Development and validation of an assay for the measurement of gentamicin concentrations in dried blood spots using UHPLC-MS/MS

被引:6
|
作者
Anibaletto dos Santos, Ana Laura [1 ,2 ]
Cezimbra da Silva, Anne Caroline [1 ,2 ]
Feltraco Lizot, Lilian de Lima [1 ,2 ]
Schneider, Anelise [2 ]
Meireles, Yasmin Fazenda [2 ]
Hahn, Roberta Zilles [1 ]
Pagnussat, Lidiane Riva [4 ]
Nonnenmacher, Julia Livia [5 ]
Hahn, Siomara Regina [3 ]
Linden, Rafael [1 ,2 ]
机构
[1] Univ Feevale, Analyt Toxicol Lab, Novo Hamburgo, RS, Brazil
[2] Univ Feevale, Grad Program Toxicol & Analyt Toxicol, Novo Hamburgo, Brazil
[3] Univ Passo Fundo, Coll Pharm, Passo Fundo, RS, Brazil
[4] Hosp Sao Vicente Paulo, Passo Fundo, RS, Brazil
[5] Univ Passo Fundo, Multidisciplinary Residency Program, Passo Fundo, RS, Brazil
关键词
Gentamicin; UHPLC-MS/MS; Dried blood spots; Therapeutic drug monitoring; Neonates; MASS-SPECTROMETRY; LC-MS/MS; HEMATOCRIT; TIME;
D O I
10.1016/j.jpba.2021.114448
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Gentamicin sulfate (GEN) is an aminoglycoside antibiotic with a narrow therapeutic range of plasma concentrations. The collection of venous blood represents a significant burden for patients, especially in neonatology. Dried blood spots (DBS) obtained from capillary blood can be an alternative for drug measurements in this particular population. This study aimed to develop and validate an assay for the quantification of GEN in DBS using UHPLC-MS/MS. Total GEN concentrations were obtained by adding the individual concentrations of the GEN forms C1, C1a, and C2. The assay used a DBS disk containing approximately 17 mu L of blood for GEN quantitation in the range of 0.1-40 mg L-1. Measurement accuracy for total GEN was in the range of 102.6-108.6%, inter-assay precision was 11.3-13.1% and intra-assay precision was 9.1-12.8.% GEN was stable for 21 days at - 20 and 8 degrees C, but only for 24 h at room temperature. Blood Hct affected the accuracy within acceptable limits (93.8-95% at Hct% of 30, 104.3-113% at Hct% of 50). Blood spotted volume did not affect GEN measurement accuracy. Concentrations of GEN in DBS obtained after heel pricks were correlated to plasma levels in a small cohort of neonatal patients. However, percentual differences between estimated plasma concentrations and actual plasma levels presented values between - 64-35.3% (average difference of - 1.9%). The use of DBS for the measurement of GEN concentrations can increase access to TDM of this antibiotic due to the ease of sample collection and the facilitated specimen transportation logistics when testing is not available onsite. (C) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Dried plasma spots for therapeutic monitoring of amikacin: Validation of an UHPLC-MS/MS assay and pharmacokinetic application
    Cezimbra da Silva, Anne Caroline
    Feltraco Lizot, Lilian de Lima
    Bastiani, Marcos Frank
    Antunes, Marina Venzon
    Brucker, Natalia
    Linden, Rafael
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2020, 184
  • [2] Development and validation of UHPLC-MS/MS methods for the quantification of colistin in plasma and dried plasma spots
    Cangemi, Giuliana
    Barco, Sebastiano
    Castagnola, Elio
    Tripodi, Gino
    Favata, Fabio
    D'Avolio, Antonio
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 129 : 551 - 557
  • [3] Development and validation of the UHPLC-MS/MS method for the quantitative determination of 25 PFAS in dried blood spots
    Galletto, Martina
    Ververi, Christina
    Massano, Marta
    Alladio, Eugenio
    Vincenti, Marco
    Salomone, Alberto
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (26) : 5671 - 5687
  • [4] A novel, rapid and simple UHPLC-MS/MS method for quantification of warfarin in dried blood spots
    Qu, Wenhao
    Li, Xianglian
    Tian, Guidi
    Liu, Linsheng
    Miao, Liyan
    ANALYTICAL BIOCHEMISTRY, 2022, 647
  • [5] Development and Validation of an UHPLC-MS/MS Assay for the Therapeutic Monitoring of Brivaracetam Plasma Concentrations in Patients with Epilepsy
    Mohamed, Susan
    Riva, Roberto
    Contin, Manuela
    THERAPEUTIC DRUG MONITORING, 2020, 42 (03) : 445 - 451
  • [6] "Center punch" and "whole spot" bioanalysis of apixaban in human dried blood spot samples by UHPLC-MS/MS
    Zheng, Naiyu
    Yuan, Long
    Ji, Qin C.
    Mangus, Heidi
    Song, Yan
    Frost, Charles
    Zeng, Jianing
    Aubry, Anne-Francoise
    Arnold, Mark E.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2015, 988 : 66 - 74
  • [7] A UHPLC-MS/MS Method for Therapeutic Drug Monitoring of Aciclovir and Ganciclovir in Plasma and Dried Plasma Spots
    Pigliasco, Federica
    Cafaro, Alessia
    Simeoli, Raffaele
    Barco, Sebastiano
    Magnasco, Alberto
    Faraci, Maura
    Tripodi, Gino
    Goffredo, Bianca Maria
    Cangemi, Giuliana
    BIOMEDICINES, 2021, 9 (10)
  • [8] A Validated UHPLC-MS/MS Method to Quantify Eight Antibiotics in Quantitative Dried Blood Spots in Support of Pharmacokinetic Studies in Neonates
    Liu, Qian
    Liu, Lanyu
    Yuan, Yu
    Xie, Feifan
    ANTIBIOTICS-BASEL, 2023, 12 (02):
  • [9] Development and validation of a sensitive LC-MS/MS method for determination of tacrolimus on dried blood spots
    Li, Qin
    Cao, Di
    Huang, Yue
    Xu, Hong
    Yu, Chen
    Li, Zhiping
    BIOMEDICAL CHROMATOGRAPHY, 2013, 27 (03) : 327 - 334
  • [10] Development and evaluation of a new test kit for determination of immunosuppressants in blood by UHPLC-MS/MS
    Wu, Lingjie
    Ye, Zhenjie
    Zhang, Xiaolong
    Zheng, Aixian
    Zhang, Xiaoying
    Chen, Lulu
    Ouyang, Dongsheng
    Zheng, Ling
    Liu, Xiaolong
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2022, 215