VitaFlow™ transcatheter valve system in the treatment of severe aortic stenosis: One-year results of a multicenter study

被引:40
作者
Zhou, Daxin [1 ]
Pan, Wenzhi [1 ]
Wang, Jianan [2 ]
Wu, Yongjian [3 ]
Chen, Mao [4 ]
Modine, Thomas [5 ]
Mylotte, Darren [6 ,7 ]
Piazza, Niccolo [8 ]
Ge, Junbo [1 ]
机构
[1] Fudan Univ, Zhongshan Hosp, Shanghai Inst Cardiovasc Dis, Dept Cardiol, 180 Fenglin Rd, Shanghai 200032, Peoples R China
[2] Zhejiang Univ, Dept Cardiol, Affiliated Hosp 2, Sch Med, Hangzhou, Peoples R China
[3] Fuwai Hosp, Natl Ctr Cardiovasc Dis, Dept Cardiol, Beijing, Peoples R China
[4] Sichuan Univ, Dept Cardiol, West China Hosp, Chengdu, Peoples R China
[5] CHRU Lille, Hop Cardiol, Pole Chirurg CardioVasc, Lille, France
[6] Natl Univ Ireland, Galway Univ Hosp, Galway, Ireland
[7] Natl Univ Ireland, SAOLTA Hlth Care Grp, Galway, Ireland
[8] McGill Univ, Dept Intervent Cardiol, Hlth Ctr, Montreal, PQ, Canada
关键词
aortic stenosis; bicuspid aortic valve; self-expanding; transcatheter aortic valve replacement; REPLACEMENT; IMPLANTATION; DEVICES;
D O I
10.1002/ccd.28226
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectiveThis study reports the 1-year clinical outcomes of the VitaFlow (TM) transcatheter aortic valve system in the treatment of severe aortic stenosis. BackgroundThe VitaFlow (TM) system (MicroPort (R), Shanghai, China) was developed as a novel transcatheter aortic valve replacement system to mitigate or circumvent some of the challenges associated with heavily calcified valves and bicuspid valves. MethodsFrom September 2014 to November 2017, a prospective, multicenter, single arm study was conducted in 11 centers in China. The primary end point was all cause mortality at 12months. ResultsOne hundred and ten symptomatic aortic stenosis patients (60 men, 50 women; mean age 77.734.78years) at prohibitive or high risk for surgery were enrolled. Mean society of thoracic surgeons score was 8.84 +/- 5.58%. All-cause mortality was 2.7% at 1-year. Major stroke, major vascular complication, coronary artery obstruction, new pacemaker implantation occurred in 2.7, 2.7, 1.8, and 19.1% at 1-year follow-up, respectively. No patients had moderate or severe paravalvular leak at 1-year. At 1 year follow-up, 97% of patients had New York heart association <= II. Patients with bicuspid valves had similar outcomes as those patients with tricuspid aortic valve stenosis. ConclusionsThe 12-month clinical results support the safety and efficacy of VitaFlow (TM) in the treatment of patients with severe aortic stenosis, including patients with bicuspid aortic valve.
引用
收藏
页码:332 / 338
页数:7
相关论文
共 20 条
[1]   Comparison of Balloon-Expandable vs Self-expandable Valves in Patients Undergoing Transcatheter Aortic Valve Replacement The CHOICE Randomized Clinical Trial [J].
Abdel-Wahab, Mohamed ;
Mehilli, Julinda ;
Frerker, Christian ;
Neumann, Franz-Josef ;
Kurz, Thomas ;
Toelg, Ralph ;
Zachow, Dirk ;
Guerra, Elena ;
Massberg, Steffen ;
Schaefer, Ulrich ;
El-Mawardy, Mohamed ;
Richardt, Gert .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2014, 311 (15) :1503-1514
[2]   Transcatheter Aortic-Valve Replacement with a Self-Expanding Prosthesis [J].
Adams, David H. ;
Popma, Jeffrey J. ;
Reardon, Michael J. ;
Yakubov, Steven J. ;
Coselli, Joseph S. ;
Deeb, G. Michael ;
Gleason, Thomas G. ;
Buchbinder, Maurice ;
Hermiller, James, Jr. ;
Kleiman, Neal S. ;
Chetcuti, Stan ;
Heiser, John ;
Merhi, William ;
Zorn, George ;
Tadros, Peter ;
Robinson, Newell ;
Petrossian, George ;
Hughes, G. Chad ;
Harrison, J. Kevin ;
Conte, John ;
Maini, Brijeshwar ;
Mumtaz, Mubashir ;
Chenoweth, Sharla ;
Oh, Jae K. .
NEW ENGLAND JOURNAL OF MEDICINE, 2014, 370 (19) :1790-1798
[3]   Comparison of the Effectiveness of Transcatheter Aortic Valve Implantation in Patients With Stenotic Bicuspid Versus Tricuspid Aortic Valves (from the German TAVI Registry) [J].
Bauer, Timm ;
Linke, Axel ;
Sievert, Horst ;
Kahlert, Philipp ;
Hambrecht, Rainer ;
Nickenig, Georg ;
Hauptmann, Karl Eugen ;
Sack, Stefan ;
Gerckens, Ulrich ;
Schneider, Steffen ;
Zeymer, Uwe ;
Zahn, Ralf .
