Evaluation of electromyographic monitoring during insertion of thoracic pedicle screws

被引:56
作者
Reidy, DP
Houlden, D
Nolan, PC
Kim, M
Finkelstein, JA
机构
[1] Sunnybrook & Womens Coll, Hlth Sci Ctr, Div Orthopaed Surg, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Toronto, ON M4N 3M5, Canada
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2001年 / 83B卷 / 07期
关键词
D O I
10.1302/0301-620X.83B7.12017
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We prospectively studied the use of intercostal EMG monitoring as an indicator of the accuracy of the placement of pedicle screws in the thoracic spine. We investigated 95 thoracic pedicles in 17 patients. Before insertion of the screw, the surgeon recorded his assessment of the integrity of the pedicle track. We then stimulated the track using a K-wire pedicle probe connected to a constant current stimulator. A compound muscle action potential (CMAP) was recorded from the appropriate intercostal or abdominal muscles. Postoperative CT was performed to establish the position of the screw. The stimulus intensity required to evoke a muscle response was correlated with the position of the screw on the CT scan. There were eight unrecognised breaches of the pedicle. Using 7.0 mA as a threshold, the sensitivity of EMG was 0.50 in detecting a breached pedicle and the specificity was 0.83. Thoracic pedicle screws were accurately placed in more than 90% of patients. EMG monitoring did not significantly improve the reliability of placement of the screw.
引用
收藏
页码:1009 / 1014
页数:6
相关论文
共 25 条
[11]   Intraoperative electromyography during thoracolumbar spinal surgery [J].
Holland, NR .
SPINE, 1998, 23 (17) :1915-1922
[12]   Higher electrical stimulus intensities are required to activate chronically compressed nerve roots - Implications for intraoperative electromyographic pedicle screw testing [J].
Holland, NR ;
Lukaczyk, TA ;
Riley, LH ;
Kostuik, JP .
SPINE, 1998, 23 (02) :224-227
[13]   Continuous electromyographic monitoring to detect nerve root injury during thoracolumbar scoliosis surgery [J].
Holland, NR ;
Kostuik, JP .
SPINE, 1997, 22 (21) :2547-2550
[14]   Pedicle screw instrumentation of the thoracic spine in idiopathic scoliosis [J].
Liljenqvist, UR ;
Halm, HFH ;
Link, TM .
SPINE, 1997, 22 (19) :2239-2245
[15]   ANATOMIC ANALYSIS OF PEDICLE CORTICAL AND CANCELLOUS DIAMETER AS RELATED TO SCREW SIZE [J].
MISENHIMER, GR ;
PEEK, RD ;
WILTSE, LL ;
ROTHMAN, SLG ;
WIDELL, EH .
SPINE, 1989, 14 (04) :367-372
[16]  
RAISZADEH K, 2000, P N AM SPIN SOC NEW, P131
[17]  
ROYCAMILLE R, 1986, ORTHOP CLIN N AM, V17, P147
[18]   Evaluation of pedicle screw position in thoracic and lumbar spine fixation using plain radiographs and computed tomography - A prospective study of 35 patients [J].
Sapkas, GS ;
Papadakis, SA ;
Stathakopoulos, DP ;
Papagelopoulos, PJ ;
Badekas, AC ;
Kaiser, JH .
SPINE, 1999, 24 (18) :1926-1929
[19]  
SJOSTROM L, 1993, J SPINAL DISORD, V6, P225
[20]   MEASURING THE ACCURACY OF DIAGNOSTIC SYSTEMS [J].
SWETS, JA .