The design, fabrication, and in-vitro evaluation of an amperometric biochip that is designed for the continuous in vivo monitoring of physiological analytes is described. The 2 x 4 x 0.5 mm biochip contains two platinum working enzyme electrodes that adopt the microdise array design to minimize diffusional limitations associated with enzyme kinetics. This configuration permits either dual analyte sensing or a differential response analytical methodology during amperometric detection of a single analyte. The working enzyme electrodes are complemented by a large area platinized platinum counter electrode and a silver reference electrode. The biorecognition layer of the working electrodes was fabricated from around 1.0-mu m-thick composite membrane of principally tetraethylene glycol (TEGDA) cross-linked poly(2-hydroxyethyl methacrylate) that also contained a derivatized polypyrrole component and a biomimetic methacrylate component with pendant phosphorylcholine groups. These two additional components were introduced to provide interference screening and in vivo biocompatibility, respectively. This composite membrane was used to immobilize glucose oxidase and lactate oxidase onto both planar and microdisc array electrode designs, which were then used to assay for in vitro glucose and lactate, respectively. The glucose biosensor exhibited a dynamic linear range of 0.10-13.0 mM glucose with a response time (t(95)) of 50 s. The immobilized glucose oxidase within the hydrogel yielded a K-m(app) of 35 mM, not significantly different from that for the native, solution-borne enzyme (33 mM). The microdisc array biosensor displayed linearity for assayed lactate up to 90 mM, which represented a 30-fold increase in linear dynamic lactate range compared to the biosensor with the planar electrode configuration. Preliminary in vitro operational stability tests performed with the microdisc array lactate biosensor demonstrated retention of 80% initial biosensor response after five days of continuous operation in buffer under physiologic conditions of pH and temperature.
引用
收藏
页码:345 / 355
页数:11
相关论文
共 50 条
  • [41] LACTATE MONITORING WITH SUBCUTANEOUS MICRODIALYSIS IN PATIENTS WITH SHOCK - A PILOT-STUDY
    DEBOER, J
    POTTHOFF, H
    MULDER, POM
    DOFFERHOFF, ASM
    VANTHIEL, RJ
    PLIJTERGROENDIJK, H
    KORF, J
    CIRCULATORY SHOCK, 1994, 43 (02) : 57 - 63
  • [42] A MEMS SENSOR FOR CONTINUOUS MONITORING OF GLUCOSE IN SUBCUTANEOUS TISSUE
    Huang, X.
    Li, S.
    Schultz, J.
    Wang, Q.
    Lin, Q.
    IEEE 22ND INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS 2009), 2009, : 352 - 355
  • [43] Glucose prediction from continuous subcutaneous monitoring data
    Maran, A
    Sparacino, G
    Pavanini, A
    Crepaldi, C
    Poscia, A
    Tiengo, A
    Avogaro, A
    Cobelli, C
    DIABETES, 2004, 53 : A105 - A105
  • [44] Continuous glucose monitoring on the ICU using a subcutaneous sensor
    Punke, M. A.
    Decker, C.
    Wodack, K.
    Reuter, D. A.
    Kluge, S.
    MEDIZINISCHE KLINIK-INTENSIVMEDIZIN UND NOTFALLMEDIZIN, 2015, 110 (05) : 360 - 363
  • [45] Subcutaneous continuous glucose monitoring in severe burn patients
    Klonoff, David C.
    CRITICAL CARE MEDICINE, 2007, 35 (05) : 1445 - 1445
  • [46] A bifunctional biosensor for subcutaneous glucose monitoring by reverse iontophoresis
    Liu, Jian
    Jiang, Liying
    Liu, Hongmin
    Cai, Xinxia
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2011, 660 (01) : 8 - 13
  • [47] Role of continuous subcutaneous glucose monitoring in intensive care
    Marics, Gabor
    Koncz, Levente
    Korner, Anna
    Mikos, Borbala
    Toth-Heyn, Peter
    ORVOSI HETILAP, 2013, 154 (27) : 1043 - 1048
  • [48] Subcutaneous glucose monitoring - Microdialysis vs. intracorporeal
    Ward, WK
    DIABETES CARE, 2002, 25 (02) : 410 - 410
  • [49] Design of a dual-band implantable antenna and development of skin mimicking gels for continuous glucose monitoring
    Karacolak, Tutku
    Hood, Aaron Z.
    Topsakal, Erdem
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2008, 56 (04) : 1001 - 1008
  • [50] IMPLANTABLE MONITORING SYSTEMS - DESIGN CONSIDERATIONS AND CHALLENGES
    YOMTOV, BM
    KIM, J
    AYERS, GM
    JOURNAL OF ELECTROCARDIOLOGY, 1994, 27 : 74 - 77