Accuracy of CBCT Measurements of a Human Skull

被引:45
作者
Kamburoglu, Kivanc [1 ]
Kolsuz, Eray [1 ]
Kurt, Hakan [1 ]
Kihc, Cenk [2 ]
Ozen, Tuncer [3 ]
Paksoy, Candan Semra [1 ]
机构
[1] Ankara Univ, Dept Oral Diag & Radiol, Fac Dent, TR-06100 Ankara, Turkey
[2] Gulhane Mil Med Acad, Dept Anat, Fac Med, Ankara, Turkey
[3] Gulhane Mil Med Acad, Dent Sci Ctr, Oral Diag & Radiol Dept, Ankara, Turkey
关键词
CBCT; Direct measurement; Radiography; 3-D imaging (imaging; three-dimensional); 3-D reconstruction; Automated measurement; Computed tomography; BEAM COMPUTED-TOMOGRAPHY; CONE-BEAM; LINEAR MEASUREMENTS; CT; IMAGES; RELIABILITY; DOSIMETRY; SCANS;
D O I
10.1007/s10278-010-9339-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to assess the accuracy and reproducibility of cone-beam computed tomography (CBCT) measurements of a human dry skull by comparing them to direct digital caliper measurements. Heated gutta-percha was used to mark 13 specific distances on a human skull, and the distances were directly measured using a digital caliper and on CBCT images obtained with Iluma (3M Imtec, OK, USA) and 3D Accuitomo 170 (3D Accuitomo; J Morita Mfg. Corp., Kyoto, Japan) CBCT imaging systems. Iluma images were obtained at 120 kVp and 3.8 mA and reconstructed using voxel sizes of 0.2 and 0.3 mm(3). Accuitomo images were obtained at 60 kVp and 2 mA and a voxel size of 0.250 mm(3). In addition, 3-D reconstructions were produced from images obtained from both systems. All measurements were made independently by three trained observers and were repeated after an interval of 1 week. Agreement between observers and image type was assessed by calculating Pearson correlation coefficients, with a level of significance set at p < 0.05. Pearson correlation coefficients between readings ranged from 0.995 to 1 for all image types. Correlations among observers were also very high, ranging from 0.992 to 1 for the first reading and from 0.992 to 1 for the second reading for the different image types. All CBCT image measurements were identical and highly correlated with digital caliper measurements. Accuracy of measurements of various distances on a human skull obtained from different CBCT units and image types is comparable to that of digital caliper measurements.
引用
收藏
页码:787 / 793
页数:7
相关论文
共 30 条
  • [1] Accuracy assessment of three-dimensional surface reconstructions of teeth from Cone Beam Computed Tomography scans
    Al-Rawi, B.
    Hassan, B.
    Vandenberge, B.
    Jacobs, R.
    [J]. JOURNAL OF ORAL REHABILITATION, 2010, 37 (05) : 352 - 358
  • [2] Accuracy and reliability of linear cephalometric measurements from cone-beam computed tomography scans of a dry human skull
    Berco, Mauricio
    Rigali, Paul H., Jr.
    Miner, R. Matthew
    DeLuca, Stephelynn
    Anderson, Nina K.
    Will, Leslie A.
    [J]. AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2009, 136 (01) : 17.e1 - 17.e9
  • [3] Linear Accuracy of Cone Beam CT Derived 3D Images
    Brown, April A.
    Scarfe, William C.
    Scheetz, James P.
    Silveira, Anibal M.
    Farman, Allan G.
    [J]. ANGLE ORTHODONTIST, 2009, 79 (01) : 150 - 157
  • [4] Comparative study of jaws with multislice computed tomography and cone-beam computed tomography
    Carrafiello, G.
    Dizonno, M.
    Colli, V.
    Strocchi, S.
    Taubert, S. Pozzi
    Leonardi, A.
    Giorgianni, A.
    Barresi, M.
    Macchi, A.
    Bracchi, E.
    Conte, L.
    Fugazzola, C.
    [J]. RADIOLOGIA MEDICA, 2010, 115 (04): : 600 - 611
  • [5] Comparison of reliability in anatomical landmark identification using two-dimensional digital cephalometrics and three-dimensional cone beam computed tomography in vivo
    Chien, P. C.
    Parks, E. T.
    Eraso, F.
    Hartsfield, J. K., Jr.
    Roberts, W. E.
    Ofner, S.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2009, 38 (05) : 262 - 273
  • [6] Accuracy of linear measurements from cone-beam computed tomography-derived surface models of different voxel sizes
    Damstra, Janalt
    Fourie, Zacharias
    Slater, James J. R. Huddleston
    Ren, Yijin
    [J]. AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (01) : 16.e1 - 16.e6
  • [7] Beam hardening artefacts occur in dental implant scans with the NewTom® cone beam CT but not with the dental 4-row multidetector CT
    Draenert, F. G.
    Coppenrath, E.
    Herzog, P.
    Muller, S.
    Mueller-Lisse, U. G.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (04) : 198 - 203
  • [8] Accuracy of three-dimensional measurements obtained from cone beam computed tomography surface-rendered images for cephalometric analysis: influence of patient scanning position
    Hassan, Bassam
    van der Stelt, Paul
    Sanderink, Gerard
    [J]. EUROPEAN JOURNAL OF ORTHODONTICS, 2009, 31 (02) : 129 - 134
  • [9] Accuracy of linear temporomandibular joint measurements with cone beam computed tomography and digital cephalometric radiography
    Hilgers, ML
    Scarfe, WC
    Scheetz, JP
    Farman, AG
    [J]. AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2005, 128 (06) : 803 - 811
  • [10] Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of view
    Hirsch, E.
    Wolf, U.
    Heinicke, F.
    Silva, M. A. G.
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2008, 37 (05) : 268 - 273