Dynamic volume vs respiratory correlated 4DCT for motion assessment in radiation therapy simulation

被引:26
作者
Coolens, Catherine [1 ,2 ]
Bracken, John [1 ]
Driscoll, Brandon [1 ]
Hope, Andrew [1 ,2 ]
Jaffray, David [1 ,2 ,3 ]
机构
[1] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5S 3S2, Canada
[3] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
关键词
motion; computed tomography; radiation therapy; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; LUNG-CANCER PATIENTS; TUMOR MOTION; CT SCANS; RADIOTHERAPY; VARIABILITY; ARTIFACTS; QUALITY;
D O I
10.1118/1.4704498
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Conventional (i.e., respiratory-correlated) 4DCT exploits the repetitive nature of breathing to provide an estimate of motion; however, it has limitations due to binning artifacts and irregular breathing in actual patient breathing patterns. The aim of this work was to evaluate the accuracy and image quality of a dynamic volume, CT approach (4D(vol)) using a 320-slice CT scanner to minimize these limitations, wherein entire image volumes are acquired dynamically without couch movement. This will be compared to the conventional respiratory-correlated 4DCT approach (RCCT). Methods: 4D(vol) CT was performed and characterized on an in-house, programmable respiratory motion phantom containing multiple geometric and morphological "tumor" objects over a range of regular and irregular patient breathing traces obtained from 3D fluoroscopy and compared to RCCT. The accuracy of volumetric capture and breathing displacement were evaluated and compared with the ground truth values and with the results reported using RCCT. A motion model was investigated to validate the number of motion samples needed to obtain accurate motion probability density functions (PDF). The impact of 4D image quality on this accuracy was then investigated. Dose measurements using volumetric and conventional scan techniques were also performed and compared. Results: Both conventional and dynamic volume 4DCT methods were capable of estimating the programmed displacement of sinusoidal motion, but patient breathing is known to not be regular, and obvious differences were seen for realistic, irregular motion. The mean RCCT amplitude error averaged at 4 mm (max. 7.8 mm) whereas the 4D(vol) CT error stayed below 0.5 mm. Similarly, the average absolute volume error was lower with 4D(vol) CT. Under irregular breathing, the 4D(vol) CT method provides a close description of the motion PDF (cross-correlation 0.99) and is able to track each object, whereas the RCCT method results in a significantly different PDF from the ground truth, especially for smaller tumors (cross-correlation ranging between 0.04 and 0.69). For the protocols studied, the dose measurements were higher in the 4D(vol) CT method (40%), but it was shown that significant mAs reductions can be achieved by a factor of 4-5 while maintaining image quality and accuracy. Conclusions: 4D(vol) CT using a scanner with a large cone-angle is a promising alternative for improving the accuracy with which respiration-induced motion can be characterized, particularly for patients with irregular breathing motion. This approach also generates 4DCT image data with a reduced total scan time compared to a RCCT scan, without the need for image binning or external respiration signals within the 16 cm scan length. Scan dose can be made comparable to RCCT by optimization of the scan parameters. In addition, it provides the possibility of measuring breathing motion for more than one breathing cycle to assess stability and obtain a more accurate motion PDF, which is currently not feasible with the conventional RCCT approach. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4704498]
引用
收藏
页码:2669 / 2681
页数:13
相关论文
共 29 条
  • [21] Four-dimensional computed tomography: Image formation and clinical protocol
    Rietzel, E
    Pan, TS
    Chen, GTY
    [J]. MEDICAL PHYSICS, 2005, 32 (04) : 874 - 889
  • [22] Design of 4D treatment planning target volumes
    Rietzel, Eike
    Liu, Arthur K.
    Doppke, Karen P.
    Wolfgang, John A.
    Chen, Aileen B.
    Chen, George T. Y.
    Choi, Noah C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (01): : 287 - 295
  • [23] Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy
    Seppenwoolde, Y
    Shirato, H
    Kitamura, K
    Shimizu, S
    van Herk, M
    Lebesque, JV
    Miyasaka, K
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (04): : 822 - 834
  • [24] Intrafractional tumor motion: Lung and liver
    Shirato, H
    Seppenwoolde, Y
    Kitamura, K
    Onimura, R
    Shimizu, S
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) : 10 - 18
  • [25] Pulmonary Tumor Measurements from X-Ray Computed Tomography in One, Two, and Three Dimensions
    Villemaire, Lauren
    Owrangi, Amir M.
    Etemad-Rezai, Roya
    Wilson, Laura
    O'Riordan, Elaine
    Keller, Harry
    Driscoll, Brandon
    Bauman, Glenn
    Fenster, Aaron
    Parraga, Grace
    [J]. ACADEMIC RADIOLOGY, 2011, 18 (11) : 1391 - 1402
  • [26] Motion effects in (intensity modulated) radiation therapy: a review
    Webb, S.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (13) : R403 - R425
  • [27] Mid-ventilation CT scan construction from four-dimensional respiration-correlated CT scans for radiotherapy planning of lung cancer patients
    Wolthaus, Jochem W. H.
    Schneider, Christoph
    Sonke, Jan-Jakob
    van Herk, Marcel
    Belderbos, Jose S. A.
    Rossi, Maddalena M. G.
    Lebesque, Joos V.
    Damen, Eugene M. F.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (05): : 1560 - 1571
  • [28] RETROSPECTIVE ANALYSIS OF ARTIFACTS IN FOUR-DIMENSIONAL CT IMAGES OF 50 ABDOMINAL AND THORACIC RADIOTHERAPY PATIENTS
    Yamamoto, Tokihiro
    Langner, Ulrich
    Loo, Billy W., Jr.
    Shen, John
    Keall, Paul J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (04): : 1250 - 1258
  • [29] Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms
    Zhang, Di
    Savandi, Ali S.
    Demarco, John J.
    Cagnon, Chris H.
    Angel, Erin
    Turner, Adam C.
    Cody, Dianna D.
    Stevens, Donna M.
    Primak, Andrew N.
    McCollough, Cynthia H.
    McNitt-Gray, Michael F.
    [J]. MEDICAL PHYSICS, 2009, 36 (03) : 1025 - 1038