Solid Phase Extraction of Trace Elements in Water and Tissue Samples on a Mini Column with Diphenylcarbazone Impregnated Nano-TiO2 and Their Determination by Inductively Coupled Plasma Optical Emission Spectrometry

被引:12
作者
Baytak, Sitki [1 ]
Arslan, Zikri [2 ]
机构
[1] Suleyman Demirel Univ, Dept Chem Engn, Fac Engn, TR-32260 Isparta, Turkey
[2] Jackson State Univ, Dept Chem & Biochem, Jackson, MS USA
关键词
Chelating agents; Multi-element determination; Nanoscale materials; Preconcentration; Trace metals; ATOMIC-ABSORPTION-SPECTROMETRY; LIQUID-LIQUID MICROEXTRACTION; IONIC IMPRINTED POLYMER; ENVIRONMENTAL-SAMPLES; SILICA-GEL; HEAVY-METALS; ICP-AES; PRECONCENTRATION; CADMIUM; FE(III);
D O I
10.1002/clen.201400348
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This study presents a simple, robust, and environmentally friendly solid phase extraction procedure for multi-element determination by inductively coupled plasma optical emission spectrometry (ICP-OES) using diphenylcarbazone (DPC) impregnated TiO2 nanopowder (n-TiO2). DPC was successfully impregnated onto n- TiO2 in a colloidal solution. A number of elements, including Co(II), Cr(III), Cu(II), Fe(III), Mn(II), and Zn(II) were quantitatively preconcentrated from aqueous solutions between pH 8 and 8.5 at a flow rate of 2mL min (-)1, and then eluted with 2 mL of 5% HNO3 (v/v). A mini-column packed with 0.12 g DPC impregnated n- TiO2 retained all elements quantitatively from an up to 250 mL multi-element solution (2.5 mu g per analyte) affording an enrichment factor of 125. The limits of detection for the preconcentration of 50 mL blank solutions (n = 12) were 0.28, 0.15, 0.25, 0.22, 0.12, and 0.10 mu g L (1) for Co, Cr, Cu, Fe, Mn, and Zn, respectively. The relative standard deviation for five replicate determinations was 0.8, 3.4, 2.6, 2.2, 1.2, and 3.3% for Co, Cr, Cu, Fe, Mn, and Zn, respectively, at a 5 mu g L-1 level. The method was validated with analysis of freshwater and lobster hepatopancreas certified reference materials, and then applied to the determination of the elements from tap water and lake water samples by ICP-OES.
引用
收藏
页码:822 / 829
页数:8
相关论文
共 42 条
[1]   Selective dispersive liquid-liquid microextraction and preconcentration of Ni(II) into a micro droplet followed by ETAAS determination using a yellow Schiffs base bisazanyl derivative [J].
Alizadeh, Kamal ;
Nemati, Hadi ;
Zohrevand, Somaieh ;
Hashemi, Payman ;
Kakanejadifard, Ali ;
Shamsipur, Mojtaba ;
Ganjali, Mohammad Reza ;
Faridbod, Farnoush .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2013, 33 (02) :916-922
[2]   Analysis of biogenic carbonates by inductively coupled plasma-mass spectrometry (ICP-MS). Flow injection on-line solid-phase preconcentration for trace element determination in fish otoliths [J].
Arslan, Z ;
Paulson, AJ .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 372 (7-8) :776-785
[3]   Preconcentration and matrix elimination for the determination of Pb(II), Cd(II), Ni(II), and Co(II) by 8-hydroxyquinoline anchored poly(styrene-divinylbenzene) microbeads [J].
Askun, Hakan ;
Gulbakan, Basri ;
Celikbicak, Omur ;
Uzun, Cengiz ;
Guven, Gun ;
Salih, Bekir .
JOURNAL OF APPLIED POLYMER SCIENCE, 2008, 107 (04) :2714-2722
[4]   Determination of lead and nickel in environmental samples by flame atomic absorption spectrometry after column solid-phase extraction on Ambersorb-572 with EDTA [J].
Baytak, S ;
Türker, AR .
JOURNAL OF HAZARDOUS MATERIALS, 2006, 129 (1-3) :130-136
[5]   Atomic absorption spectrometric determination of Fe(III) and Cr(III) in various samples after preconcentration by solid-phase extraction with pyridine-2-carbaldehyde thiosemicarbazone [J].
Baytak, S ;
Balaban, A ;
Türker, AR ;
Erk, B .
JOURNAL OF ANALYTICAL CHEMISTRY, 2006, 61 (05) :476-482
[6]   Penicillium digitatum immobilized on pumice stone as a new solid phase extractor for preconcentration and/or separation of trace metals in environmental samples [J].
Baytak, Sitki ;
Kenduzler, Erdal ;
Turker, Ali Rehber ;
Gok, Nuray .
JOURNAL OF HAZARDOUS MATERIALS, 2008, 153 (03) :975-983
[7]  
Baytak S, 2007, ACTA CHIM SLOV, V54, P385
[8]   Preconcentration of trace elements from water samples on a minicolumn of yeast (Yamadazyma spartinae) immobilized TiO2 nanoparticles for determination by ICP-AES [J].
Baytak, Sitki ;
Zereen, Fahmida ;
Arslan, Zikri .
TALANTA, 2011, 84 (02) :319-323
[9]   Preconcentration of copper using double-imprinted polymer via solid phase extraction [J].
Birlik, E ;
Ersöz, A ;
Denizli, A ;
Say, R .
ANALYTICA CHIMICA ACTA, 2006, 565 (02) :145-151
[10]   Separation and preconcentration system for lead and cadmium determination in natural samples using 2-aminoacetylthiophenol modified polyurethane foam [J].
Burham, N. .
DESALINATION, 2009, 249 (03) :1199-1205