Acetabular positioning is more consistent with the use of a novel miniature computer-assisted device

被引:9
作者
Jacob, Ivan [1 ]
Benson, Jessica [2 ]
Shanaghan, Kate [1 ]
Gonzalez Della Valle, Alejandro [1 ]
机构
[1] Hosp Special Surg, Dept Orthpaed Surg, 535 E 70th St, New York, NY 10021 USA
[2] Intellijoint Surg Inc., Waterloo, ON, Canada
关键词
Total hip arthroplasty; Computer-assisted surgery; Cup position; Anterior pelvic plane; TOTAL HIP-ARTHROPLASTY; COMPONENT ORIENTATION; IMAGELESS NAVIGATION; CUP ANTEVERSION; ACCURACY; PLACEMENT; INCLINATION; REPLACEMENT; SYSTEM; TOOL;
D O I
10.1007/s00264-020-04484-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction Computer-assisted surgery (CAS) relying on registration of the anterior pelvic plane (APP) allows precise acetabular component placement. We determined the variability of cup placement in patients who underwent THA with and without the use of CAS that does not rely on the registration of APP. Methods Fifty-one patients who underwent staged-bilateral THAs, one without CAS (control), and a subsequent one with CAS (study group), were included. Acetabular inclination and anteversion were measured on standardized post-operative radiographs. Variance in cup position and Lewinnek's zone proportionality were compared between the groups. Multiple regressions were performed to identify factors affecting variability in acetabular component placement. Results The mean inclination for the control and study group was 42.7 degrees (SD 4.5) and 42.5 degrees (SD 2.9), respectively. The inclination variance was 20.5 degrees and 8.2 degrees respectively (p = < 0.001). Cup inclination was more consistent in the study than in the control group (deviation from the mean: 2.3 degrees vs. 3.8 degrees, p < 0.001). The mean anteversion for the control and the study group was 25.5 degrees (SD 7.4) and 26.8 degrees (SD 4.3), respectively. The anteversion variance was 54.2 degrees and 18.2 degrees respectively (p = <0.001). Consistency in cup anteversion was significantly improved with CAS (deviation from the mean: 3.4 degrees vs. 5.8 degrees; p = 0.002). Lewinnek's zone proportionality was not affected by the use of CAS. In the linear regression analysis, CAS significantly increased consistency in cup inclination (p = 0.01). Patient's factors including BMI and laterality affected consistency of cup placement. Conclusion CAS without referencing the APP allows a more consistent orientation of the acetabular component when compared to freehand placement.
引用
收藏
页码:429 / 435
页数:7
相关论文
共 50 条
  • [41] Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials?
    Cheng, Tao
    Zhao, Song
    Peng, Xiaochun
    Zhang, Xianlong
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (07) : 1307 - 1322
  • [42] A Novel Technique for Restricted Kinematic Alignment Total Knee Arthroplasty Without Computer-Assisted Devices
    Sakakibara, Yuzuru
    Yasutani, Yusuke
    Oyama, Akimitsu
    Teramoto, Atsushi
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (08)
  • [43] Computer-assisted virtual preoperative planning in orthopedic surgery for acetabular fractures based on actual computed tomography data
    Wang, Guang-Ye
    Huang, Wen-Jun
    Song, Qi
    Qin, Yun-Tian
    Liang, Jin-Feng
    COMPUTER ASSISTED SURGERY, 2016, 21 (01): : 160 - 165
  • [44] Preliminary application of computer-assisted patient-specific acetabular navigational template for total hip arthroplasty in adult single development dysplasia of the hip
    Zhang, Yuan Z.
    Chen, Bin
    Lu, Sheng
    Yang, Yong
    Zhao, Jian M.
    Liu, Rui
    Li, Yan B.
    Pei, Guo X.
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2011, 7 (04) : 469 - 474
  • [45] Effect of surgical experience on imageless computer-assisted femoral component positioning in hip resurfacing – a preclinical study
    Maik Stiehler
    Jens Goronzy
    Stephan Kirschner
    Albrecht Hartmann
    Torsten Schäfer
    Klaus-Peter Günther
    European Journal of Medical Research, 20
  • [46] Computer-assisted stereotaxic navigation improves the accuracy of mechanical alignment and component positioning in total knee arthroplasty
    Liow, Ming Han Lincoln
    Goh, Graham Seow-Hng
    Pang, Hee-Nee
    Tay, Darren Keng Jin
    Lo, Ngai Nung
    Yeo, Seng Jin
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2016, 136 (08) : 1173 - 1180
  • [47] Virtual implantation of a novel LVAD: toward computer-assisted surgery for heart failure
    Anselmi, Amedeo
    Collin, Sophie
    Haigron, Pascal
    Verhoye, Jean-Philippe
    Flecher, Erwan
    JOURNAL OF SURGICAL RESEARCH, 2016, 205 (01) : 204 - 207
  • [48] Optimization of cup positioning in THA - Comparison between conventional mechanical instrumentation and computer-assisted implanted cups by using the orthopilot navigation system
    Ottersbach, A
    Haaker, R
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2005, 143 (06): : 611 - 615
  • [49] A combination of three-dimensional printing and computer-assisted virtual surgical procedure for preoperative planning of acetabular fracture reduction
    Zeng, Canjun
    Xing, Weirong
    Wu, Zhanglin
    Huang, Huajun
    Huang, Wenhua
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (10): : 2223 - 2227
  • [50] Advancing computer-assisted orthopaedic surgery using a hexapod device for closed diaphyseal fracture reduction
    Du, Hailong
    Hu, Lei
    Li, Changsheng
    Wang, Tianmiao
    Zhao, Lu
    Li, Yang
    Mao, Zhi
    Liu, Daohong
    Zhang, Lining
    He, Chunqing
    Zhang, Licheng
    Hou, Hongping
    Zhang, Lihai
    Tang, Peifu
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2015, 11 (03) : 348 - 359