AMERICAN JOURNAL OF CARDIOLOGY, 2014, 113 (03) :518-521
[4]  
Baumgartner H, 2018, EUR HEART J, V39, P1980, DOI [10.1093/eurheartj/ehx636, 10.1093/eurheartj/ehx391]
[5]   Morphological Characteristics of Severe Aortic Stenosis in China: Imaging Corelab Observations From the First Chinese Transcatheter Aortic Valve Trial [J].
Jilaihawi, Hasan ;
Wu, Yongjian ;
Yang, Yuejin ;
Xu, Liang ;
Chen, Mao ;
Wang, Jianan ;
Kong, Xiangqing ;
Zhang, Ruiyan ;
Wang, Moyang ;
Lv, Bin ;
Wang, Wei ;
Xu, Bo ;
Makkar, Raj R. ;
Sievert, Horst ;
Gao, Runlin .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2015, 85 :752-761
[6]   Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document [J].
Kappetein, A. Pieter ;
Head, Stuart J. ;
Genereux, Philippe ;
Piazza, Nicolo ;
van Mieghem, Nicolas M. ;
Blackstone, Eugene H. ;
Brott, Thomas G. ;
Cohen, David J. ;
Cutlip, Donald E. ;
van Es, Gerrit-Anne ;
Hahn, Rebecca T. ;
Kirtane, Ajay J. ;
Krucoff, Mitchell W. ;
Kodali, Susheel ;
Mack, Michael J. ;
Mehran, Roxana ;
Rodes-Cabau, Josep ;
Vranckx, Pascal ;
Webb, John G. ;
Windecker, Stephan ;
Serruys, Patrick W. ;
Leon, Martin B. .
EUROPEAN HEART JOURNAL, 2012, 33 (19) :2403-+
[7]   Paravalvular regurgitation after transcatheter aortic valve replacement with the Edwards sapien valve in the PARTNER trial: characterizing patients and impact on outcomes [J].
Kodali, Susheel ;
Pibarot, Philippe ;
Douglas, Pamela S. ;
Williams, Mathew ;
Xu, Ke ;
Thourani, Vinod ;
Rihal, Charanjit S. ;
Zajarias, Alan ;
Doshi, Darshan ;
Davidson, Michael ;
Tuzcu, E. Murat ;
Stewart, William ;
Weissman, Neil J. ;
Svensson, Lars ;
Greason, Kevin ;
Maniar, Hersh ;
Mack, Michael ;
Anwaruddin, Saif ;
Leon, Martin B. ;
Hahn, Rebecca T. .
EUROPEAN HEART JOURNAL, 2015, 36 (07) :449-+
[8]   Transcatheter or Surgical Aortic-Valve Replacement in Intermediate-Risk Patients [J].
Leon, Martin B. ;
Smith, Craig R. ;
Mack, Michael J. ;
Makkar, Raj R. ;
Svensson, Lars G. ;
Kodali, Susheel K. ;
Thourani, Vinod H. ;
Tuzcu, E. Murat ;
Miller, D. Craig ;
Herrmann, Howard C. ;
Doshi, Darshan ;
Cohen, David J. ;
Pichard, Augusto D. ;
Kapadia, Samir ;
Dewey, Todd ;
Babaliaros, Vasilis ;
Szeto, Wilson Y. ;
Williams, Mathew R. ;
Kereiakes, Dean ;
Zajarias, Alan ;
Greason, Kevin L. ;
Whisenant, Brian K. ;
Hodson, Robert W. ;
Moses, Jeffrey W. ;
Trento, Alfredo ;
Brown, David L. ;
Fearon, William F. ;
Pibarot, Philippe ;
Hahn, Rebecca T. ;
Jaber, Wael A. ;
Anderson, William N. ;
Alu, Maria C. ;
Webb, John G. .
NEW ENGLAND JOURNAL OF MEDICINE, 2016, 374 (17) :1609-1620
[9]   Transcatheter Aortic-Valve Implantation for Aortic Stenosis in Patients Who Cannot Undergo Surgery. [J].
Leon, Martin B. ;
Smith, Craig R. ;
Mack, Michael ;
Miller, D. Craig ;
Moses, Jeffrey W. ;
Svensson, Lars G. ;
Tuzcu, E. Murat ;
Webb, John G. ;
Fontana, Gregory P. ;
Makkar, Raj R. ;
Brown, David L. ;
Block, Peter C. ;
Guyton, Robert A. ;
Pichard, Augusto D. ;
Bavaria, Joseph E. ;
Herrmann, Howard C. ;
Douglas, Pamela S. ;
Petersen, John L. ;
Akin, Jodi J. ;
Anderson, William N. ;
Wang, Duolao ;
Pocock, Stuart .
NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (17) :1597-1607
[10]   Transcatheter Aortic Valve Implantation with the Self-Expandable Venus A-Valve and CoreValve Devices: Preliminary Experiences in China [J].
Liao, Yan-Biao ;
Zhao, Zhen-Gang ;
Wei, Xin ;
Xu, Yuan-Ning ;
Zuo, Zhi-Liang ;
Li, Yi-jian ;
Zheng, Ming-Xia ;
Feng, Yuan ;
Chen, Mao .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 89 :528-533