Evaluation of in vivo wear of vitamin E-diffused highly crosslinked polyethylene at five years A MULTICENTRE RADIOSTEREOMETRIC ANALYSIS STUDY

被引:39
作者
Galea, V. P. [2 ]
Connelly, J. W. [2 ]
Shareghi, B. [1 ,3 ]
Karrholm, J. [1 ,3 ]
Skoldenberg, O. [1 ,4 ]
Salemyr, M. [1 ,4 ]
Laursen, M. B. [1 ,5 ]
Muratoglu, O. [2 ,6 ]
Bragdon, C. [2 ,6 ]
Malchau, H. [2 ,6 ]
机构
[1] Massachusetts Gen Hosp, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Harris Orthopaed Lab, Boston, MA 02114 USA
[3] Univ Gothenburg, Sahlgrenska Acad, Dept Orthopaed, Inst Clin Sci, Gothenburg, Sweden
[4] Karolinska Inst, Div Orthopaed, Danderyd Hosp, Dept Clin Sci, Stockholm, Sweden
[5] Aalborg Univ Hosp, Dept Orthopaed, Aalborg, Denmark
[6] Harvard Med Sch, Dept Orthopaed Surg, Boston, MA USA
关键词
MOLECULAR-WEIGHT POLYETHYLENE; TOTAL HIP-ARTHROPLASTY; OXIDATION; RSA; OSTEOLYSIS; RESISTANCE;
D O I
10.1302/0301-620X.100B12.BJJ-2018-0371.R1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aims The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure. Patients and Methods A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery. Results We observed similar bedding in one year postoperatively and wear two years postoperatively between the two types of liner. However, there was significantly more penetration of the femoral head in the ModXLPE cohort compared with the VEPE cohort five years postoperatively (p < 0.001). The only variables independently predictive of increased wear were ModXLPE (vs VEPE) liner type (beta = 0.22, p = 0.010) and metal (vs ceramic) femoral head (beta = 0.21, p = 0.013). There was no association between increased wear and the development of radiolucency (p = 0.866) or PROMs. No patient had evidence of osteolysis. Conclusion Five years postoperatively, patients with VEPE (vs ModXLPE) and ceramic (vs metal) femoral heads had decreased wear. The rates of wear for both liners were very low and have not led to any osteolysis or implant failure due to aseptic loosening.
引用
收藏
页码:1592 / 1599
页数:8
相关论文
共 29 条
  • [1] American Joint Replacement Registry (AJRR), 2017, ANN REPORT
  • [2] Tradeoffs amongst fatigue, wear, and oxidation resistance of cross-linked ultra-high molecular weight polyethylene
    Atwood, Sara A.
    Van Citters, Douglas W.
    Patten, Eli W.
    Furmanski, Jevan
    Ries, Michael D.
    Pruitt, Lisa A.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2011, 4 (07) : 1033 - 1045
  • [3] Clarke IC, 2000, CLIN ORTHOP RELAT R, P34
  • [4] A literature review of the association between wear rate and osteolysis in total hip arthroplasty
    Dumbleton, JH
    Manley, MT
    Edidin, AA
    [J]. JOURNAL OF ARTHROPLASTY, 2002, 17 (05) : 649 - 661
  • [5] The basis for a second-generation highly cross-linked UHMWPE
    Dumbleton, John H.
    D'Antonio, James A.
    Manley, Michael T.
    Capello, William N.
    Wang, Aiguo
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (453) : 265 - 271
  • [6] Effectiveness of Vitamin-E-Doped Polyethylene in Joint Replacement: A Literature Review
    Gigante, Antonio
    Bottegoni, Carlo
    Ragone, Vincenza
    Banci, Lorenzo
    [J]. JOURNAL OF FUNCTIONAL BIOMATERIALS, 2015, 6 (03): : 889 - 900
  • [8] Harris WH, 2001, CLIN ORTHOP RELAT R, P66
  • [9] Highly cross-linked polyethylene - The debate is over - In the affirmative
    Jasty, M
    Rubash, HE
    Muratoglu, O
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (04) : 55 - 58
  • [10] Kurtz SM, 2016, PDL HANDB SER, P57