Micro-computed Tomographic Evaluation of the Shaping Ability of 3 Reciprocating Single-File Nickel-Titanium Systems on Single- and Double-Curved Root Canals

被引:12
作者
Haupt, Franziska [1 ]
Pult, Jonas Robert Wilhelm [1 ]
Huelsmann, Michael [1 ]
机构
[1] Univ Med Ctr, Dept Prevent Dent Periodontol & Cariol, Robert Koch Str 40, D-37075 Gottingen, Germany
关键词
Canal transportation; centering ability; micro-computed tomography; S-shaped root canal; single-file reciprocating nickel-titanium systems; CYCLIC FATIGUE RESISTANCE; APICAL TRANSPORTATION; STAINLESS-STEEL; SHAPED CANALS; ROTARY; INSTRUMENTATION; GEOMETRY; REMOVAL; EFFICACY; WAVEONE;
D O I
10.1016/j.joen.2020.05.005
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction We performed a microcomputed tomographic assessment of the preparation of moderately single- and double-curved root canals using 3 single-file reciprocating nickel-titanium systems: S1 Plus Standard (Sendoline, Taby, Sweden), WaveOne Gold Primary (Dentsply Sirona, Ballaigues, Switzerland), and Reciproc R25 (VDW, Munich, Germany). Methods Seventy-five moderately curved mandibular molars with 2 separate mesial root canals were assigned to 3 experimental groups (n = 25) (ie, S1 Plus Standard, WaveOne Gold Primary, and Reciproc 25 groups) by forming matched triples according to curvature (15 degrees 40 degrees), radius (<= 18 mm), and type of curvature (single or double curved). Teeth were scanned before and after root canal preparation with a resolution of 10.5 mu m using microcomputed tomographic imaging (Bruker SkyScan 1272; Bruker microCT, Kontich, Belgium). The following parameters were assessed: changes in root canal volume and surface area, percentage of unshaped canal walls, structure model index, canal transportation, and centering ratio. Data were analyzed using 2- and 3-way analysis of variance with Tukey and Scheffe post hoc tests (significance level of 5%). Results No significant differences among groups were observed concerning all parameters. The type of curvature had no significant effect on all tested parameters. Within all experimental groups, canal transportation increased significantly from the apical to the coronal region, of which the majority was directed toward the furcational area. Conclusions Preparation with the 3 nickel-titanium systems did not result in significantly different dimensional changes, and there was no significant effect of the type of curvature on all tested parameters.
引用
收藏
页码:1130 / 1135
页数:6
相关论文
共 31 条
[1]   Effect of finishing instrumentation using NiTi hand files on volume, surface area and uninstrumented surfaces in C-shaped root canal systems [J].
Amoroso-Silva, P. ;
Alcalde, M. P. ;
Hungaro Duarte, M. A. ;
De-Deus, G. ;
Ordinola-Zapata, R. ;
Freire, L. G. ;
Cavenago, B. C. ;
De Moraes, I. G. .
INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (06) :604-611
[2]   Shaping ability of reciprocating single-file systems in severely curved canals: WaveOne and Reciproc versus WaveOne Gold and Reciproc blue [J].
Buerklein, Sebastian ;
Fluech, Stefanie ;
Schaefer, Edgar .
ODONTOLOGY, 2019, 107 (01) :96-102
[3]   Shaping Ability of Different Nickel-Titanium Systems in Simulated S-shaped Canals with and without Glide Path [J].
Buerklein, Sebastian ;
Poschmann, Thomas ;
Schaefen, Edgar .
JOURNAL OF ENDODONTICS, 2014, 40 (08) :1231-1234
[4]   Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement [J].
De-Deus, G. ;
Moreira, E. J. L. ;
Lopes, H. P. ;
Elias, C. N. .
INTERNATIONAL ENDODONTIC JOURNAL, 2010, 43 (12) :1063-1068
[5]   In Vitro Canal and Isthmus Debris Removal of the Self-Adjusting File, K3, and WaveOne Files in the Mesial Root of Human Mandibular Molars [J].
Dietrich, Matthew A. ;
Kirkpatrick, Timothy C. ;
Yaccino, John M. .
JOURNAL OF ENDODONTICS, 2012, 38 (08) :1140-1144
[6]   Counterclockwise or clockwise reciprocating motion for oval root canal preparation: a micro-CT analysis [J].
Espir, C. G. ;
Nascimento-Mendes, C. A. ;
Guerreiro-Tanomaru, J. M. ;
Freire, L. G. ;
Gavini, G. ;
Tanomaru-Filho, M. .
INTERNATIONAL ENDODONTIC JOURNAL, 2018, 51 (05) :541-548
[7]   Changes in Geometry and Transportation of Root Canals with Severe Curvature Prepared by Different Heat-treated Nickel-titanium Instruments: A Micro-computed Tomographic Study [J].
Filizola de Oliveira, Daniel Jose ;
Leoni, Graziela Bianchi ;
Goulart, Rafael da Silva ;
de Sousa-Neto, Manoel Damiao ;
Correa Silva Sousa, Yara Teresinha ;
Silva, Ricardo Gariba .
JOURNAL OF ENDODONTICS, 2019, 45 (06) :768-773
[8]   Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography [J].
Gambill, JM ;
Alder, M ;
delRio, CE .
JOURNAL OF ENDODONTICS, 1996, 22 (07) :369-375
[9]   Effects of three nickel titanium instrument systems on root canal geometry assessed by micro-computed tomography [J].
Gergi, R. ;
Osta, N. ;
Bourbouze, G. ;
Zgheib, C. ;
Arbab-Chirani, R. ;
Naaman, A. .
INTERNATIONAL ENDODONTIC JOURNAL, 2015, 48 (02) :162-170
[10]   Preparation of Oval-shaped Canals with TRUShape and Reciproc Systems: A Micro-Computed Tomography Study Using Contralateral Premolars [J].
Guimaraes, Ludmila S. ;
Gomes, Cinthya C. ;
Marceliano-Alves, Marilia F. ;
Cunha, Rodrigo S. ;
Provenzano, Jose C. ;
Siqueira, Jose F., Jr. .