Hyperecho as the indicator of tissue necrosis during microbubble-assisted high intensity focused ultrasound: Sensitivity, specificity and predictive value

被引:44
作者
Yu, Tinghe [1 ,3 ]
Xu, Chuanshan [2 ,3 ]
机构
[1] Chongqing Med Univ, Inst Ultrasound Engn Med, Chongqing 400016, Peoples R China
[2] Chongqing Med Univ, Inst Ultrasound Imaging, Chongqing 400016, Peoples R China
[3] Third Mil Med Univ, Southwest Hosp, Chongqing, Peoples R China
基金
中国国家自然科学基金;
关键词
high-intensity focused ultrasound; hyperecho; microbubble; grayscale ultrasound image; monitoring;
D O I
10.1016/j.ultrasmedbio.2008.01.012
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this study was to determine the sensitivity, specificity and predictive values of hyperecho in grayscale ultrasonic images as the indicator of tissue necrosis in microbubble-assisted high-intensity focused ultrasound (HIFU) exposure in vivo. Livers and kidneys of the rabbit were exposed to HIFU (control group) or microbubble-assisted HIFU (experimental group); a continuous line of ablation, viz. linear scan, was performed to destruct tissues. Tissue responses were evaluated macroscopically and microscopically 24 h after HIFU. The cases of positive (hyperecho occurred and tissue necrotized), false positive (tissue was unaffected although hyperecho appeared), negative (echo was not changed and tissue was intact) and false negative (tissue was destructed despite the lack of hyperecho) were counted, and then the sensitivity, specificity and positive and negative predictive values of hyperecho were calculated. The sensitivity, specificity, positive predictive value and negative predictive value were 49.25% vs. 79.63% (p < 0.001), 45.45% vs. 30.00%,84.62% vs. 86.00% and 12.82% vs. 21.43% for liver and 76.06% vs. 81.25%, 26.53% vs. 41.67%, 60.00% vs. 82.28% (p = 0.002) and 43.33% vs. 40.00% for kidney, in control and experimental groups, respectively. Rates varied between tissue types in control group. These findings indicated that the use of microbubble during HIFU improved the sensitivity in liver and the positive predictive value in kidney. The specificity and negative predictive value were poor. Hyperecho could only be used as the indicator of tissue necrosis in some tissue types.
引用
收藏
页码:1343 / 1347
页数:5
相关论文
共 19 条
[1]   Influence of ablated tissue on the formation of high-intensity focused ultrasound lesions [J].
Chen, LL ;
terHaar, G ;
Hill, CR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (06) :921-931
[2]   Role of acoustic cavitation in the delivery and monitoring of cancer treatment by high-intensity focused ultrasound (HIFU) [J].
Coussios, C. C. ;
Farny, C. H. ;
Ter Haar, G. ;
Roy, R. A. .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2007, 23 (02) :105-120
[3]   Elastography for the follow-up of high-intensity focused ultrasound prostate cancer treatment:: Initial comparison with MRI [J].
Curiel, L ;
Souchon, R ;
Rouvière, O ;
Gelet, A ;
Chapelon, JY .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2005, 31 (11) :1461-1468
[4]   Use of a microbubble agent to increase the effects of high intensity focused ultrasound on liver tissue [J].
Kaneko, Y ;
Maruyama, T ;
Takegami, K ;
Watanabe, T ;
Mitsui, H ;
Hanajiri, K ;
Nagawa, HA ;
Matsumoto, Y .
EUROPEAN RADIOLOGY, 2005, 15 (07) :1415-1420
[5]   Contrast-enhanced ultrasound assessment of tissue response to high-intensity focused ultrasound [J].
Kennedy, JE ;
Ter Haar, GR ;
Wu, F ;
Gleeson, FV ;
Roberts, ISD ;
Middleton, MR ;
Cranston, D .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2004, 30 (06) :851-854
[6]  
Kennedy JE, 2004, ULTRASONICS, V42, P931, DOI 10.1016/j.ultras.2004.01.089
[7]   Attenuation coefficient and sound speed in human myometrium and uterine fibroid tumors [J].
Keshavarzi, A ;
Vaezy, S ;
Kaczkowski, PJ ;
Keilman, G ;
Martin, R ;
Chi, EY ;
Garcia, R ;
Fujimoto, V .
JOURNAL OF ULTRASOUND IN MEDICINE, 2001, 20 (05) :473-480
[8]   Hyperecho in ultrasound images of HIFU therapy: Involvement of cavitation [J].
Rabkin, BA ;
Zderic, V ;
Vaezy, S .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2005, 31 (07) :947-956
[9]   Biological and physical mechanisms of HIFU-induced hyperecho in ultrasound images [J].
Rabkin, Brian A. ;
Zderic, Vfsna ;
Crum, Lawrence A. ;
Vaezy, Shahram .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2006, 32 (11) :1721-1729
[10]   Enhanced heat deposition using ultrasound contrast agent - Modeling and experimental observations [J].
Razansky, D ;
Einziger, PD ;
Adam, DR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (01) :137-147