Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor

被引:76
作者
Jones, CR [1 ]
Taylor, K [1 ]
Poston, L [1 ]
Shennan, AH [1 ]
机构
[1] St Thomas Hosp, Guys Kings & St Thomas Sch Med, Fetal Hlth Res Grp, Dept Obstet & Gynaecol, London SE1 7EH, England
关键词
blood pressure; measurement; validation; oscillometric;
D O I
10.1038/sj.jhh.1001141
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
This study has determined the accuracy of the Welch Allyn 'Vital Signs' blood pressure (BP) monitoring device in an adult population, according to the criteria of the British Hypertension Society (British Hypertension Society, 1993). Validation was also classified by the American 'Association for the Advancement of Medical Instrumentation' (AAMI). The Welch Allyn 'Vital Signs' monitor achieved a BHS grade A for both systolic and diastolic BP. It also met the criteria for the AAMI protocol. The mean differences between trained observers using a mercury sphygmomanometer and the device were -1 +/- 5 mm Hg for systolic BP and -3 +/- 4 mm Hg for diastolic BP. The Welch Allyn 'Vital Signs' monitor can be recommended for clinical use in the adult population. This is one of the first oscillometric monitors designed for routine use in a hospital environment that has been found to provide an accurate assessment of BP by accepted validation procedures.
引用
收藏
页码:191 / 195
页数:5
相关论文
共 10 条
[1]  
*ASS ADV MED INSTR, 1992, AM NAT STAND EL AUT
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   AMBULATORY BLOOD-PRESSURE MONITORING USE IN HYPERTENSION RESEARCH AND CLINICAL-PRACTICE [J].
MANCIA, G ;
DIRIENZO, M ;
PARATI, G .
HYPERTENSION, 1993, 21 (04) :510-524
[4]  
O'Brien, 1996, Blood Press Monit, V1, P55
[5]  
O'Brien, 1998, Blood Press Monit, V3, P205
[6]  
O'Brien E., 1993, J HYPERTENSION S2, V11, P43, DOI DOI 10.1097/00004872-199306000-00013
[7]   Inaccuracy of the Dinamap 8100 portable monitor [J].
OBrien, E ;
Atkins, N .
LANCET, 1997, 349 (9057) :1026-1026
[8]   SHORT REPORT - ACCURACY OF THE PROFILOMAT AMBULATORY BLOOD-PRESSURE MEASURING SYSTEM DETERMINED BY THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL [J].
OBRIEN, E ;
MEE, F ;
ATKINS, N ;
OMALLEY, K .
JOURNAL OF HYPERTENSION, 1992, 10 (10) :1285-1286
[9]  
OBRIEN E, 1995, ABC HYPERTENSION, P5
[10]   NATIONAL STANDARD FOR MEASUREMENT OF RESTING AND AMBULATORY BLOOD PRESSURES WITH AUTOMATED SPHYGMOMANOMETERS [J].
WHITE, WB ;
BERSON, AS ;
ROBBINS, C ;
JAMIESON, MJ ;
PRISANT, LM ;
ROCCELLA, E ;
SHEPS, SG .
HYPERTENSION, 1993, 21 (04) :504-509