A new optional vena cava filter: Retrieval at 12 weeks in an animal model

被引:34
作者
Brountzos, EN
Kaufman, JA
Venbrux, AC
Brown, R
Harry, J
Kinst, TF
Kleshinski, S
Ravenscroft, AC
机构
[1] Oregon Hlth Sci Univ, Dotter Intervent Inst, Portland, OR 97201 USA
[2] Univ Athens, Sch Med, Dept Radiol 2, GR-10679 Athens, Greece
[3] George Washington Univ Hosp, Div Intervent Radiol, Washington, DC USA
[4] Johns Hopkins Univ, Div Comparat Med, Baltimore, MD USA
[5] Johns Hopkins Univ, Dept Surg, Baltimore, MD USA
[6] NMT Med Inc, Boston, MA USA
关键词
D O I
10.1097/01.RVI.0000079986.80153.CC
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To test the feasibility and safety of percutaneous retrieval of a new inferior vena cava (IVC) filter, the Recovery Filter (RF), acutely and after 12-week implantation in sheep. MATERIALS AND METHODS: The RF is a bilevel filter with stabilizing arms and elastic hooks that allow retrieval with a unique retrieval cone after incorporation into the wall of the IVC. Twenty-four filters were placed in the infrarenal IVCs of 18 sheep. In six sheep, two filters were placed and then removed immediately; three sheep were killed acutely and three were killed after a healing period of 3 weeks. In 12 sheep, a single filter was placed and then removed 12 weeks later; six were killed after retrieval and six were killed after an 8-week healing period. RESULTS: The mean (+/-SD) transverse vena caval diameter was 15.3 mm +/- 2. All filters were deployed as intended and retrieved without difficulty. At sacrifice, there was no evidence of IVC perforation or retroperitoneal abnormality. The IVCs of the animals in the acute retrieval group showed minimal acute superficial injury that was largely reversed at 3 weeks. At 12 weeks, there was evidence of transmural incorporation of filter elements with narrowing of the IVCs. Solitary fibrotic abnormalities were present in the aorta adjacent to IVC lesions in nine of the 12 animals in the 12-week group. The IVC and aortic abnormalities were largely healed, with reversal of IVC narrowing after 8 weeks. CONCLUSION: The recovery filter can be reliably and safely retrieved acutely and 12 weeks after implantation in sheep.
引用
收藏
页码:763 / 772
页数:10
相关论文
共 26 条
[1]   Initial experience in humans with a new retrievable inferior vena cava filter [J].
Asch, MR .
RADIOLOGY, 2002, 225 (03) :835-844
[2]   The Tempofilter((R)): A multicenter study of a new temporary caval filter implantable for up to six weeks [J].
Bovyn, G ;
Gory, P ;
Reynaud, P ;
Ricco, JB .
ANNALS OF VASCULAR SURGERY, 1997, 11 (05) :520-528
[3]   Caval incorporation of the LGM Vena Tech filter: An experimental study [J].
Crochet, D ;
BachLijour, B ;
Grossetete, R ;
Raffin, T ;
Nguyen, JM ;
Hurtrel, M ;
Daculsi, G ;
LeNihouannen, JC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (03) :419-425
[4]   SHORT-TERM PROPHYLAXIS OF PULMONARY-EMBOLISM BY USING A RETRIEVABLE VENA-CAVA FILTER [J].
DARCY, MD ;
SMITH, TP ;
HUNTER, DW ;
CASTANEDAZUNIGA, W ;
LUND, G ;
AMPLATZ, K .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 147 (04) :836-838
[5]   Animal experience in the Gunther Tulip retrievable inferior vena cava filter [J].
de Gregorio, MA ;
Gimeno, MJ ;
Tobio, R ;
Lostalé, F ;
Mainar, A ;
Beltrán, JM ;
Madariaga, B ;
Alfonso, ER ;
Medrano, J ;
Viloria, A .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 24 (06) :413-417
[6]   A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis [J].
Decousus, H ;
Leizorovicz, A ;
Parent, F ;
Page, Y ;
Tardy, B ;
Girard, P ;
Laporte, S ;
Faivre, R ;
Charbonnier, B ;
Barral, FG ;
Huet, Y ;
Simonneau, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (07) :409-415
[7]   In vivo evaluation of a retrievable vena cava filter - The Dibie-Musset filter: Experimental results [J].
Dibie, A ;
Musset, D ;
Heissler, M ;
Fournet, JC ;
Palau, R ;
Laborde, F .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 21 (02) :151-157
[8]  
EICHELTER P, 1967, REV SURG, V24, P455
[9]   STAGING OF FIXATION AND RETRIEVABILITY OF GREENFIELD FILTERS [J].
GREENFIELD, LJ ;
PROCTOR, MC ;
JAMES, EA ;
ABRAMS, GD ;
MOURSI, MM .
JOURNAL OF VASCULAR SURGERY, 1994, 20 (05) :744-750
[10]  
Hunter D W, 1987, Cardiovasc Intervent Radiol, V10, P32, DOI 10.1007/BF02583304