Single event upsets in implantable cardioverter defibrillators

被引:60
作者
Bradley, PD [1 ]
Normand, E
机构
[1] Univ Wollongong, Dept Engn Phys, Wollongong, NSW 2522, Australia
[2] Boeing Def & Space Grp, Seattle, WA 98124 USA
关键词
D O I
10.1109/23.736549
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Single event upsets (SEU) have been observed in implantable cardiac defibrillators. The incidence of SEUs is well modeled by upset rate calculations attributable to the secondary cosmic ray neutron flux. The effect of recent interpretations of the shape of the heavy ion cross-section curve on neutron burst generation rate calculations is discussed. The model correlates well with clinical experience and is consistent with the expected geographical variation of the secondary cosmic ray neutron flux. The observed SER was 9.3 x 10(-12) upsets/bit-hr from 22 upsets collected over a total of 284672 device days. This is the first clinical data set obtained indicating the effects of cosmic radiation on implantable devices. Importantly, it may be used to predict the susceptibility of future implantable device designs to cosmic radiation. The significance of cosmic radiation effects relative to other radiation sources applicable to implantable devices is discussed.
引用
收藏
页码:2929 / 2940
页数:12
相关论文
共 47 条
[1]   ESTIMATING THE DIMENSIONS OF THE SEU-SENSITIVE VOLUME [J].
ABDELKADER, WG ;
MCNULTY, PJ ;
ELTELEATY, S ;
LYNCH, JE ;
KHONDKER, AN .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (06) :1300-1304
[2]   CALCULATIONS OF NEUTRON-FLUX SPECTRA INDUCED IN EARTHS ATMOSPHERE BY GALACTIC COSMIC-RAYS [J].
ARMSTRONG, TW ;
CHANDLER, KC ;
BARISH, J .
JOURNAL OF GEOPHYSICAL RESEARCH, 1973, 78 (16) :2715-2726
[3]  
BRADLEY PD, 1996, THESIS U S WALES
[4]  
BROWNING R, 1985, IEEE T NUCL SCI, V32, P4137
[5]   AN ANALYTICAL METHOD FOR PREDICTING CMOS SRAM UPSETS WITH APPLICATION TO ASYMMETRICAL MEMORY CELLS [J].
BUEHLER, MG ;
ALLEN, RA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (06) :1637-1641
[6]   INFLUENCES ON SOFT ERROR RATES IN STATIC RAMS [J].
CARTER, PM ;
WILKINS, BR .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1987, 22 (03) :430-436
[7]   THE EFFECTS OF TEST CONDITIONS ON MOS RADIATION-HARDNESS RESULTS [J].
DRESSENDORFER, PV ;
SODEN, JM ;
HARRINGTON, JJ ;
NORDSTROM, TV .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (06) :4281-4287
[8]  
Fletcher GH., 1973, Textbook of Radiotherapy, V2nd ed
[9]   RADIATION DOSES DELIVERED TO SKIN BONE MARROW AND GONADS OF PATIENTS DURING CARDIAC CATHETERISATION AND ANGIOCARDIOGRAPHY [J].
GOUGH, JH ;
DAVIS, R ;
STACEY, AJ .
BRITISH JOURNAL OF RADIOLOGY, 1968, 41 (487) :508-&
[10]   OBSERVATION AND PREDICTION OF SEU IN HITACHI SRAMS IN LOW-ALTITUDE POLAR ORBITS [J].
HARBOESORENSEN, R ;
DALY, EJ ;
ADAMS, L ;
UNDERWOOD, CI ;
MULLER, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (06) :1498-1504