Interrater reliability of algometry in measuring pressure pain thresholds in healthy humans, using multiple raters

被引:505
作者
Chesterton, Linda S. [1 ]
Sim, Julius [1 ]
Wright, Christine C. [1 ]
Foster, Nadine E. [1 ]
机构
[1] Univ Keele, Primary Care Musculoskeletal Res Ctr, Keele ST5 5BG, Staffs, England
关键词
reliability; multiple raters; algometry; pressure pain threshold; quantitative sensory testing;
D O I
10.1097/AJP.0b013e318154b6ae
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective: To evaluate interrater reliability using 5 newly trained observers in the assessment of pressure pain threshold (PPT) using a fixed-angle algometer. Methods: The study design comprised 2 phases. Phase 1: 5 undergraduate physical therapists were trained in algometry at a predefined angle, at a rate of 5 Newtons (N)/s, to the first dorsal interosseous muscle. Each observer then underwent a competency test of the application speed. The aim was to achieve repeated applications at 5 N/s without visual feedback from the algometer. Phase 2: the 5 observers measured PPT of 13 healthy volunteers, at the first dorsal interosseous muscle. The sequence of observer measurements for each participant was randomized. Mean PPT values for each observer were analyzed using repeated measures analysis of variance, intraclass correlation coefficient (ICC2,1), and standard error of measurement, with 95% confidence intervals (CIs). Results: No significant differences between observers' mean values were found (P = 0.094), suggesting no bias. The ICC was 0.91 (95% CI 0.82, 0.97). The standard error of measurement value was 6.27 N/cm(2) (95% CI 5.35, 7.59). Differences in PPT measurements of more than 17.39 N/cm(2) (1.77 kg/cm(2)) are likely to exceed the magnitude of measurement error, and could be used to indicate true change. This margin of error is, however, somewhat larger than a previously proposed minimum clinically important difference in PPT of 14.71 N/cm(2) (1.5 kg/cm(2)). Discussion: This study provides new evidence that trained observers can apply an algometer at a consistent rate and provide highly reliable measures of PPT in healthy humans, when PPT is calculated as the mean of 3 trials.
引用
收藏
页码:760 / 766
页数:7
相关论文
共 51 条
  • [1] [Anonymous], 1996, Health measurement scales
  • [2] Antonaci F, 1998, SCAND J REHABIL MED, V30, P3
  • [3] Batterham A.M., 2000, Phys. Ther. Sport, V1, P54, DOI DOI 10.1054/PTSP.2000.0010
  • [4] A NOTE ON THE USE OF THE INTRACLASS CORRELATION-COEFFICIENT IN THE EVALUATION OF AGREEMENT BETWEEN 2 METHODS OF MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. COMPUTERS IN BIOLOGY AND MEDICINE, 1990, 20 (05) : 337 - 340
  • [5] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [6] MEASUREMENTS OF HUMAN PRESSURE-PAIN THRESHOLDS ON FINGERS AND TOES
    BRENNUM, J
    KJELDSEN, M
    JENSEN, K
    JENSEN, TS
    [J]. PAIN, 1989, 38 (02) : 211 - 217
  • [7] EVALUATION OF PAIN THRESHOLD USING A SIMPLE PRESSURE ALGOMETER
    BUCHANAN, HM
    MIDGLEY, JA
    [J]. CLINICAL RHEUMATOLOGY, 1987, 6 (04) : 510 - 517
  • [8] Effects of TENS frequency, intensity and stimulation site parameter manipulation on pressure pain thresholds in healthy human subjects
    Chesterton, LS
    Foster, NE
    Wright, CC
    Baxter, GD
    Barlas, P
    [J]. PAIN, 2003, 106 (1-2) : 73 - 80
  • [9] Gender differences in pressure pain threshold in healthy humans
    Chesterton, LS
    Barlasb, P
    Foster, NE
    Baxter, GD
    Wright, CC
    [J]. PAIN, 2003, 101 (03) : 259 - 266
  • [10] EVALUATION OF PRESSURE PAIN THRESHOLD IN HEAD AND NECK MUSCLES BY ELECTRONIC ALGOMETER - INTRARATER AND INTERRATER RELIABILITY
    CHUNG, SC
    UM, BY
    KIM, HS
    [J]. CRANIO-THE JOURNAL OF CRANIOMANDIBULAR & SLEEP PRACTICE, 1992, 10 (01): : 28 - 34