Effect of Pneumatic Tubing System Transport on Platelet Apheresis Units

被引:4
|
作者
Zilberman-Rudenko, Jevgenia [1 ,2 ]
Zhao, Frank Z. [1 ]
Reitsma, Stephanie E. [2 ]
Mitrugno, Annachiara [2 ]
Pang, Jiaqing [2 ]
Shatzel, Joseph J. [3 ]
Rick, Beth [1 ]
Tyrrell, Christina [4 ]
Hasan, Wohaib [4 ]
McCarty, Owen J. T. [2 ]
Schreiber, Martin A. [1 ]
机构
[1] Oregon Hlth & Sci Univ, Dept Surg, Div Trauma Crit Care & Acute Care Surg, Portland, OR 97201 USA
[2] Oregon Hlth & Sci Univ, Sch Med, Dept Biomed Engn, 3303 SW Bond Ave, Portland, OR 97201 USA
[3] Oregon Hlth & Sci Univ, Dept Med, Div Hematol & Med Oncol, Portland, OR 97201 USA
[4] Oregon Hlth & Sci Univ, Knight Cardiovasc Inst, Portland, OR 97201 USA
基金
美国国家卫生研究院;
关键词
Platelet apheresis units; Platelet function; Hemo-dynamics; Hemostasis; Pneumatic tubing system; VON-WILLEBRAND-FACTOR; FUNCTION ASSAY (PFA)-100; GLYCOPROTEIN-IB-ALPHA; TUBE TRANSPORT; HEMOSTATIC FUNCTION; BLOOD-STREAM; STORAGE; TRAUMA; TRANSFUSION; ACTIVATION;
D O I
10.1007/s13239-018-0361-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Platelet apheresis units are transfused into patients to mitigate or prevent bleeding. In a hospital, platelet apheresis units are transported from the transfusion service to the healthcare teams via two methods: a pneumatic tubing system (PTS) or ambulatory transport. Whether PTS transport affects the activity and utility of platelet apheresis units is unclear. We quantified the gravitational forces and transport time associated with PTS and ambulatory transport within our hospital. Washed platelets and supernatants were prepared from platelet apheresis units prior to transport as well as following ambulatory or PTS transport. For each group, we compared resting and agonist-induced platelet activity and platelet aggregate formation on collagen or von Willebrand factor (VWF) under shear, platelet VWF-receptor expression and VWF multimer levels. Subjection of platelet apheresis units to rapid acceleration/deceleration forces during PTS transport did not pre-activate platelets or their ability to activate in response to platelet agonists as compared to ambulatory transport. Platelets within platelet apheresis units transported via PTS retained their ability to adhere to surfaces of VWF and collagen under shear, although platelet aggregation on collagen and VWF was diminished as compared to ambulatory transport. VWF multimer levels and platelet GPIb receptor expression was unaffected by PTS transport as compared to ambulatory transport. Subjection of platelet apheresis units to PTS transport did not significantly affect the baseline or agonist-induced levels of platelet activation as compared to ambulatory transport. Our case study suggests that PTS transport may not significantly affect the hemostatic potential of platelets within platelet apheresis units.
引用
收藏
页码:515 / 527
页数:13
相关论文
共 31 条
  • [1] Effect of Pneumatic Tubing System Transport on Platelet Apheresis Units
    Jevgenia Zilberman-Rudenko
    Frank Z. Zhao
    Stephanie E. Reitsma
    Annachiara Mitrugno
    Jiaqing Pang
    Joseph J. Shatzel
    Beth Rick
    Christina Tyrrell
    Wohaib Hasan
    Owen J. T. McCarty
    Martin A. Schreiber
    Cardiovascular Engineering and Technology, 2018, 9 : 515 - 527
  • [2] Triple apheresis platelet concentrate quality after pneumatic tube system, conveyor box, and courier transport: An observational study
    Reichert, Lena
    Wallner, Stefan
    Burkhardt, Ralph
    Offner, Robert
    Ahrens, Norbert
    Hahnel, Viola
    HEALTH SCIENCE REPORTS, 2022, 5 (03)
  • [3] The effect of plasma removal from apheresis platelet concentrates on platelet function
    Jilma-Stohlawetz, P.
    Eichelberger, B.
    List, J.
    Jilma, B.
    Panzer, S.
    Leitner, G.
    VOX SANGUINIS, 2012, 102 (03) : 258 - 260
  • [4] Investigating the effect of a platelet additive solution on apheresis platelet and fibrin network ultrastructure
    Pretorius, E.
    Crookes, R.
    Oberholzer, H. M.
    van der Spuy, W. J.
    TRANSFUSION AND APHERESIS SCIENCE, 2010, 43 (03) : 347 - 351
  • [5] Sample transport by pneumatic tube system alters results of multiple electrode aggregometry but not rotational thromboelastometry
    Glas, Michael
    Mauer, Dietmar
    Kassas, Hazim
    Volk, Thomas
    Kreuer, Sascha
    PLATELETS, 2013, 24 (06) : 454 - 461
  • [6] THE INTERVENTION EFFECT OF HIGH QUALITY NURSING MODE ON APHERESIS PLATELET DONORS
    Teng, Yali
    Shao, Qing
    ACTA MEDICA MEDITERRANEA, 2022, 38 (02): : 1371 - 1376
  • [7] Therapeutic Effect of Apheresis and Buffy Coat-Derived Platelet Concentrates
    Zimmermann, Robert
    CLINICAL LABORATORY, 2024, 70 (08) : 1431 - 1433
  • [8] Pneumatic tube transport affects platelet function measured by multiplate electrode aggregometry
    Thalen, Simon
    Forsling, Ida
    Eintrei, Jaak
    Soderblom, Lisbeth
    Antovic, Jovan P.
    THROMBOSIS RESEARCH, 2013, 132 (01) : 77 - 80
  • [9] Pre-analytical effects of pneumatic tube transport on impedance platelet aggregometry
    Bolliger, Daniel
    Seeberger, Manfred D.
    Tanaka, Kenichi A.
    Dell-Kuster, Salome
    Gregor, Michael
    Zenklusen, Urs
    Grapow, Martin
    Tsakiris, Dimitrios A.
    Filipovic, Miodrag
    PLATELETS, 2009, 20 (07) : 458 - 465
  • [10] Evaluation of a rapid colorimetric assay for detection of bacterial contamination in apheresis and pooled random-donor platelet units
    Heaton, W. Andrew
    Good, Caryn E.
    Galloway-Haskins, Rakijah
    Yomtovian, Roslyn A.
    Jacobs, Michael R.
    TRANSFUSION, 2014, 54 (06) : 1634 - 1641