Acetabular positioning is more consistent with the use of a novel miniature computer-assisted device

被引:9
|
作者
Jacob, Ivan [1 ]
Benson, Jessica [2 ]
Shanaghan, Kate [1 ]
Gonzalez Della Valle, Alejandro [1 ]
机构
[1] Hosp Special Surg, Dept Orthpaed Surg, 535 E 70th St, New York, NY 10021 USA
[2] Intellijoint Surg Inc., Waterloo, ON, Canada
关键词
Total hip arthroplasty; Computer-assisted surgery; Cup position; Anterior pelvic plane; TOTAL HIP-ARTHROPLASTY; COMPONENT ORIENTATION; IMAGELESS NAVIGATION; CUP ANTEVERSION; ACCURACY; PLACEMENT; INCLINATION; REPLACEMENT; SYSTEM; TOOL;
D O I
10.1007/s00264-020-04484-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction Computer-assisted surgery (CAS) relying on registration of the anterior pelvic plane (APP) allows precise acetabular component placement. We determined the variability of cup placement in patients who underwent THA with and without the use of CAS that does not rely on the registration of APP. Methods Fifty-one patients who underwent staged-bilateral THAs, one without CAS (control), and a subsequent one with CAS (study group), were included. Acetabular inclination and anteversion were measured on standardized post-operative radiographs. Variance in cup position and Lewinnek's zone proportionality were compared between the groups. Multiple regressions were performed to identify factors affecting variability in acetabular component placement. Results The mean inclination for the control and study group was 42.7 degrees (SD 4.5) and 42.5 degrees (SD 2.9), respectively. The inclination variance was 20.5 degrees and 8.2 degrees respectively (p = < 0.001). Cup inclination was more consistent in the study than in the control group (deviation from the mean: 2.3 degrees vs. 3.8 degrees, p < 0.001). The mean anteversion for the control and the study group was 25.5 degrees (SD 7.4) and 26.8 degrees (SD 4.3), respectively. The anteversion variance was 54.2 degrees and 18.2 degrees respectively (p = <0.001). Consistency in cup anteversion was significantly improved with CAS (deviation from the mean: 3.4 degrees vs. 5.8 degrees; p = 0.002). Lewinnek's zone proportionality was not affected by the use of CAS. In the linear regression analysis, CAS significantly increased consistency in cup inclination (p = 0.01). Patient's factors including BMI and laterality affected consistency of cup placement. Conclusion CAS without referencing the APP allows a more consistent orientation of the acetabular component when compared to freehand placement.
引用
收藏
页码:429 / 435
页数:7
相关论文
共 50 条
  • [1] Acetabular positioning is more consistent with the use of a novel miniature computer-assisted device
    Ivan Jacob
    Jessica Benson
    Kate Shanaghan
    Alejandro Gonzalez Della Valle
    International Orthopaedics, 2020, 44 : 429 - 435
  • [2] Accuracy of Acetabular Component Positioning Using Computer-assisted Navigation in Direct Anterior Total Hip Arthroplasty
    Bradley, Michael P.
    Benson, Jessica R.
    Muir, Jeffrey M.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2019, 11 (04)
  • [3] Enhanced acetabular component positioning through computer-assisted navigation
    Thomas Ybinger
    Wolfgang Kumpan
    International Orthopaedics, 2007, 31 : 35 - 38
  • [4] Computer-assisted surgery for acetabular cup positioning in total hip arthroplasty: comparative prospective randomized study
    Parratte, S.
    Argenson, J.-N.
    Flecher, X.
    Aubaniac, J.-M.
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2007, 93 (03): : 238 - 246
  • [5] Does Robotic-Assisted Computer Navigation Affect Acetabular Cup Positioning in Total Hip Arthroplasty in the Obese Patient? A Comparison Study
    Gupta, Asheesh
    Redmond, John M.
    Hammarstedt, Jon E.
    Petrakos, Alexandra E.
    Vemula, S. Pavan
    Domb, Benjamin G.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (12) : 2204 - 2207
  • [6] Acetabular Component Positioning in Total Hip Arthroplasty With and Without a Computer-Assisted System: A Prospective, Randomized and Controlled Study
    Gurgel, Henrique M. C.
    Croci, Alberto T.
    Cabrita, Henrique A. B. A.
    Vicente, Jose Ricardo N.
    Leonhardt, Marcos C.
    Rodrigues, Joao Carlos
    JOURNAL OF ARTHROPLASTY, 2014, 29 (01) : 167 - 171
  • [7] Accuracy of acetabular cup placement in computer-assisted, minimally-invasive THR in a lateral decubitus position
    Sendtner, Ernst
    Schuster, Tibor
    Woerner, Michael
    Kalteis, Thomas
    Grifka, Joachim
    Renkawitz, Tobias
    INTERNATIONAL ORTHOPAEDICS, 2011, 35 (06) : 809 - 815
  • [8] Direct Anatomic Registration in Computer-Assisted Total Hip Arthroplasty Improves Accuracy of Acetabular Cup Alignment
    Patamarat, Apisit
    Laoruengthana, Artit
    Kitcharanant, Nitchanant
    Khantasit, Jakkaphong
    Thremthakanpon, Witoon
    ARTHROPLASTY TODAY, 2023, 21
  • [9] Use of bone anchoring device in electromagnetic computer-assisted navigation in lateral skull base surgery
    Bernardeschi, Daniele
    Nguyen, Yann
    Villepelet, Aude
    Ferrary, Evelyne
    Mazalaigue, Stephane
    Kalamarides, Michel
    Sterkers, Olivier
    ACTA OTO-LARYNGOLOGICA, 2013, 133 (10) : 1047 - 1052
  • [10] Abnormal spinopelvic mobility as a risk factor for acetabular placement error in total hip arthroplasty using optical computer-assisted surgical navigation system
    Jang, S. J.
    Vigdorchik, J. M.
    Windsor, E. W.
    Schwarzkopf, R.
    Mayman, D. J.
    Sculco, P. K.
    BONE & JOINT OPEN, 2022, 3 (06): : 475 - +