Determination of uranium by flow injection inductively coupled plasma mass spectrometry

被引:61
作者
Aldstadt, JH [1 ]
Kuo, JM [1 ]
Smith, LL [1 ]
Erickson, MD [1 ]
机构
[1] ARGONNE NATL LAB,DIV CHEM TECHNOL,ARGONNE,IL 60439
关键词
inductively coupled plasma-mass spectrometry; flow injection; uranium;
D O I
10.1016/0003-2670(95)00502-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection (FI) system for the on-line determination of uranium in environmental samples has been characterized and tested. The simple Fl manifold uses off-the-shelf components to precisely manipulate the carrier, sample, eluent, and washing streams through a column separation module. The separation module preconcentrates the sample by a factor of 30 and removes the potentially interfering environmental matrix. TRU(TM), a commercially available resin for extraction chromatography, is packed into a glass separation column (5 X 50 mm) in the limited dispersion (D = 2.8) FI system. Detection is performed by using inductively coupled plasma mass spectrometry (ICP-MS). Results are presented for the U-238 in groundwater samples. The total analysis time for each 5.0-ml aqueous sample is 6 min. The determination of detection limit for U-238 is 0.30 ng/l. By using certified groundwater samples, a precision of 1.1% (n = 4) and accuracy of +/-1.8% for U-238 at 62.6 ng/l were observed. We also report the design and preliminary testing of a second FI-ICP-MS system that minimizes the influence of polyatomic ions (e.g., (UH+)-U-238) On the determination of minor actinides (e.g., Pu-239(+)) by ICP-MS.
引用
收藏
页码:135 / 143
页数:9
相关论文
共 27 条
[1]   DETERMINATION OF TRACE-METALS IN AN OPEN OCEAN WATER REFERENCE MATERIAL BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
MYKYTIUK, AP ;
BERMAN, SS .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (02) :305-308
[2]   DETERMINATION OF TRACE-METALS IN REFERENCE WATER STANDARDS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH ONLINE PRECONCENTRATION [J].
BEAUCHEMIN, D ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1989, 61 (17) :1857-1862
[3]   SAMPLE HANDLING AND PRETREATMENT USING FLOW-INJECTION ANALYSIS [J].
CLARK, GD ;
WHITMAN, DA ;
CHRISTIAN, GD ;
RUZICKA, J .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1990, 21 (05) :357-375
[4]   HYDRIDE INTERFERENCE ON THE DETERMINATION OF MINOR ACTINIDE ISOTOPES BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
CRAIN, JS ;
ALVARADO, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (11) :1223-1227
[5]   ONLINE TRACE ENRICHMENT AND DETERMINATION OF URANIUM IN WATERS BY FLOW-INJECTION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
DADFARNIA, S ;
MCLEOD, CW .
APPLIED SPECTROSCOPY, 1994, 48 (11) :1331-1336
[6]  
DECASTRO MDL, 1989, TRAC-TREND ANAL CHEM, V8, P172
[7]  
FANG Z, 1993, INJECTION SEPARATION
[8]  
Giddings J.C., 1991, Unified separation science
[9]  
GRAY AL, 1991, HDB INDUCTIVELY COUP, P2861
[10]  
HAVEL J, 1992, TALANTA, V7, P795