Fracture of the tibial tray after total knee arthroplasty

被引:19
作者
Altintas, F
Sener, N
Ugutmen, E
机构
[1] Dept. Orthopaedics and Traumatology, SSK Goztepe Training Hospital, Goztepe
[2] 81112 Bostanci, Istanbul
关键词
knee; arthroplasty; prosthesis; fracture; breakage; complication;
D O I
10.1016/S0883-5403(99)90212-7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a case in which fracture of the metal tibial plate was thought to be caused by a combination of neutral anatomic rotation of the femoral component and an undersized tibial plate. This resulted in medial subsidence and baseplate fracture with loosening of the polyethylene and dislocation of the prosthesis.
引用
收藏
页码:112 / 114
页数:3
相关论文
共 9 条
[1]   FRACTURE OF THE TIBIAL TRAY OF THE PCA KNEE - A CASE-REPORT OF EARLY FAILURE CAUSED BY IMPROPER DESIGN [J].
FLIVIK, G ;
LJUNG, P ;
RYDHOLM, U .
ACTA ORTHOPAEDICA SCANDINAVICA, 1990, 61 (01) :26-28
[2]  
Gradisar I A Jr, 1989, J Arthroplasty, V4, P27, DOI 10.1016/S0883-5403(89)80050-6
[3]  
Hungerford DS, 1995, AAOS INSTR COURS LEC, V44, P455
[4]  
Krackow KA, 1995, ORTHOPAEDIC KNOWLEDG, P269
[5]  
LEWIS P, 1994, CLIN ORTHOP RELAT R, P11
[6]   FRACTURE OF THE TIBIAL TRAY FOLLOWING PCA KNEE REPLACEMENT - A REPORT OF 2 CASES [J].
MARUYAMA, M ;
TERAYAMA, K ;
SUNOHARA, H ;
ADACHI, T ;
SUZUKI, S ;
FUKUZAWA, T .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1994, 113 (06) :330-333
[7]  
McPherson E J, 1993, J Arthroplasty, V8, P561
[8]   FRACTURE OF THE METALLIC TIBIAL TRAY FOLLOWING TOTAL KNEE REPLACEMENT - REPORT OF 2 CASES [J].
SCOTT, RD ;
EWALD, FC ;
WALKER, PS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (05) :780-782
[9]  
SKINNER HB, 1987, ORTHOPEDICS, V10, P581