Characteristics of the photoneutron contamination present in a high-energy radiotherapy treatment room

被引:25
作者
Garnica-Garza, HM [1 ]
机构
[1] Univ Autonoma Estado Mexico, Fac Med, Toluca 50180, Mexico
关键词
D O I
10.1088/0031-9155/50/3/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The photoneutron contamination arising from a high-energy medical lineal accelerator is calculated using Monte Carlo simulation, as a function of the radiation field size. The information is used to model the neutron propagation in a radiotherapy treatment room and the transmission across concrete mazes. The Monte Carlo code MCNP4C is used to model the main components of a medical lineal accelerator. Simulations were performed to calculate the photoneutron yields and spectra as a function of the radiation field size. The yield of contaminant photoneutrons is observed to increase with the size of the radiation beam, but the energy spectra remain the same, suggesting that the contamination arises from above the movable collimator. The transport of the photoneutrons across a treatment room corroborates the validity of empirical models, but the transmission across a concrete maze produces a dose-equivalent tenth-value layer that differs from previous data.
引用
收藏
页码:531 / 539
页数:9
相关论文
共 9 条
[1]   NEUTRON YIELDS FROM TARGETS BOMBARDED BY ELECTRONS [J].
BARBER, WC ;
GEORGE, WD .
PHYSICAL REVIEW, 1959, 116 (06) :1551-1559
[2]  
Briesmeister J.F., 2000, LA-13709M
[3]   TRANSPORT OF ACCELERATOR PRODUCED NEUTRONS IN A CONCRETE ROOM [J].
MCCALL, RC ;
JENKINS, TM ;
SHORE, RA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (01) :1593-1602
[4]  
MCCALL RC, 1979, 2441 STANF LIN ACC R
[5]  
*NCRP, 1995, 79 NCRP
[6]  
Ongaro C., 1999, Proceedings of the 1999 Particle Accelerator Conference (Cat. No.99CH36366), P2531, DOI 10.1109/PAC.1999.792756
[7]   BEAM - A MONTE-CARLO CODE TO SIMULATE RADIOTHERAPY TREATMENT UNITS [J].
ROGERS, DWO ;
FADDEGON, BA ;
DING, GX ;
MA, CM ;
WE, J ;
MACKIE, TR .
MEDICAL PHYSICS, 1995, 22 (05) :503-524
[8]   Comparison of measured and Monte Carlo calculated dose distributions from the NRC linac [J].
Sheikh-Bagheri, D ;
Rogers, DWO ;
Ross, CK ;
Seuntjens, JP .
MEDICAL PHYSICS, 2000, 27 (10) :2256-2266
[9]   Photonuclear physics in radiation transport - II: Implementation [J].
White, MC ;
Little, RC ;
Chadwick, MB ;
Young, PG ;
MacFarlane, RE .
NUCLEAR SCIENCE AND ENGINEERING, 2003, 144 (02) :174-189