Acetabular positioning is more consistent with the use of a novel miniature computer-assisted device

被引:9
作者
Jacob, Ivan [1 ]
Benson, Jessica [2 ]
Shanaghan, Kate [1 ]
Gonzalez Della Valle, Alejandro [1 ]
机构
[1] Hosp Special Surg, Dept Orthpaed Surg, 535 E 70th St, New York, NY 10021 USA
[2] Intellijoint Surg Inc., Waterloo, ON, Canada
关键词
Total hip arthroplasty; Computer-assisted surgery; Cup position; Anterior pelvic plane; TOTAL HIP-ARTHROPLASTY; COMPONENT ORIENTATION; IMAGELESS NAVIGATION; CUP ANTEVERSION; ACCURACY; PLACEMENT; INCLINATION; REPLACEMENT; SYSTEM; TOOL;
D O I
10.1007/s00264-020-04484-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction Computer-assisted surgery (CAS) relying on registration of the anterior pelvic plane (APP) allows precise acetabular component placement. We determined the variability of cup placement in patients who underwent THA with and without the use of CAS that does not rely on the registration of APP. Methods Fifty-one patients who underwent staged-bilateral THAs, one without CAS (control), and a subsequent one with CAS (study group), were included. Acetabular inclination and anteversion were measured on standardized post-operative radiographs. Variance in cup position and Lewinnek's zone proportionality were compared between the groups. Multiple regressions were performed to identify factors affecting variability in acetabular component placement. Results The mean inclination for the control and study group was 42.7 degrees (SD 4.5) and 42.5 degrees (SD 2.9), respectively. The inclination variance was 20.5 degrees and 8.2 degrees respectively (p = < 0.001). Cup inclination was more consistent in the study than in the control group (deviation from the mean: 2.3 degrees vs. 3.8 degrees, p < 0.001). The mean anteversion for the control and the study group was 25.5 degrees (SD 7.4) and 26.8 degrees (SD 4.3), respectively. The anteversion variance was 54.2 degrees and 18.2 degrees respectively (p = <0.001). Consistency in cup anteversion was significantly improved with CAS (deviation from the mean: 3.4 degrees vs. 5.8 degrees; p = 0.002). Lewinnek's zone proportionality was not affected by the use of CAS. In the linear regression analysis, CAS significantly increased consistency in cup inclination (p = 0.01). Patient's factors including BMI and laterality affected consistency of cup placement. Conclusion CAS without referencing the APP allows a more consistent orientation of the acetabular component when compared to freehand placement.
引用
收藏
页码:429 / 435
页数:7
相关论文
共 50 条
  • [11] Reliability and Correlation of the Force-PRO Device and Computer-Assisted Navigation System for Measurement of Acetabular Cup Position in Total Hip Arthroplasty
    Vijittrakarnrung, Chaiyanun
    Manupibul, Udomporn
    Charoensuk, Warakorn
    Wongsak, Siwadol
    Sa-Ngasoongsong, Paphon
    SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2020, 36
  • [12] Effect of surgical experience on imageless computer-assisted femoral component positioning in hip resurfacing - a preclinical study
    Stiehler, Maik
    Goronzy, Jens
    Kirschner, Stephan
    Hartmann, Albrecht
    Schaefer, Torsten
    Guenther, Klaus-Peter
    EUROPEAN JOURNAL OF MEDICAL RESEARCH, 2015, 20
  • [13] Computer-Assisted Total Hip Arthroplasty Improves Acetabular Prosthesis Placement Accuracy: A Multicenter, Randomized Controlled Clinical Study
    Zhou, Ge
    Geng, Xiao
    Zhang, Ming
    Sun, Zhiwen
    Li, Feng
    Zhao, Minwei
    Tian, Hua
    ORTHOPAEDIC SURGERY, 2024, 16 (12) : 3078 - 3087
  • [14] Computer-assisted positioning of the acetabular cup for total hip arthroplasty based on joint kinematics without prior imaging: preliminary results with computed tomographic assessment
    Laffargue, P.
    Pinoit, Y.
    Tabutin, J.
    Giraud, F.
    Puget, J.
    Migaud, H.
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2006, 92 (04): : 316 - 325
  • [15] Computer-assisted needle positioning for liver tumour radiofrequency ablation (RFA)
    Mundeleer, Laurent
    Wikler, David
    Leloup, Thierry
    Lucidi, Valerio
    Donckier, Vincent
    Warzee, Nadine
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2009, 5 (04) : 458 - 464
  • [16] A Modified Novel Technique for Condylar Positioning in Mandibular Bilateral Sagittal Split Osteotomy Using Computer-Assisted Designed and Computer-Assisted Manufactured Surgical Guides
    Cortese, Antonio
    Chandran, Ravi
    Borri, Antonio
    Cataldo, Emilio
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2019, 77 (05) : 1069.e1 - 1069.e9
  • [17] Use of computer-assisted surgery in the orbit
    Campbell, Ashley A.
    Mahoney, Nicholas R.
    ORBIT-THE INTERNATIONAL JOURNAL ON ORBITAL DISORDERS-OCULOPLASTIC AND LACRIMAL SURGERY, 2022, 41 (02): : 226 - 234
  • [18] Analysis of anatomic positioning in computer-assisted and conventional anterior cruciate ligament reconstruction
    Plaweski, S.
    Rossi, J.
    Merloz, P.
    Julliard, R.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2011, 97 (06) : S80 - S85
  • [19] Is computer-assisted positioning of the cup necessary in total hip replacement?
    Lahmer, A
    Wiesel, U
    Boerner, M
    CARS 2000: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2000, 1214 : 964 - 964
  • [20] Novel Approach to Reducing Discrepancies in Radiographic and Navigational Limb Alignments in Computer-Assisted TKA
    Chung, Byung June
    Dileep, Isaac
    Chang, Chong Bum
    Kang, Yeon Gwi
    Park, Yong Bum
    Kim, Tae Kyun
    ORTHOPEDICS, 2010, 33 (10) : 62 - 67