Robust Adaptive Extended Kalman Filtering for Real Time MR-Thermometry Guided HIFU Interventions

被引:36
作者
Roujol, Sebastien [1 ,2 ]
de Senneville, Baudouin Denis [1 ]
Hey, Silke [1 ]
Moonen, Chrit [1 ]
Ries, Mario [1 ]
机构
[1] Univ Victor Segalen Bordeaux, FRE CNRS 3313, Lab Mol & Funct Imaging Physiol Therapy, F-33076 Bordeaux, France
[2] Univ Bordeaux, UMR CNRS 5800, Lab Bordelais Rech Informat, F-33405 Talence, France
关键词
Biomedical signal processing; Kalman filters; magnetic resonance imaging (MRI); real time systems;
D O I
10.1109/TMI.2011.2171772
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Real time magnetic resonance (MR) thermometry is gaining clinical importance formonitoring and guiding high intensity focused ultrasound (HIFU) ablations of tumorous tissue. The temperature information can be employed to adjust the position and the power of the HIFU system in real time and to determine the therapy endpoint. The requirement to resolve both physiological motion of mobile organs and the rapid temperature variations induced by state-of-the-art high-power HIFU systems require fast MRI-acquisition schemes, which are generally hampered by low signal-to-noise ratios (SNRs). This directly limits the precision of real time MR-thermometry and thus in many cases the feasibility of sophisticated control algorithms. To overcome these limitations, temporal filtering of the temperature has been suggested in the past, which has generally an adverse impact on the accuracy and latency of the filtered data. Here, we propose a novel filter that aims to improve the precision of MR-thermometry while monitoring and adapting its impact on the accuracy. For this, an adaptive extended Kalman filter using a model describing the heat transfer for acoustic heating in biological tissues was employed together with an additional outlier rejection to address the problem of sparse artifacted temperature points. The filter was compared to an efficient matched FIR filter and outperformed the latter in all tested cases. The filter was first evaluated on simulated data and provided in the worst case (with an approximate configuration of the model) a substantial improvement of the accuracy by a factor 3 and 15 during heat up and cool down periods, respectively. The robustness of the filter was then evaluated during HIFU experiments on a phantom and in vivo in porcine kidney. The presence of strong temperature artifacts did not affect the thermal dose measurement using our filter whereas a high measurement variation of 70% was observed with the FIR filter.
引用
收藏
页码:533 / 542
页数:10
相关论文
共 18 条
  • [1] Chauvenet A., 1863, MANUAL SPHERICAL PRA
  • [2] MR-GUIDED FOCUSED ULTRASOUND SURGERY
    CLINE, HE
    SCHENCK, JF
    HYNYNEN, K
    WATKINS, RD
    SOUZA, SP
    JOLESZ, FA
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (06) : 956 - 965
  • [3] SIGNAL-TO-NOISE IN PHASE-ANGLE RECONSTRUCTION - DYNAMIC-RANGE EXTENSION USING PHASE REFERENCE OFFSETS
    CONTURO, TE
    SMITH, GD
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) : 420 - 437
  • [4] Non-invasive determination of tissue thermal parameters from high intensity focused ultrasound treatment monitored by volumetric MRI thermometry
    Dragonu, Iulius
    de Oliveira, Philippe Lourenco
    Laurent, Christophe
    Mougenot, Charles
    Grenier, Nicolas
    Moonen, Chrit T. W.
    Quesson, Bruno
    [J]. NMR IN BIOMEDICINE, 2009, 22 (08) : 843 - 851
  • [5] Kalman R.E., 1960, NEW APPROACH LINEAR, DOI [DOI 10.1115/1.3662552, 10.1115/1.3662552]
  • [6] Real-time monitoring of hepatocellular carcinoma radiofrequency ablation by quantitative temperature MRI
    Laumonier, Herve
    Blanc, Jean-Frederic
    Quesson, Bruno
    Seror, Olivier
    Laurent, Christophe
    Bioulac-Sage, Paulette
    Balabaud, Charles
    Trillaud, Herve
    [J]. SEMINARS IN LIVER DISEASE, 2006, 26 (04) : 391 - 397
  • [7] Three-Dimensional Spatial and Temporal Temperature Control With MR Thermometry-Guided Focused Ultrasound (MRgHIFU)
    Mougenot, Charles
    Quesson, Bruno
    de Senneville, Baudouin Denis
    de Oliveira, Philippe Lourenco
    Sprinkhuizen, Sara
    Palussiere, Jean
    Grenier, Nicolas
    Moonen, Chrit T. W.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (03) : 603 - 614
  • [8] Oppenheim A.V., 2009, Discrete-time signal processing, V3rd
  • [10] CONCURRENT HYPERTHERMIA ESTIMATION SCHEMES BASED ON EXTENDED KALMAN FILTERING AND REDUCED-ORDER MODELING
    POTOCKI, JK
    THARP, HS
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1993, 9 (06) : 849 - 865