Establishment of the 222Rn traceability chain with the Romanian standard system

被引:8
作者
Sahagia, M. [1 ]
Luca, A. [1 ]
Waetjen, A. C. [1 ]
Antohe, A. [1 ]
Ivan, C. [1 ]
Varlam, C. [2 ]
Faurescu, I. [2 ]
Cassette, P. [3 ]
机构
[1] IFIN HH, Horia Hulubei Natl Inst R&D Phys & Nucl Engn, Bucharest 077125, Romania
[2] ICSI Rm, Natl Inst R&D Cryogen & Isotop Technol, Valcea, Romania
[3] LNE CEA, LIST, Lab Natl Henri Becquerel, F-91191 Gif Sur Yvette, France
关键词
Rn-222; measurement; Traceability chain; Secondary standards; ABSOLUTE MEASUREMENT; RADON; COUNTER;
D O I
10.1016/j.nima.2010.12.019
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper presents the results obtained in the establishment of the upper level of the traceability chain, primary and secondary Rn-222 measurements, with the standard system realized at IFIN-HH, Romania. The following subjects are presented: (i) the principle of establishing the traceability chain; (ii) the setting up of the final metallic system for the radon standardization and the experimental method; (iii) results obtained in the realization of the traceability chain, by absolute radon standardization with the liquid scintillation counting method (LSC) and relative measurements of activity, using the HPGe gamma-ray spectrometry method and the CENTRONIC IG12/20A ionization chamber, and its validation. The final purpose of the work is to calibrate adequately the secondary standard systems. HPGe spectrometer and ionization chamber, for using them in the measurement of the working standards, various types of recipients containing radon gas, prepared in the laboratory with the radon system. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:73 / 79
页数:7
相关论文
共 20 条
[1]  
[Anonymous], 0090209 ICRP
[2]  
BE MM, 2008, MONOGRAPHIE BUREAU I, P79
[3]   Absolute measurement of the activity of 222Rn using a proportional counter [J].
Busch, I ;
Greupner, H ;
Keyser, U .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 481 (1-3) :330-338
[4]   Standardization of 222Rn by LSC and comparison with α- and γ-spectrometry [J].
Cassette, P. ;
Sahagia, M. ;
Grigorescu, L. ;
Lépy, M. C. ;
Picolo, J. L. .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (10-11) :1465-1470
[5]   Primary standards of radon [J].
De Felice, P. .
METROLOGIA, 2007, 44 (04) :S82-S86
[6]   Primary and secondary measurements of 222Rn [J].
Dersch, R .
APPLIED RADIATION AND ISOTOPES, 2004, 60 (2-4) :387-390
[7]  
*ISO, 1995, ISO IEC OIML BIPM GU
[8]  
LEPY MC, 2001, LNHB0109A ETNA
[9]  
LUCA A, 2007, IFIN HH RADIONUCLIDE
[10]   Primary activity measurements with 4πγ NaI(T1) counting and Monte Carlo calculated efficiencies [J].
Nedjadi, Youcef ;
Spring, Philippe ;
Bailat, Claude ;
Decombaz, Marc ;
Triscone, Gilles ;
Gostely, Jean-Jacques ;
Laedermann, Jean-Pascal ;
Bochud, Francois O. .
APPLIED RADIATION AND ISOTOPES, 2007, 65 (05) :534-538