Isometric Muscle Strength in Youth Assessed by Hand-held Dynamometry: A Feasibility, Reliability, and Validity Study

被引:132
作者
Hebert, Luc J. [1 ]
Maltais, Desiree B. [2 ]
Lepage, Celine [2 ,3 ]
Saulnier, Joanne [3 ]
Crete, Melanie [3 ]
Perron, Marc [4 ]
机构
[1] Natl Def Canada, CFHS HQ D Med Pol, Ottawa, ON, Canada
[2] Univ Laval, Ctr Interdisciplinary Res Rehabil & Social Integr, Quebec City, PQ, Canada
[3] Quebec Rehabil Ctr, Quebec City, PQ, Canada
[4] Ctr Sante Valcartier, Quebec City, PQ, Canada
关键词
adolescent; child; muscle strength; muscle strength dynamometer; reliability and validity; CEREBRAL-PALSY; CHILDREN; TORQUE;
D O I
10.1097/PEP.0b013e318227ccff
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Purpose: To determine, with respect to measurement of maximal isometric torque (MIT) using a specific hand-held dynamometer (HHD) protocol, (1) protocol feasibility over a wide age range, (2) intra-and inter-rater reliability, (3) standard error of measurement, and (4) concurrent validity. Methods: The MIT of selected upper and lower limb muscle groups was assessed (n = 74; age = 4-17.5 years) using a standardized, HHD protocol. Testing was repeated in 20 adolescents (n = 10 for each muscle group), who were also assessed with a Cybex dynamometer. Results: The protocol was feasible for all participants. Mean intra-and interrater reliability [intraclass correlation coefficient (ICC)] varied from 0.75 to 0.98, except for ankle dorsiflexor interrater reliability (mean ICC = 0.67). The standard error of measurement varied from 0.5 to 4.9 Nm and was highest for hip extensors. Mean concurrent validity (ICC) varied from 0.78 to 0.93, except for ankle plantar flexors (mean ICC = 0.48). Conclusions: Our HHD protocol was feasible over a wide age range and most MIT values were valid and reliable. (Pediatr Phys Ther 2011;23:289-299)
引用
收藏
页码:289 / 299
页数:11
相关论文
共 26 条
[1]  
[Anonymous], 2008, 2008 Physical Activity Guidelines for Americans
[2]  
AUFSESSER P, 1996, ADAPT PHYS ACT Q, V13, P153
[3]  
Blimkie CJR, 1998, PEDIATRIC ANAEROBIC PERFORMANCE, P193
[4]  
Bohannon R.W., 1990, Critical Reviews in Physical Medicine and Rehabilitation, V2, P75
[5]   Intertester reliability of hand-held dynamometry: A concise summary of published research [J].
Bohannon, RW .
PERCEPTUAL AND MOTOR SKILLS, 1999, 88 (03) :899-902
[6]   Muscle force production and functional performance in spastic cerebral palsy: Relationship of cocontraction [J].
Damiano, DL ;
Martellotta, TL ;
Sallivan, DJ ;
Granata, KP ;
Abel, MF .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2000, 81 (07) :895-900
[7]  
Dawson C., 1992, Pediatr. Exerc. Sci, V4, P340, DOI [10.1123/pes.4.4.340, DOI 10.1123/PES.4.4.340]
[8]  
DOMHOLDT E, 2005, REHABILITATION RES P, P245
[9]  
Dvir Z., 2004, ISOKINETICS MUSCLE T
[10]   Isometric muscle torque in children 5 to 15 years of age: Normative data [J].
Eek, Meta Nystrom ;
Kroksmark, Anna-Karin ;
Beckung, Eva .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2006, 87 (08) :1091-1099