabricated to have identical aspect ratios of 2.0 to maintain similar weighting potential distributions. Pulse-height spectra were acquired from both Frisch collar devices with a standard calibration gamma-ray source of Cs-137, and the results are presented. As known, the Frisch collar alters the weighting potential within the planar device and enhances the CCE distributions. Thus, the parameters affecting these distributions have great impact on the pulse-height spectrum. The device length and mobility-lifetime product mu tau have great impacts on CCE. Primarily, crystal (device) length L directly affects CCE because more charge carriers are trapped in longer devices with longer traveling distances. Alternatively, the better mobility-lifetime product of the charge carriers enhances CCE of the fabricated device. It is shown in this study that as a result of similarity in shape for both devices (equal aspect ratio), the weighting potential distributions resemble each other. However, as a result of the trapping effect (due to both length and mu tau), the CCE profiles are not the same, and the CdZnTe detector shows more uniform response to gamma rays and, therefore, better spectroscopic performance (even with a longer device length), which is confirmed through CCE simulations. Finally, by applying the CCE model to the HgI2 Frisch collar device, the mobility-lifetime products mu(e,h)tau(e,h) of electrons and holes were estimated to be 0.0008 and 0.00003 cm(2).V-1, respectively, for the HgI2 crystal.
引用
收藏
页码:131 / 137
页数:7
相关论文
共 15 条
[1]   Fabrication and spectroscopy results of mercuric iodide Frisch collar detectors [J].
Ariesanti, E. ;
Kargar, A. ;
McGregor, D. S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 624 (03) :656-661
[2]   Comparison of 5 and 10 mm thick HgI2 pixelated γ-ray spectrometers [J].
Baciak, JE ;
He, Z .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 505 (1-2) :191-194
[3]   Charge collection characteristics of Frisch collar CdZnTe gamma-ray spectrometers [J].
Harrison, Mark J. ;
Kargar, Alireza ;
McGregor, Douglas S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (01) :134-137
[4]   CdZnTe Frisch collar detectors for γ-ray spectroscopy [J].
Kargar, A ;
Jones, AM ;
McNeil, WJ ;
Harrison, MJ ;
McGregor, DS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 558 (02) :497-503
[5]   Charge collection efficiency characterization of a HgI2 Frisch collar spectrometer with collimated high energy gamma rays [J].
Kargar, A. ;
Ariesanti, E. ;
James, S. ;
McGregor, D. S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01) :186-192
[6]   Effect of Crystal Length on CdZnTe Frisch Collar Device Performance [J].
Kargar, A. ;
Brooks, A. C. ;
Harrison, M. J. ;
Chen, H. ;
Awadalla, S. ;
Bindley, G. ;
McGregor, D. S. .
2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, :2017-+
[7]  
Kargar A., 2009, SPIE P, V7449, P744908
[8]  
Kargar A., 2009, Characterization and Optimization of Cadmium Zinc Telluride Frisch Collar Gamma-Ray Spectrometers and Their Development in an Array of Detectors
[9]  
KARGAR A, 2007, SPIE P, V6706
[10]   Characterization of charge carrier collection in a CdZnTe Frisch collar detector with a highly collimated 137Cs source [J].
Kargar, Alireza ;
Harrison, Mark J. ;
Brooks, Adam C. ;
McGregor, Douglas S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 620 (2-3) :270-278